COVERA AVSM09100Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 9 mm x 100 mm (80 cm delivery catheter) (straight) · Class 3 · Implant
All models of COVERA (65)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741106361), received between Oct 20, 2021 and May 16, 2025 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Misfire” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Misfire4
- Device Contamination with Chemical or Other Material3
- Loss of or Failure to Bond3
- Packaging Problem3
- Positioning Failure3
- Material Deformation2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Restenosis1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 16, 2025Material Deformation; Material Deformation
ON, (B)(6) 2025 , A PATIENT UNDERWENT A STENT GRAFT PLACEMENT PROCEDURE USING THE COVERA VASCULAR COVERED STENT IN RECURRENT STENOSIS VENOUS ANASTOMOSIS INTO THE BASILIC VEIN. DURING THE PROCEDURE, THE STENT DEPLOYED IN LEFT ARM. PHYSICIAN STATED THAT THE DISTAL ENDS OF THE STENT ARE FOLDED IN. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22034289 ↗MalfunctionDec 3, 2024Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY NOT DEPLOYED. IT WAS FURTHER REPORTED THAT THE DISTAL PORTION OF THE DEVICE APPEARED BUNCHED UP. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20829296 ↗MalfunctionJun 10, 2024Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem
IT WAS REPORTED THAT DURING PREPARATION OF A STENT GRAFT PLACEMENT PROCEDURE, THE PACKAGE WAS ALLEGEDLY FOUND TO BE ARRIVED WITH WATER DAMAGE. IT WAS FURTHER REPORTED THAT THERE WAS WATER DAMAGE TO THE INNER PACKAGING AND THE DEVICE WAS ALLEGEDLY FOUND TO BE DAMAGED. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19495993 ↗MalfunctionJun 10, 2024Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem
IT WAS REPORTED THAT DURING PREPARATION OF A STENT GRAFT PLACEMENT PROCEDURE, THE PACKAGE WAS ALLEGEDLY FOUND TO BE ARRIVED WITH WATER DAMAGE. IT WAS FURTHER REPORTED THAT THERE WAS WATER DAMAGE TO THE INNER PACKAGING AND THE DEVICE WAS ALLEGEDLY FOUND TO BE DAMAGED. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19495990 ↗MalfunctionJun 10, 2024Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem
IT WAS REPORTED THAT DURING PREPARATION OF A STENT GRAFT PLACEMENT PROCEDURE, THE PACKAGE WAS ALLEGEDLY FOUND TO BE ARRIVED WITH WATER DAMAGE. IT WAS FURTHER REPORTED THAT THERE WAS WATER DAMAGE TO THE INNER PACKAGING AND THE DEVICE WAS ALLEGEDLY FOUND TO BE DAMAGED. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19495989 ↗MalfunctionMar 26, 2024Fracture; Material Puncture/Hole; Fracture; Material Puncture/Hole
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE LEFT CEPHALIC VEIN AT CEPHALIC ARCH VIA THE DISTAL CEPHALIC VEIN, THE STENT ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT ADDITIONAL STENT WAS PLACED INSIDE THE RUPTURED STENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18984423 ↗MalfunctionAug 8, 2023Misfire; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY NOT FULLY DEPLOYED AND WAS COMPACTED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17479553 ↗MalfunctionDec 12, 2022Loss of or Failure to Bond; Positioning Failure; Misfire; Loss of or Failure to Bond; Positioning Failure; Misfire
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15957240 ↗MalfunctionAug 1, 2022Loss of or Failure to Bond; Positioning Failure; Misfire; Loss of or Failure to Bond; Positioning Failure; Misfire
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15137456 ↗MalfunctionJun 8, 2022Loss of or Failure to Bond; Misfire; Loss of or Failure to Bond; Misfire
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY PARTIALLY DEPLOYED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14630676 ↗InjuryOct 20, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY FOUR MONTHS AND TWENTY SEVEN DAYS POST INDEX PROCEDURE, THE SUBJECT DEVELOPED 100% OF TARGET LESION STENOSIS. A STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE SUBJECT EXPIRED DUE TO SEPTICEMIA, SPINAL INFECTION AND URINARY TRACT INFECTION (UTI) WITH BOWEL BLADDER FISTULA AND THE CAUSE OF DEATH WAS NOT RELATED TO THE DEVICE, NOT RELATED TO PROCEDURE AND NOT RELATED TO ARTERIOVENOUS (AV) ACCESS CIRCUIT.
Patient problems: Restenosis
MAUDE report 12666320 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page