COVERA AVFM08100Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 8 mm x 100 mm (80 cm delivery catheter) (flared) · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741106538), received between Aug 17, 2020 and Sep 15, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Material Deformation2
- Adverse Event Without Identified Device or Use Problem1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Overstimulation of Tissue1
- Injury1
- Ischemia1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Material Deformation
On (b)(6) 2026, a patient underwent a stent placement procedure using a Covera Vascular Covered Stent. It was reported that additional interventions were performed in (b)(6) 2026. During the (b)(6) 2026 procedure, the stent demonstrated evidence of collapse, and balloon angioplasty was performed. On (b)(6) 2026, recurrent collapse of the proximal portion of the stent was observed. The physician performed balloon angioplasty using an 8 × 80 mm Vaccess balloon and subsequently relined the Covera stent with a Wraysody stent. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50769778 ↗MalfunctionJul 19, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE BRACHIOCEPHALIC FISTULA, THE STENT ALLEGEDLY CRUNCHED UP. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY DID NOT COVER THE ENTIRE AREA INTENDED. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19781371 ↗InjurySep 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT DURING POST-DILATION OF THE VASCULAR COVERED STENT, SPASM OF THE OUTFLOW VEIN OCCURRED AND WAS RESOLVED WITH PTA. THE EVENT WAS CONSIDERED RESOLVED AND THE PATIENT WAS DISCHARGED THE SAME DAY. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Device Overstimulation of Tissue; Injury; Ischemia
MAUDE report 10605778 ↗InjuryAug 17, 2020Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY ONE MONTH POST STENT PLACEMENT THROUGH CEPHALIC ARCH, THE HEALTH CARE PROVIDER IDENTIFIED UNDER IMAGING AN ALLEGED STENT COMPRESSION. IT WAS FURTHER REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE TO TREAT THE NARROWING. THE PATIENT WAS STABLE POST ADDITIONAL BALLOON DILATATION.
Patient problems: Pain
MAUDE report 10412370 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.