COVERA AVFM08100Recalls & reported problems

Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 8 mm x 100 mm (80 cm delivery catheter) (flared) · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741106538), received between Aug 17, 2020 and Sep 15, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 22, 2022Open, Classified

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗

Most reported problems

Device problems

  • Material Deformation2
  • Adverse Event Without Identified Device or Use Problem1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Overstimulation of Tissue1
  • Injury1
  • Ischemia1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Material Deformation

    On (b)(6) 2026, a patient underwent a stent placement procedure using a Covera Vascular Covered Stent. It was reported that additional interventions were performed in (b)(6) 2026. During the (b)(6) 2026 procedure, the stent demonstrated evidence of collapse, and balloon angioplasty was performed. On (b)(6) 2026, recurrent collapse of the proximal portion of the stent was observed. The physician performed balloon angioplasty using an 8 × 80 mm Vaccess balloon and subsequently relined the Covera stent with a Wraysody stent. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769778 ↗
  • MalfunctionJul 19, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE BRACHIOCEPHALIC FISTULA, THE STENT ALLEGEDLY CRUNCHED UP. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY DID NOT COVER THE ENTIRE AREA INTENDED. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19781371 ↗
  • InjurySep 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT DURING POST-DILATION OF THE VASCULAR COVERED STENT, SPASM OF THE OUTFLOW VEIN OCCURRED AND WAS RESOLVED WITH PTA. THE EVENT WAS CONSIDERED RESOLVED AND THE PATIENT WAS DISCHARGED THE SAME DAY. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.

    Patient problems: Device Overstimulation of Tissue; Injury; Ischemia

    MAUDE report 10605778 ↗
  • InjuryAug 17, 2020Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT APPROXIMATELY ONE MONTH POST STENT PLACEMENT THROUGH CEPHALIC ARCH, THE HEALTH CARE PROVIDER IDENTIFIED UNDER IMAGING AN ALLEGED STENT COMPRESSION. IT WAS FURTHER REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE TO TREAT THE NARROWING. THE PATIENT WAS STABLE POST ADDITIONAL BALLOON DILATATION.

    Patient problems: Pain

    MAUDE report 10412370 ↗

Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices

All System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.