Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741106569), received between Nov 7, 2024 and Jan 8, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 22, 2022Open, Classified

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Misfire1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2026Misfire

    ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING A COVERA STENT. IT WAS REPORTED THAT WHEN THE STENT WAS DEPLOYED, PART OF THE STENT ALLEGEDLY STUCK TO THE PAD AND WOULD NOT FULLY OPEN. THE PROCEDURE WAS COMPLETED USING ANOTHER STENT. THERE WERE NO REPORTED PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24000715 ↗
  • MalfunctionNov 7, 2024Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE LEFT UPPER ARM FISTULA, THE STENT SUCCESSFULLY UNLOCKED AND BEGAN TURNING THEN ALLEGEDLY STOPPED. IT WAS FURTHER REPORTED THAT ALLEGEDLY THERE WAS A POP SOUND HEARD AND THE PHYSICIAN UNDERSTOOD THE STRING PROBABLY SNAPPED. REPORTEDLY, THE STENT ALLEGEDLY DID NOT DEPLOY AT ALL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20627164 ↗

Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices

All System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.