COVERA AVFM10100Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 10 mm x 100 mm (80 cm delivery catheter) (flared) · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741106613), received between Oct 16, 2020 and Jul 19, 2026 — classified as 5 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Material Deformation3
- Misfire2
- Break1
- Deformation Due to Compressive Stress1
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Restenosis2
- Cramp(s) /Muscle Spasm(s)1
- Device Overstimulation of Tissue1
- Reduced Blood Flow1
- Stenosis1
- Thrombosis/Thrombus1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2026Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions; Reduced Blood Flow
MAUDE report 25867847 ↗MalfunctionMay 5, 2025Misfire; Material Deformation; Misfire; Material Deformation
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE 10X100 FLARED COVERA NOT DEPLOY. IT WAS FURTHER REPORTED THAT PART OF THE STENT STARTED TO DEPLOY BUT THEN NEVER COMPLETELY DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21961865 ↗InjuryApr 8, 2025Material Deformation; Material Deformation
A PATIENT UNDERWENT A STENT GRAFT PLACEMENT PROCEDURE USING THE COVERA VASCULAR COVERED STENT. DURING THE PROCEDURE, THE STENT BEING USED FOLDED ON ITSELF. ANOTHER COVERA WAS NEEDED TO OPEN UP THE CRUSHED COVERA. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombosis/Thrombus
MAUDE report 21794479 ↗MalfunctionJun 2, 2023Break; Misfire; Break; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN LEFT BASILIC VEIN VIA LEFT UPPER ARM, WHEN STARTED TO DEPLOY THE STENT USING THE LARGE PINWHEEL, THE PINWHEEL BECAME STUCK AND THE STENT ALLEGEDLY STOPPED DEPLOYMENT ABOUT HALF WAY. IT WAS FURTHER REPORTED THAT THE STENT WAS ALLEGEDLY PARTIALLY DEPLOYED OF APPROXIMATELY 10-20MM. REPORTEDLY, THE STENT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17048696 ↗InjuryOct 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY ONE YEAR ELEVEN MONTHS AND TWENTY-TWO DAYS POST INDEX PROCEDURE, THE SUBJECT DEVELOPED 70% OF TARGET LESION STENOSIS. STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION. THE CURRENT PATIENT STATUS IS NOT PROVIDED.
Patient problems: Restenosis
MAUDE report 15568639 ↗MalfunctionJun 16, 2022Malposition of Device; Malposition of Device
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE, THE DEVICE ALLEGEDLY DEPLOYED INCORRECTLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14710642 ↗InjurySep 9, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY FIVE MONTHS POST INDEX PROCEDURE, THE PATIENT DEVELOPED 80% OF TARGET LESION STENOSIS AND COMPRESSION AT THE APEX OF THE STENT. A STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS USED TO SUCCESSFULLY TREAT THE TARGET LESION AND THE STENT COMPRESSION WAS RECOVERED. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Stenosis
MAUDE report 12445330 ↗InjuryJun 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, POST DEVICE IMPLANTATION SPASM OF OUTFLOW VEIN WAS REPORTED. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Cramp(s) /Muscle Spasm(s)
MAUDE report 11997695 ↗InjuryOct 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT DURING POST-DILATION OF THE VASCULAR COVERED STENT, SPASM OF THE OUTFLOW VEIN OCCURRED AND WAS RESOLVED WITH PTA. THE EVENT WAS CONSIDERED RESOLVED AND THE PATIENT WAS DISCHARGED THE SAME DAY. THE CURRENT PATIENT STATUS WAS NOT PROVIDED.
Patient problems: Device Overstimulation of Tissue; Restenosis
MAUDE report 10691973 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.