COVERA AVSL08080Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · COVERA™ Vascular Covered Stent 8 mm x 80 mm (120 cm delivery catheter) (straight) · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741106750), received between Oct 5, 2023 and Oct 5, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 22, 2022Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
Z-1108-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Loss of or Failure to Bond1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 5, 2023Loss of or Failure to Bond; Positioning Failure; Loss of or Failure to Bond; Positioning Failure
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE CEPHALIC ARCH, THE STENT WAS ALLEGEDLY FAILED TO DEPLOY. IT WAS FURTHER REPORTED THAT THE STENT WAS REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17876723 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.