Phasix 1192040Recalls & reported problems

Davol Inc. · Phasix Mesh, 8" x 16" (20 cm x 40 cm), Rectangle · Class 2 · Implant

All models of Phasix (43)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741107634), received between Apr 26, 2021 and Apr 10, 2026 — classified as 9 injuries, 1 malfunction.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient-Device Incompatibility10

Patient problems

  • Seroma5
  • Unspecified Infection3
  • Wound Dehiscence2
  • Abscess1
  • Drug Resistant Bacterial Infection1
  • Hernia1
  • Muscle/Tendon Damage1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 10, 2026Patient-Device Incompatibility

    AS REPORTED PER CLINICAL TRIAL (B)(6): THE SUBJECT PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2026 AND UNDERWENT AN OPEN PRIMARY LAPAROTOMY, COLECTOMY, EXPLORATORY LAPAROTOMY AND SMALL BOWEL RESECTION DURING WHICH A PHASIX FLAT MESH ONLAY WAS PLACED AND SECURED IN PLACE WITH ABSORBABLE MONOFILAMENT SUTURES WITH 2-0 PDS. THE SURGERY WAS DONE WITH TRIMMING THE PHASIX MESH (39 CM LENGTH AND 6 CM WIDTH). THE MIDLINE FASCIA WAS COMPLETELY CLOSED WITH ETHICON SLOW ABSORBING MONOFILAMENT WITH RUNNING STITCHES APPROXIMATELY 1 CM FAR APART AND 30 CM IN LENGTH. PATIENT WAS DISCHARGED FROM HOSPITAL ON (B)(6) 2026. ON (B)(6) 2026, THE SUBJECT PATIENT DEVELOPED A SMALL SEROMA ON INCISION SITE, NO TREATMENT OR TEST DONE. THE SEROMA WAS RESOLVED ON (B)(6) 2026. THE REPORTED ADVERSE EVENT (SEROMA) HAS BEEN ASSESSED PER CLINICIANS AS POSSIBLE RELATED TO THE STUDY DEVICE AND A PROBABLE RELATIONSHIP TO THE PROCEDURE. IT HAS BEEN ASSESSED AS MILD IN SEVERITY AND RECOVERED AND RESOLVED; THE REPORTED AE DOES NOT MEET THE DEFINITION OF A SAE (SERIOUS ADVERSE EVENT) AND THE DEFINITION OF USADE (UNANTICIPATED SERIOUS ADVERSE DEVICE EVENT).

    Patient problems: Seroma

    MAUDE report 24839738 ↗
  • InjuryFeb 18, 2026Patient-Device Incompatibility

    AS REPORTED PER CLINICAL TRIAL (B)(4): THE SUBJECT PATIENT UNDERWENT AN OPEN EXPLORATORY PRIMARY LAPAROTOMY WITH CONCOMITANT PROCEDURES HYSTERECTOMY, OOPHORECTOMY, LEFT SALPINGECTOMY AND RIGHT SALPINGO-OOPHORECTOMY ON (B)(6) 2023, DURING WHICH A BARD/DAVOL PHASIX MESH WAS TRIMMED, PLACED IN ONLAY FASHION AND SECURED IN PLACE. THE MIDLINE FASCIA WAS COMPLETELY CLOSED WITH SLOW ABSORBING MONOFILAMENT 1 PDS SUTURES. A 3CM OVERLAP IN ALL DIRECTIONS WAS ACHIEVED. SKIN CLOSURE WAS ACHIEVED BY RUNNING STITCH WITH LONG TERM ABSORBABLE SUTURES. THE SUBJECT PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2023. ON (B)(6) 2023, THE SUBJECT PATIENT WAS DIAGNOSED WITH SUPERFICIAL SURGICAL SITE INFECTION AND REQUIRED MEDICATION. (NOTE THE SSI RESOLVED ON 01-FEB-2024 AND WAS REPORTED TO THE FDA REFERENCE (B)(4). ON (B)(6) 2025, THE SUBJECT PATIENT WAS DIAGNOSED WITH A HERNIA OCCURRENCE THROUGH CT SCAN. AS REPORTED, THE ADVERSE EVENT (HERNIA) HAS BEEN ASSESSED PER CLINICIAN AS POSSIBLY RELATED TO THE STUDY DEVICE WITH A CAUSAL RELATIONSHIP TO THE PROCEDURE AND NOT RECOVERED/ NOT RESOLVED. THE SEVERITY HAS BEEN ASSESSED AS MILD. THE REPORTED AE DOES NOT MEET THE DEFINITION OF A SAE (SERIOUS ADVER

    Patient problems: Hernia

    MAUDE report 24376757 ↗
  • InjuryJul 3, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER CLINICAL TRIAL DVL-HE-018: THE SUBJECT PATIENT UNDERWENT LAPAROSCOPIC PRIMARY COLECTOMY AND EXPLORATORY LAPAROTOMY ON (B)(6) 2025 DURING WHICH A PHASIX MESH WAS TRIMMED, PLACED IN AN ONLAY FASHION AND SECURED USING ABSORBABLE MULTIFILAMENT SUTURES. A 3 CM OVERLAP IN ALL DIRECTIONS WAS ACHIEVED. PATIENT WAS DISCHARGED FROM HOSPITAL ON (B)(6) 2025. ON (B)(6) 2025, PATIENT WITH BULGE IDENTIFIED AT POST-OP VISIT WITH NP. SEROMA WAS DIAGNOSED VIA CT SCAN. THE REPORTED ADVERSE EVENT (SEROMA) HAS BEEN ASSESSED PER CLINICIANS AS CAUSALLY RELATED TO THE STUDY DEVICE AND TO THE PROCEDURE. IT HAS BEEN ASSESSED AS MODERATE IN SEVERITY, AND IT HAS RECOVERED/RESOLVED ON (B)(6) 2025. THE REPORTED AE DOES NOT MEET THE DEFINITION OF A SAE (SERIOUS ADVERSE EVENT) AND THE DEFINITION OF UADE (UNANTICIPATED SERIOUS ADVERSE DEVICE EVENT).

    Patient problems: Seroma

    MAUDE report 22393653 ↗
  • InjuryFeb 22, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER CLINICAL TRIAL (B)(4) : ON (B)(6) 2024, THE SUBJECT PATIENT UNDERWENT AN OPEN EXPLORATORY PRIMARY LAPAROTOMY WITH CONCOMITANT PROCEDURE LYSIS OF ADHESIONS WITH THE IMPLANT OF A PHASIX MESH. THE MESH WAS TRIMMED, PLACED IN ONLAY FASHION, AND FIXATED BY MECHANICAL FIXATION WITH 15 FIXATION POINTS. THE MIDLINE FASCIA WAS COMPLETELY CLOSED WITH SLOW ABSORBING MONOFILAMENT 0 PDS SUTURES. A 3CM OVERLAP IN ALL DIRECTIONS WAS ACHIEVED. SKIN CLOSURE WAS ACHIEVED BY STAPLES. THE SUBJECT PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2024. ON (B)(6) 2024, THE PATIENT VISITED THE HOSPITAL AND WAS DIAGNOSED WITH SURGICAL SITE INFECTION THROUGH CLINICAL ASSESSMENT. AS REPORTED, THE MANUAL OPENING OF WOUND WAS PERFORMED AT BEDSIDE. THE FLUID WAS EXTRACTED, AND THE DRAINAGE CULTURES WERE COLLECTED. THE PATIENT WAS ADMINISTERED WITH ANTIBIOTICS/ANTIMICROBIALS INTRAVENOUSLY. PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2024. THE REPORTED ADVERSE EVENT (SURGICAL SITE INFECTION) HAS BEEN ASSESSED PER CLINICIAN AS DEFINITELY RELATED TO THE STUDY DEVICE AND POSSIBLY RELATED TO THE PROCEDURE. THE AE HAS BEEN ASSESSED AS MODERATE IN SEVERITY. THE REPORTED AE DOES MEET THE DEFINIT

    Patient problems: Unspecified Infection

    MAUDE report 18757206 ↗
  • InjuryJan 31, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER THE CLINICAL TRIAL (B)(6): THE SUBJECT PATIENT UNDERWENT AN OPEN EXPLORATORY PRIMARY LAPAROTOMY WITH CONCOMITANT PROCEDURES HYSTERECTOMY, OOPHORECTOMY, LEFT SALPINGECTOMY AND RIGHT SALPINGO-OOPHORECTOMY ON (B)(6) 2023, DURING WHICH A BARD/DAVOL PHASIX MESH WAS TRIMMED, PLACED IN ONLAY FASHION AND SECURED IN PLACE. THE MIDLINE FASCIA WAS COMPLETELY CLOSED WITH SLOW ABSORBING MONOFILAMENT 1 PDS SUTURES. A 3CM OVERLAP IN ALL DIRECTIONS WAS ACHIEVED. SKIN CLOSURE WAS ACHIEVED BY RUNNING STITCH WITH LONG TERM ABSORBABLE SUTURES. THE SUBJECT PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2023. ON (B)(6) 2023, THE SUBJECT PATIENT WAS DIAGNOSED WITH SUPERFICIAL SURGICAL SITE INFECTION AND REQUIRED MEDICATION. THE REPORTED ADVERSE EVENT (SURGICAL SITE INFECTION) HAS BEEN ASSESSED PER CLINICIAN AS POSSIBLY RELATED TO THE STUDY DEVICE, DEFINITELY RELATED TO THE PROCEDURE AND RECOVERING/RESOLVING. THE AE HAS BEEN ASSESSED AS MILD IN SEVERITY. THE REPORTED AE DOES NOT MEET THE DEFINITION OF A SAE (SERIOUS ADVERSE EVENT) AND UADE (UNANTICIPATED ADVERSE DEVICE EVENT). ADDENDUM PER ADDITIONAL INFORMATION RECEIVED: THE SUBJECT PATIENT WAS DIAGNOSED WITH SUPERFICIAL SURGICAL SITE

    Patient problems: Unspecified Infection

    MAUDE report 18612599 ↗
  • InjuryJun 7, 2023Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER THE CLINICAL TRIAL (B)(4): ON (B)(6) 2023, THE SUBJECT PATIENT UNDERWENT AN OPEN EXPLORATORY RELAPAROTOMY-COLOSTOMY REVERSAL PROCEDURE DURING WHICH A PHASIX MESH WAS TRIMMED AND IMPLANTED IN ONLAY FASHION USING ABSORBABLE MONOFILAMENT 2-0 PDS SUTURES. A 3CM OVERLAP IN ALL DIRECTIONS WAS ACHIEVED. THE SUBJECT PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2023. ON (B)(6) 2023, THE PATIENT WAS DIAGNOSED WITH SURGICAL SITE INFECTION WHICH REQUIRED HOSPITALIZATION. THIS ADVERSE EVENT HAS BEEN ASSESSED PER CLINICIAN AS POSSIBLY RELATED TO THE STUDY DEVICE AND DEFINITELY RELATED TO THE PROCEDURE. THE ADVERSE EVENT HAS BEEN ASSESSED AS RECOVERING/RESOLVING, MODERATE IN SEVERITY AND A SAE (SERIOUS ADVERSE EVENT) BUT DOES NOT MEET THE DEFINITION OF UADE (UNANTICIPATED ADVERSE DEVICE EVENT).

    Patient problems: Unspecified Infection

    MAUDE report 17078415 ↗
  • InjuryJan 19, 2023Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER PREVENT CLINICAL TRIAL DVL-HE-018: ON (B)(6) 2022, THE SUBJECT PATIENT UNDERWENT PRIMARY LAPAROTOMY-OPEN AORTOILIAC BYPASS DURING WHICH A PHASIX MESH WAS TRIMMED AND SECURED IN PLACE WITH 2-0 PDS SUTURES. PROPHYLACTIC ANTIBIOTIC CEFAZOLIN WAS USED PERIOPERATIVELY. ON (B)(6) 2022, THE PATIENT WAS DIAGNOSED WITH SEROMA AND WAS ADMITTED ON THE SAME DAY. IT WAS REPORTED THAT THE PATIENT REQUIRED IR PLACED DRAIN. FLUID CULTURES OF SEROMA WERE NEGATIVE. DURING THIS STAY THE PATIENT WAS DIAGNOSED WITH WOUND DEHISCENCE. THE ADVERSE EVENT OF SEROMA WAS ASSESSED PER CLINICIAN AS MODERATE SEVERITY, POSSIBLY RELATED TO THE STUDY DEVICE, AND DEFINITELY RELATED TO THE PROCEDURE AND RECOVERING/RESOLVING. THE WOUND DEHISCENCE WAS ASSESSED PER CLINICIAN AS MODERATE SEVERITY MODERATE SEVERITY AND NOT RECOVERED/NOT RESOLVED. THE ADVERSE EVENT HAS BEEN ASSESSED AS A SERIOUS ADVERSE EVENT (SAE) AND THE EVENT DOES NOT MEET THE DEFINITION OF AN UNANTICIPATED ADVERSE DEVICE EFFECT (UADE). ADDENDUM PER ADDITIONAL INFORMATION RECEIVED: ON (B)(6) 2023, THE PATIENT WAS DIAGNOSED WITH ABDOMINAL WALL ABSCESS AND WAS ADMITTED ON THE SAME DAY. THE ABSCESS CULTURE TESTED POSITIVE FOR A DRUG RESISTA

    Patient problems: Abscess; Drug Resistant Bacterial Infection; Seroma; Wound Dehiscence

    MAUDE report 16210134 ↗
  • MalfunctionApr 19, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED PER PREVENT CLINICAL TRIAL (B)(4): ON (B)(6) 2021, THE PATIENT UNDERWENT AN OPEN HEPATICOJEJUNOSTOMY AND CHOLECYSTECTOMY PROCEDURE AND A BARD/DAVOL PHASIX MESH WAS TRIMMED AND PLACED IN ONLAY FASHION WITH 3 CM OVERLAP IN ALL DIRECTIONS. THE MESH WAS SUTURED IN PLACE WITH NON-BARD/DAVOL ABSORBABLE MULTIFILAMENT SUTURES. AS REPORTED, THE PATIENT WAS DIAGNOSED WITH DIASTASIS RECTION (B)(6) 2021. THE AE IS REPORTED TO HAVE NOT RESOLVED, AND NO TREATMENT HAS BEEN PROVIDED. THE ADVERSE EVENT HAS BEEN ASSESSED AS MILD SEVERITY AND AS NOT A SERIOUS ADVERSE EVENT (SAE) OR AN UNANTICIPATED ADVERSE DEVICE EFFECT, (UADE) BY THE STUDY GROUP. THE ADVERSE EVENT HAS BEEN ASSESSED PER STUDY CLINICIAN AS NOT RELATED TO THE STUDY DEVICE AND NOT RELATED TO THE INDEX PROCEDURE. THE ADVERSE EVENT HAS BEEN ASSESSED PER CEC CLINICIAN AS POSSIBLY RELATED TO THE STUDY DEVICE AND POSSIBLY RELATED TO THE INDEX PROCEDURE.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 14140676 ↗
  • InjuryAug 4, 2021Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED TO DAVOL THAT A PATIENT WHO IS PART OF A CLINICAL STUDY EXPERIENCED SEROMA POST-IMPLANT OF PHASIX MESH. (B)(6) 2021 - THE PATIENT UNDERWENT AN OPEN HEPATICOJEJUNOSTOMY. A BARD/DAVOL PHASIX MESH 20X40CM, WAS TRIMMED AND PLACED IN ONLAY FASHION. THE MESH WAS TRIMMED AND 3CM OVERLAP WAS ACHIEVED IN ALL DIRECTIONS. THE PHASIX MESH WAS SECURED IN PLACE WITH SUTURE FIXATION ONLY AND THE MIDLINE FASCIA WAS CLOSED BY RUNNING STITCH USING NON BARD/DAVOL 2-0 PDS SUTURE. THE SKIN WAS FULLY CLOSED WITH SUTURE AND SURGICAL GLUE. A DRAIN WAS PLACED DURING THE INDEX PROCEDURE, INTRAPERITONEALLY IN THE RIGHT UPPER QUADRANT OF THE ABDOMEN. (B)(6) 2021 - THE DRAIN WAS REMOVED. (B)(6) 2021 - THE SUBJECT PATIENT WAS DIAGNOSED WITH A SEROMA. (B)(6) 2021 - THE AE HAS RESOLVED AND THE SUBJECT PATIENT HAS RECOVERED. THE ADVERSE EVENT HAS BEEN ASSESSED PER CLINICIAN AS DEFINITELY RELATED TO THE STUDY DEVICE AND NOT RELATED TO THE INDEX PROCEDURE. NO ADDITIONAL SURGICAL INTERVENTION HAS BEEN PROVIDED AT THIS TIME. THE ADVERSE EVENT HAS BEEN ASSESSED AS MILD SEVERITY AND NOT A SERIOUS ADVERSE EVENT (SAE) OR A (UNANTICIPATED ADVERSE DEVICE EFFECT (UADE).

    Patient problems: Seroma

    MAUDE report 12273469 ↗
  • InjuryApr 26, 2021Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED TO DAVOL THAT A PATIENT WHO IS PART OF A CLINICAL STUDY EXPERIENCED SEROMA AND SMALL SKIN DEHISCENCE POST-IMPLANT OF PHASIX MESH. (B)(6) 2021 - THE SUBJECT PATIENT UNDERWENT AN OPEN EXPLORATORY LAPAROTOMY WITH A BOWEL RESECTION, LYSIS OF ADHESIONS AND OSTOMY REVISION. A BARD/DAVOL PHASIX MESH WAS TRIMMED AND PLACED IN ONLAY FASHION. MECHANICAL FIXATION (NON-BARD/DAVOL) WAS USED TO SECURE THE MESH. THE MESH WAS TRIMMED AND ADEQUATE OVERLAP OF THE DEFECT WAS ACHIEVED. FASCIAL CLOSURE WAS ACHIEVED WITH NON-BARD/DAVOL SUTURE, IN A RUNNING STITCH TECHNIQUE. IT WAS REPORTED THAT PROPHYLACTIC ANTIBIOTIC (CEFOTETAN) WAS ADMINISTERED TO THE PATIENT PERI-OPERATIVELY. (B)(6) 2021 - THE PATIENT WAS DISCHARGED HOME. (B)(6) 2021 - THE SUBJECT PATIENT WAS DIAGNOSED WITH AN APPROXIMATELY 2CM SEROMA CAUSING SMALL SKIN DEHISCENCE. AS REPORTED, THE PATIENT HAS BEEN INSTRUCTED TO RETURN TO CLINIC IN 2 WEEKS AND TO PACK THE WOUND TWICE A DAY WITH DRY GAUZE, AND SWITCH TO SALINE GAUZE ONCE DRAINAGE DECREASES. THE PATIENT HAD ALSO BEEN INSTRUCTED TO SWITCH TO AQUACEL ROPE IN A WEEK. NO ADDITIONAL SURGICAL INTERVENTION HAS BEEN PROVIDED AT THIS TIME. AS REPORTED, THE REPORTED ADVERSE EVENT

    Patient problems: Seroma; Wound Dehiscence

    MAUDE report 11725455 ↗

FDA clearance and classification

The FDA regulates product code OOD (surgical film) under 21 CFR 878.3300 as a Class II device. 21 CFR 878.3300 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: To be implanted to reinforce soft tissue or bone where weakness exists. Intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, gynecological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera.

Common questions about the Phasix 1192040

Has the Phasix 1192040 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00801741107634), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Phasix 1192040?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Apr 26, 2021 and Apr 10, 2026. In those reports, the most common specific device problem is “Patient-Device Incompatibility” (10); the most common patient problems are “Seroma” (5), “Unspecified Infection” (3) and “Wound Dehiscence” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Phasix 1192040?

None of the reports are classified as deaths, 9 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Phasix 1192040 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Phasix 1192040, and is it still sold?

The FDA's device database lists Davol Inc. as the company responsible for it (brand Phasix, model 1192040). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other surgical film devices

All surgical film devices →