Midline Catheter Kit P4154108DRecalls & reported problems
Bard Access Systems, Inc. · 4Fr SL PowerMidline catheter Max Kit · Class 2
All models of Midline Catheter Kit (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard PowerMidline catheter kit is a kit used to place a midline catheter in a vein. Many of the 126 reports describe leaks from the line, fraying material, and introducers that were hard to separate, and some describe breaks. The reports include 1 death and 9 injuries. There is 1 open recall, from May 2026, over lidocaine ampoules made under deficient manufacturing practices.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 126
- Deaths reported
- 1
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 00801741108761), received between Oct 4, 2016 and Aug 12, 2026 — classified as 1 death, 9 injuries, 116 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (28 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2662-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak28
- Material Frayed20
- Difficult or Delayed Separation13
- Material Deformation12
- Break8
- Defective Component8
- Leak/Splash6
- Activation Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions61
- No Consequences Or Impact To Patient24
- No Known Impact Or Consequence To Patient21
- Foreign Body In Patient6
- Pain4
- Laceration(s)2
- Swelling2
- Anaphylactoid / Anaphylaxis1
Recent adverse event reports
The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2026Suction Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50220753 ↗MalfunctionJul 8, 2026Material Deformation
IT WAS REPORTED DURING A TRAINING A SKIN NICK WAS PERFORMED, FOLLOWED BY DIFFICULTY ADVANCING THE DILATOR. WHEN THE DILATOR WAS REMOVED AND INSPECTED, NURSES NOTED THAT IT HAD BUCKLED. EVENT INVOLVED THE 4F DILATOR. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25749512 ↗MalfunctionJul 1, 2026Material Deformation
IT WAS REPORTED MICROINTRODUCER WAS OBSERVED TO HAVE A DAMAGED SHEATH TIP. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25679126 ↗MalfunctionApr 13, 2026Fluid/Blood Leak
IT WAS REPORTED BY A CUSTOMER THAT THEY ARE EXPERIENCING LEAKING FROM THE POWER MIDLINE. PROCEDURE TOOK PLACE (B)(6) 2026. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865239 ↗MalfunctionApr 1, 2026Fluid/Blood Leak
IT WAS REPORTED CUSTOMER IS EXPERIENCING LEAKING FROM THE POWER MIDLINE. LINE WAS REMOVED. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24753725 ↗MalfunctionMar 25, 2026Difficult or Delayed Separation
IT WAS REPORTED THAT DURING PLACEMENT OF A 4FR ML CATHETER, THE INTRODUCER MALFUNCTIONED. WHEN BREAKING THE INTRODUCER, ONLY THE LEFT SIDE SEPARATED PROPERLY WHILE THE RIGHT SIDE REMAINED INTACT, LEAVING A TIGHT RING AROUND THE CATHETER. THE CLINICIAN HAD TO CUT THE RING WITH A SCALPEL TO FREE THE CATHETER, THOUGH NO CATHETER DAMAGE OR LEAKAGE WAS OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24687824 ↗MalfunctionMar 6, 2026Material Frayed
IT WAS REPORTED THE GUIDEWIRE IN MST MIDLINE KIT UNRAVELING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532071 ↗MalfunctionFeb 24, 2026Difficult or Delayed Separation
IT WAS REPORTED DURING PLACEMENT OF A MIDLINE THE INTRODUCER DID NOT SPLIT CORRECTLY. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24429699 ↗MalfunctionJan 20, 2026Unsealed Device Packaging
IT WAS REPORTED BY CUSTOMER THAT THE KIT WAS OPEN BEFORE ANYONE PHYSICALLY OPENED IT. IT WAS REPORTED THIS OCCURRED WITH 2 DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24118130 ↗MalfunctionAug 6, 2025Defective Component; Defective Component
IT WAS REPORTED, 12 STYLETS TO HAVE MULTIPLE MICRO RAZED AREAS THIS REPORT ADDRESSES ALL TWELVE DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22734407 ↗MalfunctionAug 4, 2025Activation Problem; Activation Problem
ADDITIONAL INFORMATION: THERE WAS NO NEEDLE STICK INJURY, IT JUST SCRATCHED THE GLOVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22701813 ↗MalfunctionAug 4, 2025Activation Problem; Activation Problem
IT WAS REPORTED THAT NEEDLE FAILED TO SAFETY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22700834 ↗MalfunctionJul 31, 2025Difficult or Delayed Separation; Difficult or Delayed Separation
IT WAS REPORTED THAT INTRODUCER FAILED TO SEPARATE PROPERLY AND HAD TO BE CUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22680829 ↗MalfunctionJun 20, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THE STYLET IN THE PICC WOULD NOT UNSCREW SO THEY COULD TAKE IT OUT. THE EVENT RESULTED IN NEEDING TO OVER THE WIRE EXCHANGE FOR A NEW MIDLINE FOR THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22298646 ↗MalfunctionApr 3, 2025Activation Problem; Activation Problem
ADDITIONAL INFORMATION: THERE WAS NO ADVERSE EVENT OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21764040 ↗MalfunctionApr 3, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT THE WIRE FRAYED WHEN THREADING INTO THE NEEDLE. NO HARM OR INJURY OCCURRED DURING THIS INCIDENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21762306 ↗MalfunctionMar 31, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY THE CUSTOMER WHEN TRYING TO ACTIVATE SAFETY, THE SAFETY SLID OFF THE NEEDLE AND DETACHED. THERE WAS NO HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21733156 ↗MalfunctionMar 5, 2025Break; Break
IT WAS REPORTED BY CUSTOMER THAT THE POWER-MIDLINE CATHETER SEVERED AT THE EXTENSION SET. ADDITIONAL INFORMATION RECEIVED 2/24/2025: IT WAS REPORTED RN ENTERED THE PATIENT'S ROOM AND DISCOVERED THE MIDLINE APPEARED TO BE SEVERED BETWEEN THE CATHETER AND CLAMP. NO HARM TO PATIENT, PATIENT REQUIRED NEW IV PLACEMENT FOR CONTINUED MEDICATION. MIDLINE CLOTTED UPON SEVERING OF LINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21528632 ↗MalfunctionDec 4, 2024Activation Problem; Activation Problem
IT WAS REPORTED A NEEDLE MALFUNCTION, AS I WAS PULLING THE GREEN SAFETY MECHANISM DOWN IT JUST CAME RIGHT OFF. I DIDN¿T PULL IT HARD. IT JUST SLIPPED RIGHT OFF THE END. I DIDN¿T GET POKED LUCKILY. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20848237 ↗MalfunctionNov 7, 2024Material Frayed; Material Frayed
IT WAS REPORTED BY THE CUSTOMER THAT THIS PARTICULAR MIDLINE KIT CANNOT PULL THE STIFFENING STYLET AND WHEN SHE TRIED TO PULL HARDLY AGAINST THE T LOCK, THE STYLET STARTED TO STRETCH AND SHRED. DECIDED TO PLACE THE CATHETER WITHOUT THE STYLET. NO OTHER INFORMATION PROVIDED AT THIS TIME. ADDITIONAL INFORMATION PROVIDED 10/28/2024: THE RESULT OF THE MALFUNCTION DID DIRECTLY AFFECT THE PATIENT AS THE NEEDLE AND GUIDEWIRE WERE IN THE PATIENT WHEN IT BECAME STUCK, AND THEY COULD NOT REMOVE OR THREAD AFTER THE VEIN WAS CANNULATED. THE GUIDEWIRE WAS PLACED WITH EASE AND NO OBVIOUS ABNORMALITIES NOTED. AFTER THE INTRODUCER WAS INSERTED, THE RN WAS NOT ABLE TO REMOVE GUIDEWIRE. MULTIPLE ATTEMPTS WERE MADE AND AFTER SEVERAL MINUTES SHE WAS ABLE TO REMOVE THE GUIDEWIRE AND IT WAS NOTED TO BE BENT AND FRAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20631503 ↗MalfunctionSep 19, 2024Material Frayed; Material Frayed
IT WAS REPORTED BY CUSTOMER THAT PUT IN A TRIMMABLE MIDLINE AND PULLED THE WIRE AND IT WAS SPLIT IN 2 PIECES. NO INJURY TO THE PATIENT OR HEALTHCARE PROFESSIONAL. PATIENT RECEIVED CHEST XRAY TO CONFIRM THERE WERE NO WIRE FRAGMENTS LEFT IN ARM. UPON REMOVAL WIRE WAS SPLIT IN 2 SEPARATE THREADS. LIKE THE WIRE SEPARATED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20263557 ↗InjurySep 12, 2024Break; Break
IT WAS REPORTED, MIDLINE CAME APART IN A PATIENT'S ARM. IT WAS A 20CM PROVENA MIDLINE, LOT NUMBER REJQ1293. UNEVENTFUL INSERTION ON (B)(6) 2024- PLACED IN LEFT BASILIC VEIN IN ONE ATTEMPT WITH GREAT BLOOD RETURN AND EASY FLUSH AFTER INSERTION. WAS USED FOR INTERMITTENT INFUSIONS THROUGHOUT THE NEXT 24 HOURS. NURSING TRIED TO FLUSH ABOUT 24 HOURS AFTER INSERTION AND COULD NOT FLUSH. IV TEAM WAS CONSULTED AND WE COULD NOT FLUSH IT EITHER. NURSE TOOK DOWN THE DRESSING AND PULLED IT BACK A FEW CENTIMETERS (IT WAS 20CM LONG) AND STILL COULD NOT GET IT TO FLUSH SO DECISION WAS MADE TO PULL LINE. LINE WAS REMOVED WITHOUT ANY RESISTANCE, BUT ONLY 7CM UPON REMOVAL. IMMEDIATELY PLACED A TOURNIQUET, GOT AN X-RAY AND ULTRASOUND (WHICH DEMONSTRATED THAT THE RETAINED MIDLINE WAS IN THE AXILLARY VEIN). IR HAD TO RETRIEVE RETAINED MIDLINE THAT NIGHT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 20205910 ↗MalfunctionAug 8, 2024Material Frayed; Defective Component; Material Frayed; Defective Component
IT WAS REPORTED VIA MEDWATCH, POWER MIDLINE CATHETER HAD A FRAYED GUIDEWIRE. THIS WAS NOTICED BEFORE THE LINE WAS PLACED IN THE PATIENT. PROCEDURE WAS PAUSED WHILE A COLLEAGUE OBTAINED A NEW MIDLINE KIT. WHEN NEW KIT ARRIVED, NEW LINE WAS INSPECTED AND FOUND TO BE WITHIN NORMAL LIMITS. PROCEDURE COMPLETED WITH NO FURTHER ISSUES. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19939071 ↗MalfunctionJul 1, 2024Difficult or Delayed Separation; Difficult or Delayed Separation
IT WAS REPORTED BY MEDWATCH PATIENT REQUIRED MIDLINE. DURING MIDLINE INSERTION, PEEL-AWAY INTRODUCER FAILED TO SEPARATE PROPERLY CAUSING CATHETER TO WITHDRAW. REQUIRED INTRODUCER TO BE CUT TO BE REMOVED. NO PATIENT HARM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19644965 ↗MalfunctionJun 28, 2024Difficult or Delayed Separation; Difficult or Delayed Separation
IT WAS REPORTED VIA MEDWATCH PATIENT REQUIRED MIDLINE FOR IV VANCOMYCIN. DURING MIDLINE INSERTION, PEEL-AWAY INTRODUCER FAILED TO SEPARATE PROPERLY - CAUSING CATHETER TO WITHDRAW. REQUIRED INTRODUCER TO BE CUT TO BE REMOVED. NO PATIENT HARM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19639481 ↗
Other Midline catheter devices
- ARROW IPN921959Teleflex Incorporated · Midline catheter2 recalls0 reports
- Midline Catheter Kit S4153108BDPBard Access Systems, Inc. · Midline catheter1 recall36 reports
- Midline Catheter Kit S4153108DBard Access Systems, Inc. · Midline catheter1 recall32 reports
- Midline Catheter Kit S4153108DGPBard Access Systems, Inc. · Midline catheter1 recall15 reports
- Midline Catheter Kit S4254108BDPBard Access Systems, Inc. · Midline catheter1 recall11 reports
- Midline Catheter Kit P4154108Bard Access Systems, Inc. · Midline catheter1 recall10 reports
- ARROW IPN037698Teleflex Incorporated · Midline catheter1 recall9 reports
- Midline Catheter Kit S4254108DBard Access Systems, Inc. · Midline catheter1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.