Midline Catheter Kit S4153108DGPRecalls & reported problems
Bard Access Systems, Inc. · 3Fr SL Provena Midline Max Barrier Kit w/GuardIVa and Probe Cover · Class 2
All models of Midline Catheter Kit (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00801741154027), received between Mar 22, 2019 and Mar 30, 2026 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2667-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation4
- Activation Problem3
- Defective Component2
- Fluid/Blood Leak2
- Difficult or Delayed Separation1
- Difficult to Remove1
- Leak/Splash1
- Suction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient2
- Pain2
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
- Thrombosis/Thrombus1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Material Deformation
IT WAS REPORTED THAT 3 FR MIDLINE FAILED WITH THE DILATOR AND WAS SAVED. THE DILATOR PORTION BUNCHED UP ON THE END AND CUSTOMER HAD TO GRAB MST KIT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24712076 ↗MalfunctionJun 24, 2025Activation Problem; Activation Problem
THE CUSTOMER RETURNED AN ADDITIONAL 21G SECURELOC INTRODUCER NEEDLE FOR EVALUATION AND THE SAFETY MECHANISM IS NOT ENGAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22319172 ↗MalfunctionMar 14, 2025Activation Problem; Activation Problem
IT WAS REPORTED THAT THE NEEDLE DID NOT SAFETY ALL THE WAY. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21607511 ↗MalfunctionMar 7, 2025Activation Problem; Activation Problem
IT WAS REPORTED INTRODUCER NEEDLE DID NOT SAFETY AFTER BEING USED FOR MIDLINE INSERTION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21548875 ↗MalfunctionOct 19, 2024Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT HAS INTRODUCER, SHEATH WOULD NOT WORK. NO PATIENT HARM REPORTED. NO OTHER INFORMATION WAS PROVIDED. 09/25/2024 IT WAS FOUND DURING AN INVESTIGATION THE PEEL-APART SHEATH EXHIBITED DEFORMATION THAT WAS CONSISTENT WITH RESISTANCE ENCOUNTERED DURING ATTEMPTED INSERTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20489280 ↗InjuryJul 1, 2024Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
IT WAS REPORTED VIA MEDWATCH THAT ACCESS TO VEIN OBTAINED, INSERTED WIRE THROUGH CATHETER, PULLED CATHETER OUT. ADVANCED WIRE AND PATIENT COMPLAINED OF PAIN, PULLED WIRE BACK. HEALTHCARE PROVIDER WAS ABLE TO PULL WIRE BACK A FEW CENTIMETERS, AND THEN IT WAS STUCK WITH RESISTANCE FELT. PATIENT REQUIRED INTERVENTIONAL RADIOLOGY (IR) INTERVENTION FOR REMOVAL. MIDLINE WAS PLACED IN LEFT UPPER ARM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Pain
MAUDE report 19650977 ↗MalfunctionJun 11, 2024
ACCESS TO VEIN OBTAINED, INSERTED WIRE THROUGH CATHETER, PULLED CATHETER OUT. ADVANCED WIRE AND PATIENT COMPLAINED OF PAIN, PULLED WIRE BACK. HEALTHCARE PROVIDER WAS ABLE TO PULL WIRE BACK A FEW CENTIMETERS, AND THEN IT WAS STUCK WITH RESISTANCE FELT. PATIENT REQUIRED INTERVENTIONAL RADIOLOGY (IR) INTERVENTION FOR REMOVAL.
Patient problems: Pain
MAUDE report 19515269 ↗MalfunctionOct 11, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT LEAKING AT THE T-CONNECTOR ON THE PICC LINE THEY USED YESTERDAY. LEAKING FROM STOPPER NOTED WHILE THE PICC WAS BEING FLUSHED DURING THE INSERTION PROCESS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17916919 ↗MalfunctionOct 11, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT LEAKING AT THE T CONNECTOR ON THE PICC LINES THEY USED YESTERDAY. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17916075 ↗MalfunctionMar 28, 2023Suction Problem; Suction Problem
IT WAS REPORTED VIA MS&S: CALLER STATES SHE IS WORKING WITH A PATIENT THAT HAS POWERPICC S4153108DGP (LOT NUMBER REGY0370) AND CAN FLUSH INTO THE PICC EASILY BUT CANNOT GET A BLOOD RETURN. CALLER ASKING IF WE CAN PROVIDE CATHFLO TO HELP WITH THE OCCLUSION AND HOW TO ADMINISTER CATHFLO. MS&S RESPONSE: EXPLAINED THAT A FIBRIN TAIL CAN FORM NEAR THE TOP OF THE CATHETER. WHEN THE PICC IS FLUSHED WITH SALINE THE FIBRIN TAIL IS NOT AN ISSUE. WHEN ASPIRATED FOR A BLOOD RETURN THE FIBRIN TAIL CAN BE DRAWN INTO THE CATHETER TIP OCCLUDING THE LUMEN. CONFIRMED THE CATHFLO CAN BE USED TO HELP ALLEVIATE OCCLUSIONS CAUSED BY FIBRIN TAILS. EXPLAINED SHE WOULD NEED TO FOLLOW THE INSTRUCTIONS FOR CATHFLO USAGE. EXPLAINED THAT BD DOES NOT PROVIDE CATHFLO TO CUSTOMERS. SUGGESTED SHE CALL THE DOCTOR AND DISCUSS FURTHER.
Patient problems: Thrombosis/Thrombus
MAUDE report 16632675 ↗MalfunctionMar 5, 2021Material Deformation; Difficult or Delayed Separation; Material Deformation; Difficult or Delayed Separation
IT WAS REPORTED "WHEN GETTING READY TO DILATE, RN SEPARATED THE INTRODUCER TO MICRO-DILATE AS MANY CLINICIANS DO WITH THE 3FR PROVENA MIDLINES SINCE THE DILATE IS FLIMSY. THE DILATOR (INTERNAL LIGHTER GREY PORTION) WAS INSERTED INTO THE VEIN WITH EASE. THEN IT WAS REMOVED AND REASSEMBLED TO THE INTRODUCER/PEEL AWAY PORTION. WHILE TRYING TO ADVANCE EXTREME RESISTANCE WAS MET. RN REMOVED THE DILATOR TO FIND THAT AT THE PORTION WHERE THE DILATOR TRANSITIONS TO THE INTRODUCER, THE DARK GREY TIP OF THE INTRODUCER WAS ROLLING BACK ON ITSELF, WHICH WE SOMETIMES REFER TO AS ¿FISH MOUTH.¿ A 2ND BARD INTRODUCER WAS USED WITH NO ISSUE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11425357 ↗MalfunctionSep 3, 2020Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE SEAL ON THE KIT WAS OPENED. IT WAS NOT USED ON A PATIENT.
Patient problems: No Patient Involvement
MAUDE report 10491536 ↗MalfunctionApr 25, 2019Fluid/Blood Leak
IT WAS REPORTED LEAKING ON THE EXTERIOR TUBING AROUND THE HUB AND/OR AT THE HUB WAS NOTICED ON 2 CATHETERS IN THIS LOT. PICC WAS PLACED IN THE BASILIC VEIN AND NO PATIENT HARM WAS REPORTED. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8552278 ↗MalfunctionApr 25, 2019Fluid/Blood Leak
IT WAS REPORTED LEAKING ON THE EXTERIOR TUBING AROUND THE HUB AND/OR AT THE HUB WAS NOTICED ON 2 CATHETERS IN THIS LOT. PICC WAS PLACED IN THE BASILIC VEIN AND NO PATIENT HARM WAS REPORTED. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8552277 ↗MalfunctionMar 22, 2019Leak/Splash
IT WAS REPORTED THAT THE POWERMIDLINE WAS LEAKING AT THE SITE WHEN FLUSHING OR INFUSING. THE MIDLINE WAS TRIMMED AT 10CM. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8445299 ↗
Other Midline catheter devices
- Midline Catheter Kit P4154108DBard Access Systems, Inc. · Midline catheter1 recall1 death126 reports
- ARROW IPN921959Teleflex Incorporated · Midline catheter2 recalls0 reports
- Midline Catheter Kit S4153108BDPBard Access Systems, Inc. · Midline catheter1 recall36 reports
- Midline Catheter Kit S4153108DBard Access Systems, Inc. · Midline catheter1 recall32 reports
- Midline Catheter Kit S4254108BDPBard Access Systems, Inc. · Midline catheter1 recall11 reports
- Midline Catheter Kit P4154108Bard Access Systems, Inc. · Midline catheter1 recall10 reports
- ARROW IPN037698Teleflex Incorporated · Midline catheter1 recall9 reports
- Midline Catheter Kit S4254108DBard Access Systems, Inc. · Midline catheter1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page