Peripheral Catheter Insertion Kit F220081Recalls & reported problems
Bard Access Systems, Inc. · DOT PFlex, 20G 8cm Full w/o GIVa · Class 2
All models of Peripheral Catheter Insertion Kit (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741110672), received between Dec 16, 2016 and Aug 16, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack1
- Material Fragmentation1
- Physical Resistance/Sticking1
- Positioning Failure1
- Reset Problem1
- Retraction Problem1
- Self-Activation or Keying1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 16, 2018Crack; Retraction Problem; Self-Activation or Keying; Reset Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT MIDLINE WAS ATTEMPTED TO BE PLACED IN PATIENT'S LEFT UPPER ARM CEPHALIC VEIN. IT WAS STATED, "UPON ENTERING MIDLINE, GUIDEWIRE WAS ADVANCED AND STARTED TO POP UP SUPERIORLY TO INSERTION SITE UNDER THE SKIN. GUIDE WIRE WAS THEN RETRACTED TO ABORT PLACEMENT HOWEVER UPON RETRACTING A CRACK WAS HEARD. GUIDE WIRE WAS DIFFICULT TO RETRACT AT THIS POINT AS WELL AS CATHETER DIFFICULT TO RETRACT WITHOUT RESISTANCE BEING MET. UPON US GUIDANCE, NEEDLE AND CATHETER STILL IN TISSUE AND CATHETER IN VEIN. INTERVENTIONAL RADIOLOGY CALLED TO BEDSIDE. TO THEIR DISCRETION, NEEDLE TIP WAS NOT IN VEIN HOWEVER CATHETER WAS IN VEIN. IR MD WAS CALLED TO BEDSIDE AND TOOK OUT MIDLINE WHICH ONCE PULLED SHOWED NEEDLE HAD PUNCTURED THROUGH CATHETER AND GUIDE WIRE WAS TORTIOUS. TIP OF GUIDE WIRE ACTUALLY FELL APART FROM MAIN GUIDE WIRE WHILE OUT OF PATIENT'S ARM. PATIENT HAD XRAY OF ARM TO EVALUATE IF ANY OTHER FOREIGN BODY IN PLACE. XRAY CONFIRMED NO FOREIGN BODY. PT SUSTAINED SMALL HEMATOMA THAT RESOLVED ON ITS OWN".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7791636 ↗InjuryDec 16, 2016Positioning Failure; Material Fragmentation
IT WAS REPORTED BY BAS CLINICAL SPECIALIST THAT SHE WAS AT THE FACILITY TRAINING WHEN THE ALLEGED FAILURE OCCURRED. WHILE INSERTING THE POWERGLIDE INTO THE LEFT BRACHIAL VEIN, THE GUIDEWIRE WAS INSERTED, THE CATHETER WAS DEPLOYED, AND WHEN THE HOUSING UNIT WAS REMOVED IT WAS REPORTED THAT THE GUIDEWIRE WAS SHEARED. CLINICAL SPECIALIST STATED THAT THE NEEDLE RETRACTED INTO THE SAFETY DEVICE AND THE CATHETER WAS THEN REMOVED. IT WAS FOUND UNDER X-RAY AND ULTRASOUND THAT A PIECE OF THE GUIDEWIRE REMAINED IN THE PATIENT. PATIENT WAS SENT TO INTERVENTIONAL RADIOLOGY FOR REMOVAL.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6183552 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page