Peripheral Catheter Insertion Kit F220101Recalls & reported problems
Bard Access Systems, Inc. · DOT PFlex, 20G 10cm Full w/o GIVa · Class 2
All models of Peripheral Catheter Insertion Kit (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741110689), received between Jun 23, 2017 and Oct 16, 2019 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak2
- Bent1
- Break1
- Detachment Of Device Component1
- Device Dislodged or Dislocated1
- Device Operates Differently Than Expected1
- Material Split, Cut or Torn1
- Split1
Patient problems
- No Known Impact Or Consequence To Patient6
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2019Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY MS&S, "CALLER STATES SHE IS A NURSE AT A LONG TERM FACILITY THAT USES THE POWERGLIDE PRO. SHE STATES THEY HAVE BEEN EXPERIENCING SOME LEAKING AT THE INSERTION SITE WITH THESE DEVICES AFTER ROUGHLY FOUR DAYS OF USE. CALLER ASKING WHAT COULD BE CAUSES OF THIS ISSUE? CALLER PROVIDE LOT NUMBER RECW0442." MS&S DISCUSSED POSSIBLE CAUSES OF LEAKING AROUND MIDLINES: INFILTRATION, FIBRIN BUILD UP, THROMBUS, OCCLUSIONS, INFLAMMATION FROM VESICANTS/MEDICATIONS, IMPROPER FLUSHING ROUTINE, LOOSE CONNECTION, ETC.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9199794 ↗MalfunctionFeb 14, 2019Material Split, Cut or Torn
IT WAS REPORTED THAT THE "CATHETER SHEARED". NO OTHER INFORMATION PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8340922 ↗MalfunctionJun 14, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT THE NURSE PLACED THE POWERGLIDE INTO THE PATIENT'S BASILIC VEIN AND THE GUIDEWIRE "DISLODGED FROM THE DEVICE INTO THE SKIN TISSUE." A NEW DEVICE WAS PLACED. AFTER THE 2ND DEVICE WAS PLACED SUCCESSFULLY, THE RN NOTICED THAT A PIECE OF THE GUIDEWIRE ON THE 1ST DEVICE WAS IN THE PATIENTS SKIN. SHE PULLED THE GUIDEWIRE OUT OF THE SKIN TISSUE, NO TESTING WAS DONE TO ENSURE ALL OF THE GUIDEWIRE WAS REMOVED FROM THE PATIENT." THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7602369 ↗MalfunctionFeb 1, 2018Split
THE FACILITY REPORTED TO THE SALES REP THAT DURING PLACEMENT OF THE CATHETER THE RN NOTED A SPLIT/SHEAR IN THE MIDDLE OF THE CATHETER. THE CATHETER WAS REMOVED AND A NEW DEVICE WAS PLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7234140 ↗MalfunctionDec 6, 2017Bent
IT WAS REPORTED BY THE FACILITY THAT WHILE REMOVING THE HOUSING UNIT, IT WAS NOTED THAT THE CATHETER WAS BENT. THE DEVICE WAS REMOVED AND A NEW DEVICE USED TO COMPLETE THE PROCEDURE. NO HARM TO THE PATIENT OTHER THAN REQUIRING TWO NEEDLE STICKS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7088162 ↗InjuryAug 31, 2017Device Operates Differently Than Expected
IT WAS REPORTED BY THE SALES REP THAT THE NURSE INSERTED THE MIDLINE IN THE LEFT UPPER ARM OF THE PATIENT, THE PATIENT BEGAN COMPLAINING OF PAIN SO THE NURSE TRIED TO REMOVE THE DEVICE WITHOUT SUCCESS. A SECOND MEMBER OF THEIR TEAM ATTEMPTED TO REMOVE THE DEVICE WAS UNSUCCESSFUL. INTERVENTIONAL RADIOLOGY HAD TO REMOVE THE DEVICE. THE DEVICE WAS SUCCESSFULLY REMOVED AND SECURED. NO PATIENT INJURY WAS REPORTED.
Patient problems: Pain
MAUDE report 6835060 ↗MalfunctionJun 27, 2017Fluid/Blood Leak
PER SALES REP, THE FACILITY REPORTED THAT AFTER THE DEVICE WAS PLACED AND ASPIRATED, THE NURSE FLUSHED THE CATHETER. IT WAS STATED THAT SALINE SQUIRTED FROM THE HUB OF THE CATHETER (WHERE THE CATHETER MEETS THE PINK HUB) WHEN THE DEVICE WAS FLUSHED. NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6672350 ↗MalfunctionJun 23, 2017Break; Detachment Of Device Component
PER SALES REP, THE FACILITY REPORTED THE NURSE TRIED TO THREAD WIRE AND THE CATHETER GOT CAUGHT UP ON THE NEEDLE AND SHEARED. NO OTHER INFORMATION WAS PROVIDED. NO PATIENT HARM WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6663771 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page