ULTRASCORE US14150430Recalls & reported problems
Bard Peripheral Vascular, Inc. · ULTRASCORE™ Focused Force PTA Balloon 4mm x 300mm, 150cm catheter · Class 2
All models of ULTRASCORE (61)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741113758), received between Aug 2, 2023 and Jul 23, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture8
- Device Dislodged or Dislocated2
- Break1
- Detachment of Device or Device Component1
- Device-Device Incompatibility1
- Difficult to Remove1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2025Material Rupture; Material Rupture
ON (B)(6) 2025 DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING AN ULTRASCORE 014 BALLOON. DURING THE PROCEDURE, A RUPTURE OCCURRED DURING TREATMENT OF A SEVERELY CALCIFIED LESION, MAKING IT UNUSABLE DURING THE FIRST INFLATION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22574229 ↗MalfunctionJun 20, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE OF A CALCIFIED TARGET LESION USING THE ULTRASCORE. DURING THE PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 6ATM. IT WAS FURTHER REPORTED THAT ANOTHER BALLOON OF THE SAME SIZE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22294822 ↗MalfunctionJun 18, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRASCORE 014 PTA BALLOON CATHETER TO TREAT SUPERFICIAL FEMORAL ARTERY OBSTRUCTION VIA THE CONTRALATERAL APPROACH. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY RUPTURED UNDER NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22281674 ↗MalfunctionMay 9, 2025Material Rupture; Material Rupture
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRASCORE 014 PTA BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY RUPTURED AT 10 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21991860 ↗MalfunctionJan 18, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED BELOW THE NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21186191 ↗MalfunctionJan 9, 2025Break; Inflation Problem; Material Rupture; Break; Inflation Problem; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE PTA BALLOON WAS ALLEGEDLY SLOWLY EXPANDED AND PARTIAL POOR EXPANSION WAS OBSERVED. IT WAS FURTHER REPORTED THAT WHEN THE BALLOON WAS EXPANDED TO NOMINAL PRESSURE, IT RUPTURED. REPORTEDLY, WHEN IT WAS PULLED OUT OF THE BODY AND CHECKED, THE BODY WIRE WAS ALSO BROKEN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21110093 ↗MalfunctionOct 18, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA IPSILATERAL APPROACH, THE PTA BALLOON ALLEGEDLY RUPTURED IN LESS THAN 6 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20479848 ↗MalfunctionJun 19, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING THE PREPARATION OF AN ANGIOPLASTY PROCEDURE, THE HUB WAS ALLEGEDLY DETACHED. IT WAS FURTHER REPORTED THAT IT ALLEGEDLY CAME OFF EVEN THOUGH LITTLE FORCE WAS APPLIED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19564602 ↗MalfunctionMar 4, 2024Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE HUB OF THE PTA BALLOON ALLEGEDLY CAME OFF THE CATHETER AND THE CATHETER SHAFT WAS ALLEGEDLY SEPERATED AND SLID. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18824222 ↗MalfunctionFeb 20, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED BELOW NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738581 ↗MalfunctionAug 2, 2023Difficult to Remove; Device-Device Incompatibility; Difficult to Remove; Device-Device Incompatibility
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE PTA BALLOON ALLEGEDLY GOT STUCK WITH THE GUIDEWIRE AND WAS UNABLE TO BE REMOVED. IT WAS FURTHER REPORTED THAT THE DEVICE WAS ALLEGEDLY UNABLE TO BE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17440970 ↗
Other Catheter, percutaneous, cutting/scoring devices
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring41 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring31 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring28 reports
- 2cm Peripheral Cutting Balloon® M001PCB8020900Boston Scientific Corporation · Catheter, percutaneous, cutting/scoring18 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2105-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- ULTRASCORE US14150530Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring15 reports
- ULTRASCORE US3513058Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page