ULTRASCORE US14150630Recalls & reported problems

Bard Peripheral Vascular, Inc. · ULTRASCORE™ Focused Force PTA Balloon 6mm x 300mm, 150cm catheter · Class 2

All models of ULTRASCORE (61)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741114106), received between Feb 11, 2022 and Feb 2, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove7
  • Device-Device Incompatibility5
  • Failure to Advance2
  • Material Deformation2
  • Material Rupture2
  • Break1
  • Inflation Problem1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2026Difficult to Remove

    ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE VIA THE SUPERFICIAL FEMORAL ARTERY USING THE ULTRASCORE 014 BALLOON CATHETER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE BALLOON CATHETER COULD NOT BE REINSERTED AFTER INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232631 ↗
  • MalfunctionJul 15, 2025Difficult to Remove; Device-Device Incompatibility; Material Deformation; Difficult to Remove; Device-Device Incompatibility; Material Deformation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE ULTRASCORE 014 IN THE SUPERFICIAL FEMORAL ARTERY VIA IPSILATERAL APPROACH, THE PTA BALLOON CATHETER ALLEGEDLY WAS STUCK WITH THE GUIDEWIRE AND COULD NOT BE REMOVED. IT WAS FURTHER REPORTED THAT THE PTA BALLOON CATHETER WAS ALLEGEDLY STUCK WITH THE INTRODUCER SHEATH TIP AND COULD NOT BE INSERTED INTO THE INTRODUCER SHEATH AND WAS REMOVED TOGETHER WITH THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22504098 ↗
  • MalfunctionJun 9, 2025Break; Break

    A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE ULTRASCORE 014 PTA BALLOON CATHETER. DURING THE PROCEDURE, AN ATTEMPT WAS MADE TO PASS THE LESION, THE BALLOON SHAFT BROKE AND CONTINUED USE WAS NOT POSSIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168812 ↗
  • MalfunctionApr 30, 2025Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Difficult to Remove; Failure to Advance; Device-Device Incompatibility

    ON (B)(6) 2025, DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING AN ULTRASCORE 014 BALLOON, IT GOT STUCK UNDER NOMINAL PRESSURE. FURTHER THERE WAS DIFFICULTY IN ADVANCING THE DEVICE THROUGH THE GUIDEWIRE. ADDITIONALLY, THERE WAS DIFFICULTY RETRACTING THE DEVICE THRU THE INTRODUCER SHEATH OR REMOVING FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21932471 ↗
  • MalfunctionApr 11, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED AT 5 ATM DURING LESION EXPANSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21822929 ↗
  • MalfunctionJan 20, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21189563 ↗
  • MalfunctionJul 22, 2024Difficult to Remove; Device-Device Incompatibility; Difficult to Remove; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, AFTER EXPANSION AT 7 ATM, THE PTA BALLOON ALLEGEDLY DID NOT COME OUT. IT WAS FURTHER REPORTED THAT THE ENTIRE BALLOON ALONG WITH THE WIRE HAD TO BE REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19790248 ↗
  • MalfunctionNov 6, 2023Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA AN IPSILATERAL APPROACH, THE DEVICE ALLEGEDLY FAILED TO BE RETRACTED FROM THE SHEATH AFTER INFLATING. IT WAS FURTHER REPORTED THAT THE CATHETER AND THE SHEATH WERE REMOVED AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18071276 ↗
  • MalfunctionMar 17, 2022Difficult to Remove; Device-Device Incompatibility; Difficult to Remove; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE USING 5FR SHEATH, THE PTA BALLOON WAS ALLEGEDLY STUCK UPON REMOVAL FROM THE PATIENT. IT WAS FURTHER REPORTED THAT THE BALLOON AND THE SHEATH WERE REMOVED AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13798469 ↗
  • MalfunctionFeb 11, 2022Inflation Problem; Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Material Twisted/Bent; Material Split, Cut or Torn; Inflation Problem; Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Material Twisted/Bent; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, A PORTION OF THE PTA BALLOON ALLEGEDLY TWISTED AT NOMINAL PRESSURE. IT WAS FURTHER REPORTED THAT THERE WAS STRONG RESISTANCE FELT UPON REMOVAL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13512778 ↗

Other Catheter, percutaneous, cutting/scoring devices

All Catheter, percutaneous, cutting/scoring devices →