ULTRASCORE US35130520Recalls & reported problems
Bard Peripheral Vascular, Inc. · ULTRASCORE™ Focused Force PTA Balloon 5mm x 200mm, 130cm catheter · Class 2
All models of ULTRASCORE (61)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741116803), received between Dec 27, 2018 and Sep 4, 2024 — classified as 7 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Break1
- Leak/Splash1
Patient problems
- Vascular Dissection6
- Stenosis3
- Fistula1
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2024Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE PTA BALLOON ALLEGEDLY LEAKED FROM THE HUB. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20135091 ↗InjuryJan 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT AFTER USE OF A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, DISSECTION WAS IDENTIFIED. THIRTY DAYS AFTER AN ANGIOPLASTY PROCEDURE WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER WAS PERFORMED, STENOSIS WAS IDENTIFIED VIA DUPLEX ULTRASOUND. APPROXIMATELY ONE YEAR POST INDEX PROCEDURE, RE-INTERVENTION WAS PERFORMED USING A BARE METAL STENT.
Patient problems: Stenosis; Vascular Dissection
MAUDE report 9586859 ↗InjuryJun 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY 7-MONTH POST ANGIOPLASTY PROCEDURE WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, STENOSIS WAS IDENTIFIED IN THE SUPERFICIAL FEMORAL ARTERY. IT WAS FURTHER REPORTED REVASCULARIZATION WAS PERFORMED USING PTA BALLOON.
Patient problems: Stenosis; Vascular Dissection
MAUDE report 8689001 ↗InjuryMar 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT FOLLOWING AN ANGIOPLASTY PROCEDURE IN THE MID 1/3 OF THE LEFT SUPERFICIAL FEMORAL ARTERY WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, DISSECTION WAS IDENTIFIED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN. NEW INFORMATION: IT WAS REPORTED THROUGH THE RESULT OF A CLINICAL TRIAL THAT APPROXIMATELY 7 MONTHS POST INDEX PROCEDURE, TARGET LESION REVASCULARIZATION WAS OCCURRED. IT WAS FURTHER REPORTED THAT A STENT WAS PLACED WITH FOLLOWING BALLOON DILATATION CATHETER. THE PROCEDURE WAS COMPLETED WITHOUT ANY ISSUES. NEW INFORMATION: IT WAS REPORTED THAT APPROXIMATELY 12 MONTHS POST INDEX PROCEDURE, DUPLEX ULTRASOUND WAS PERFORMED AND IDENTIFIED STENOSIS.
Patient problems: Vascular Dissection
MAUDE report 8397371 ↗InjuryFeb 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT DURING AN ANGIOPLASTY PROCEDURE WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, NON FLOW-LIMITING VESSEL DISSECTION WAS IDENTIFIED AFTER TWO INFLATIONS IN THE RIGHT SUPERFICIAL FEMORAL ARTERY AND THE PATIENT WAS HOSPITALIZED. REPORTEDLY, PTA, DRUG COATED BALLOON PTA, BARE METAL STENTING AND MECHANICAL THROMBECTOMY WERE PERFORMED AND THE EVENT WAS CONSIDERED TO BE RESOLVED. THE PATIENT WAS DISCHARGED ONE DAY POST ADMISSION.
Patient problems: Stenosis; Vascular Dissection
MAUDE report 8356501 ↗InjuryFeb 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT DURING AN ANGIOPLASTY PROCEDURE IN THE LEFT SUPERFICIAL FEMORAL ARTERY WITH A FOCUSED FORCED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, FOLLOWING USE OF THE STUDY DEVICE PTA ANGIOGRAPHY DEMONSTRATED MICROPERFORATIONS AND AV FISTULA. THE TARGET LESION WAS TREATED WITH A PTA BALLOON AND ADEQUATE FLOW WAS IDENTIFIED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Fistula; Perforation
MAUDE report 8346329 ↗InjuryJan 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT FOLLOWING AN ANGIOPLASTY PROCEDURE IN THE MID 1/3 AND DISTAL 1/3 OF THE RIGHT SUPERFICIAL FEMORAL ARTERY WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, DISSECTION WAS IDENTIFIED WHICH REQUIRED PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY, DRUG COATED PTA BALLOON, AND PLACEMENT OF TWO STENTS. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Vascular Dissection
MAUDE report 8276555 ↗InjuryDec 27, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT DURING AN ANGIOPLASTY PROCEDURE WITH A FOCUSED FORCE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER, VESSEL DISSECTION WAS IDENTIFIED AFTER ONE INFLATION IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. REPORTEDLY, THE DISSECTION WAS TREATED WITH A STENT. THE EVENT WAS CONSIDERED TO BE RESOLVED. THE PATIENT WAS DISCHARGED THE SAME DAY AS THE INDEX PROCEDURE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Vascular Dissection
MAUDE report 8200874 ↗
Other Catheter, percutaneous, cutting/scoring devices
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring41 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring31 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring28 reports
- 2cm Peripheral Cutting Balloon® M001PCB8020900Boston Scientific Corporation · Catheter, percutaneous, cutting/scoring18 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2105-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- ULTRASCORE US14150530Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring15 reports
- ULTRASCORE US3513058Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page