Skip to content

AllPoints Port Access System PowerLoc Max AN142075GRecalls & reported problems

Bard Access Systems, Inc. · AllPoints Port Access System PowerLoc Max 20G x 0.75 inch + GuardIVa · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741119316), received between Oct 31, 2024 and Feb 6, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • Needle Stick/Puncture1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED 28 JAN. 2026. THE KITS WERE SUPPOSED TO HAVE THE BALLOONED DRESSING THAT DOES NOT STICK TO THE HUB OF THE NEEDLE. INSTEAD, THE DRESSING WAS SUBSTITUTED WITH A REGULAR TEGADERM WITH CHG, WHICH STUCK TO THE ENTIRE HUBER NEEDLE AND WOULD NOT ALLOW THE SAFETY TO ENGAGE. THE NURSE WAS PEELING THE DRESSING OFF WHEN THE NEEDLE CAME OUT OF THE PATIENT AND SHE WAS STUCK. THE PATIENT DID NOT HAVE ANY BLOOD BORNE DISEASES, BUT BLOOD WORK WAS COMPLETED FOR THE NURSE AS A PRECAUTION. THERE WAS NO LONG TERM HARM REPORTED.

    Patient problems: Needle Stick/Puncture

    MAUDE report 24277323 ↗
  • MalfunctionOct 31, 2024Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER, POWERLOC MAX NEEDLE FROM WITHIN ALL POINTS KIT BENT WHEN ACCESSED PATIENT. CUSTOMER SAYS THE NEEDLE WAS NOT INITIALLY BENT AND IS UNSURE AT WHAT POINT THE NEEDLE BENT. 10/10/2024 - THE SAFETY MECHANISM FAILED TO ACTIVATE ON THE RETURNED DEVICE DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20580842 ↗

Other Non-coring (Huber) needle devices

All Non-coring (Huber) needle devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page