AllPoints Port Access System PowerLoc Max AN142075GRecalls & reported problems
Bard Access Systems, Inc. · AllPoints Port Access System PowerLoc Max 20G x 0.75 inch + GuardIVa · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741119316), received between Oct 31, 2024 and Feb 6, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- Needle Stick/Puncture1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED 28 JAN. 2026. THE KITS WERE SUPPOSED TO HAVE THE BALLOONED DRESSING THAT DOES NOT STICK TO THE HUB OF THE NEEDLE. INSTEAD, THE DRESSING WAS SUBSTITUTED WITH A REGULAR TEGADERM WITH CHG, WHICH STUCK TO THE ENTIRE HUBER NEEDLE AND WOULD NOT ALLOW THE SAFETY TO ENGAGE. THE NURSE WAS PEELING THE DRESSING OFF WHEN THE NEEDLE CAME OUT OF THE PATIENT AND SHE WAS STUCK. THE PATIENT DID NOT HAVE ANY BLOOD BORNE DISEASES, BUT BLOOD WORK WAS COMPLETED FOR THE NURSE AS A PRECAUTION. THERE WAS NO LONG TERM HARM REPORTED.
Patient problems: Needle Stick/Puncture
MAUDE report 24277323 ↗MalfunctionOct 31, 2024Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER, POWERLOC MAX NEEDLE FROM WITHIN ALL POINTS KIT BENT WHEN ACCESSED PATIENT. CUSTOMER SAYS THE NEEDLE WAS NOT INITIALLY BENT AND IS UNSURE AT WHAT POINT THE NEEDLE BENT. 10/10/2024 - THE SAFETY MECHANISM FAILED TO ACTIVATE ON THE RETURNED DEVICE DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20580842 ↗
Other Non-coring (Huber) needle devices
- SafeStep CPA00049Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00040Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00042ABard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- INTERA AP-07014USIntera Oncology Inc. · Non-Coring (Huber) Needle37 reports
- SURECAN SAFETY II 4447006-02B. Braun Medical Inc. · Non-Coring (Huber) Needle13 reports
- SURECAN SAFETY II 4447005-02B. Braun Medical Inc. · Non-Coring (Huber) Needle6 reports
- AllPoints Port Access System SafeStep AN152075MBard Access Systems, Inc. · Non-coring (Huber) needle5 reports
- SURECAN SAFETY II 4447012-02B. Braun Medical Inc. · Non-Coring (Huber) Needle4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page