PowerPICC SOLO CK000544ARecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC SOLO (78)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741121593), received between Oct 4, 2016 and Oct 4, 2016 — classified as 1 malfunction.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Bent1
- Break1
- Component Falling1
- Physical Resistance1
- Retraction Problem1
Patient problems
- No Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2016Bent; Break; Component Falling; Retraction Problem; Physical Resistance
PER SALES REP, FACILITY REPORTED THAT THE PICC WAS DIFFICULT TO PLACE AND WAS GOING UP THE IJ. IT WAS STATED THAT THE NURSE WITHDREW THE CATHETER AND NOTICED THE WIRE WAS BENT AND IT FELL OFF IN HER HAND. THEY REPORTED THAT THERE WAS NO HARM TO THE PATIENT. THE FACILITY REPORTED THE GUIDEWIRE COMPLETELY BROKE AS THE NURSE RETRACTED IT FROM THE CATHETER. THE NURSE STATED, "I HAD MET SOME RESISTANCE ADVANCING THE CATHETER WITH THE GUIDEWIRE IN. I RETRACTED THE CATHETER AND ATTEMPTED TO RE- ADVANCE IT WHILE FLUSHING IT. STILL MET WITH RESISTANCE SO I REMOVED THE GUIDE WIRE. IT IS THEN I OBSERVED THE WIRE WITH A BEND AND WHEN I TOUCHED IT, IT BROKE OFF. I READVANCED THE CATHETER WHILE FLUSHING AND WITHOUT SHERLOCK CONFIRMATION OF THE TIP LOCATION. IT WAS IN THE SVC BUT I OBSERVED THAT THE PATH TAKEN WAS A SHARP 90 DEGREE DROP SO I THINK THE CATHETER WAS HITTING AT THE 90 TURN AND TRIED FLIPPING BACKWARDS WHICH WEAKENED THE WIRE."
Patient problems: No Information
MAUDE report 5998429 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗