PowerPICC SOLO CK000525ARecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC SOLO (78)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741124075), received between Sep 20, 2018 and Mar 26, 2021 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Remove1
- Failure to Advance1
- Material Deformation1
- Material Frayed1
- Material Split, Cut or Torn1
- Material Twisted/Bent1
- Problem with Sterilization1
Patient problems
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pain1
- Swelling1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2021Material Frayed; Material Frayed
IT WAS REPORTED THAT A GUIDE WIRE FRAYED. ADDITIONAL INFORMATION RECEIVED 3/17/2021: IT WAS REPORTED THE TIP OF THE GUIDE WIRE WAS FRAYED UPON REMOVAL DURING A PICC LINE INSERTION PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11576702 ↗MalfunctionMar 25, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT THE INTRODUCER WAS NOT STRONG AND MUSHROOMED BACK/BENT CAUSING A NEED TO NICK SKIN OF THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9879885 ↗MalfunctionJul 31, 2019Difficult to Remove; Material Twisted/Bent
IT WAS REPORTED THAT THE PICC LINE KNOTTED AND DIFFICULT TO REMOVE. REMOVED WITHOUT ANY PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8848313 ↗InjuryDec 21, 2018Failure to Advance
IT WAS REPORTED THAT A DUAL LUMEN POWERPICC SOLO INSERTION WAS ATTEMPTED ON (B)(6) 2018, UNSUCCESSFUL W/ MULTIPLE ATTEMPTS. IT WAS STATED THAT RN WAS ABLE TO CANNULATE BUT WAS NOT PLACED IN SVC- ABORTED. THEN A SECOND CLINICIAN ATTEMPTED WITH A SINGLE LUMEN, AND UNSUCCESSFUL. THE PATIENT CAME BACK TO THE HOSPITAL (MONDAY (B)(6) 2018) WITH COMPLAINTS OF SWELLING AND PAIN IN RIGHT ARM. THEN PATIENT WAS ADMITTED AFTER ASSESSMENTS AND TESTS IT WAS NOTED A 2.6 CM SEGMENT OF WIRE IN RIGHT SUBCLAVIAN VEIN. THIS REPORT ADDRESSES THE FIRST CATHETER. ADDITIONAL INFORMATION RECEIVED STATED, "WE WERE UNABLE TO ADVANCE THE CATHETER PAST THE LEVEL OF THE RIGHT SUBCLAVIAN. RESISTANCE TO THE CATHETER WAS NOTED BY BOTH PRACTITIONERS AT THE SAME POINT (THE LEVEL OF THE RIGHT SUBCLAVIAN). NO ATTEMPTS WERE MADE TO ADVANCE THE CATHETER OR THE WIRE BEYOND THAT. CONTRALATERAL ATTEMPT NOT POSSIBLE DUE TO RADICAL MASTECTOMY WITH LYMPH NODE RESECTION."
Patient problems: Foreign Body In Patient; Pain; Swelling
MAUDE report 8189527 ↗MalfunctionSep 20, 2018Problem with Sterilization; Material Split, Cut or Torn
IT WAS REPORTED POWERPICC SOLO WAS TORN AT BASE OF CATHETER OUT OF STERILE PACKAGE. IT WAS NOT USED ON PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7896182 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page