PowerPICC Provena S3173335Recalls & reported problems
Bard Access Systems, Inc. · 3 Fr Single Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741129964), received between Dec 3, 2020 and Aug 26, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Markings/Labelling Problem2
- Material Frayed1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2022Material Frayed; Material Frayed
IT WAS REPORTED "ACCOUNT HAD A PICC WIRE UNRAVEL WHILE IN THE PATIENT. THEY WERE ABLE TO GET THE WIRE OUT. IT WAS NOT LEFT IN THE PATIENT." ON 08/11/2022 ADD INFO REC'D: CATHETER SECUREMENT WE USED A SINGLE SUTURE TO SECURE THE CATHETER. FLUORO GUIDANCE WAS UTILIZED AS ALWAYS FOR OUR PICC LINE PLACEMENTS. THE PATIENT DID NOT REQUIRE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15303251 ↗MalfunctionDec 3, 2020Device Markings/Labelling Problem
IT WAS REPORTED "THE PICC LINES COME MANUFACTURED WITH MARKINGS ON THE CATHETER. THE MARKINGS REPRESENT A CENTIMETER AND A NUMBER IS REPRESENTED EVERY 5TH MARKING (EX. 5, 10, 15 ETC). IT IS DOCUMENTED IN THE PATIENT¿S CHART THE LENGTH THE CATHETER IS MEASURED AND CUT. WE HAD TO OPEN AND USE TWO SEPARATE PICC LINE PACKAGES FOR THIS PATIENT BECAUSE THE MARKINGS WERE INACCURATE. ONE CATHETER HAD THE NUMBER 10 MARKED AFTER 12 CENTIMETER MARKS AND THE OTHER CATHETER HAD THE NUMBER 10 AFTER 11 MARKS. THIS DISCREPANCY MAKES IT DIFFICULT TO GET AN ACCURATE MEASUREMENT AND CAN RESULT IN OPENING AND USING MULTIPLE PICC CATHETER KITS PER PATIENT. ALSO, FOR REMOVAL PURPOSES THE LENGTH SHOULD BE ACCURATE AND CORRELATE WITH THE DOCUMENTED MEASUREMENT TO ENSURE NO PART OF THE CATHETER WAS RETAINED DURING REMOVAL." THIS REPORT ADDRESSES THE CATHETER THAT HAD THE NUMBER 10 MARKED AFTER 11 CENTIMETER MARK.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10942951 ↗MalfunctionDec 3, 2020Device Markings/Labelling Problem
IT WAS REPORTED "THE PICC LINES COME MANUFACTURED WITH MARKINGS ON THE CATHETER. THE MARKINGS REPRESENT A CENTIMETER AND A NUMBER IS REPRESENTED EVERY 5TH MARKING (EX. 5, 10, 15 ETC). IT IS DOCUMENTED IN THE PATIENT¿S CHART THE LENGTH THE CATHETER IS MEASURED AND CUT. WE HAD TO OPEN AND USE TWO SEPARATE PICC LINE PACKAGES FOR THIS PATIENT BECAUSE THE MARKINGS WERE INACCURATE. ONE CATHETER HAD THE NUMBER 10 MARKED AFTER 12 CENTIMETER MARKS AND THE OTHER CATHETER HAD THE NUMBER 10 AFTER 11 MARKS. THIS DISCREPANCY MAKES IT DIFFICULT TO GET AN ACCURATE MEASUREMENT AND CAN RESULT IN OPENING AND USING MULTIPLE PICC CATHETER KITS PER PATIENT. ALSO, FOR REMOVAL PURPOSES THE LENGTH SHOULD BE ACCURATE AND CORRELATE WITH THE DOCUMENTED MEASUREMENT TO ENSURE NO PART OF THE CATHETER WAS RETAINED DURING REMOVAL." THIS REPORT ADDRESSES THE CATHETER THAT HAD THE NUMBER 10 MARKED AFTER 12 CENTIMETER MARK.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10942910 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page