PowerPICC Provena S1274108Recalls & reported problems
Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00801741130120), received between Nov 1, 2017 and Nov 7, 2024 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak7
- Break4
- Material Deformation4
- Material Frayed3
- Leak/Splash2
- Chemical Spillage1
- Crack1
- Defective Component1
Patient problems
- No Known Impact Or Consequence To Patient11
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient1
- Underdose1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2024Defective Component; Defective Component
IT WAS REPORTED THAT PATIENT CAME FOR A DRESSING CHANGE AND TO CHANGE THE BROKEN CLAMP TO THE RED LUMEN. ACCORDING TO PATIENT, THE CLAMP HAS BEEN BROKEN FOR A WHILE BUT NURSE UNABLE TO CHANGE IT OUT DUE TO NO AVAILABILITY OF THE REPLACEMENT CLAMP. I REMOVED THE BROKEN CLAMP AND REPLACED WITH A NEW CLAMP WITHOUT ANY DIFFICULTY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20631511 ↗MalfunctionJan 28, 2023Fluid/Blood Leak; Fluid/Blood Leak
BD VASCULAR ACCESS DEVICES FIELD ASSURANCE RECEIVED A LEAKING PICC FOR EVALUATION. A NOTE FROM THE CUSTOMER ATTACHED TO THE SAMPLE STATED PICC WAS INSERTED IN THE BASILIC VEIN. ADDITIONAL INFORMATION RECEIVED 1/24/2023: CUSTOMER REPORTS THE PATIENT HAD THE PICC REPLACED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261233 ↗MalfunctionJan 18, 2022Material Deformation; Material Deformation
IT WAS REPORTED "DURING PLACEMENT OF ANOTHER 4 FR. DUAL LUMEN PICC WE HAD A WIRE WITH A SIGNIFICANT BEND IN IT UPON REMOVAL FROM THE PATIENT. THE WIRE WAS INTACT." (B)(6) 2022: RETURNED STYLET WIRE IS KINKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13283248 ↗MalfunctionSep 23, 2021Material Frayed; Material Frayed
IT WAS REPORTED "A PICC LINE WAS PLACED FOR THIS PATIENT DUE TO THE NEED FOR TPN AND LONG TERM VENOUS ACCESS. THE LEFT UPPER EXTREMITY WAS CHOSEN FOR PICC PLACEMENT. THE VEIN WAS ACCESSED, A NITINOL GUIDEWIRE WAS PLACED AND A PEEL AWAY INTRODUCER WAS PLACED OVER THE GUIDEWIRE. THE GUIDEWIRE WAS REMOVED INTACT AND DISCARDED. THE PICC LINE WAS THEN THREADED THROUGH THE INTRODUCER, AFTER APPROX. 20 CM OF THE PICC LINE BEING INSERTED RESISTANCE WAS MET. THIS CAN BE A COMMON FINDING, I REPOSITIONED THE ARM MULTIPLE TIMES AS THIS CAN AID WITH PASSING THE CATHETER THROUGH TO THE SVC. AFTER MULTIPLE REPOSITIONING ATTEMPTS I WAS ABLE TO GET THE CATHETER TO PASS WITHOUT RESISTANCE, THE NAVIGATION SENSOR WE USE WAS SHOWING THAT THE CATHETER WAS TRAVELING THE PROPER DIRECTION, HOWEVER THE ECG TRACING THAT WE USE FOR DEFINITIVE PLACEMENT WAS ACTING UP, THE SIGNAL WAS NOT CONSISTENT. I MADE THE DECISION TO PULL THE 3CG WIRE/STYLET AND OBTAIN A CHEST FILM FOR DEFINITIVE PLACEMENT VERIFICATION. UPON REMOVING THE 3CG STYLET I NOTED THAT THE DISTAL END OF THE STYLET WAS DEFORMED AND THERE WAS A NOTICEABLE HARSH BEND IN THE WIRE 5 CM BACK FROM THE DISTAL TIP. I PHYSICALLY EXAMINED THE END OF THE STYL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12520504 ↗MalfunctionAug 26, 2021Material Frayed; Material Frayed
IT WAS REPORTED THAT TWO GUIDEWIRE STYLETS BENT AND/OR FRACTURED OFF DURING PICC INSERTION. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12374046 ↗MalfunctionAug 26, 2021Material Frayed; Material Frayed
IT WAS REPORTED THAT TWO GUIDEWIRE STYLETS BENT AND/OR FRACTURED OFF DURING PICC INSERTION. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12374003 ↗MalfunctionAug 20, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT PICC REQUIRED REPLACEMENT DUE TO INTERNAL KINKS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12351419 ↗MalfunctionAug 20, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT PICC REQUIRED REPLACEMENT DUE TO INTERNAL KINKS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12351410 ↗MalfunctionMay 14, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT SOME MECHANICAL OCCLUSIONS ARE HAPPENING DUE TO KINKING OF THE LINES, PRIMARILY ON 4F DL AND 5F TL LINES. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11826711 ↗MalfunctionMay 9, 2018Break; Nonstandard Device
IT WAS REPORTED THAT THE PATIENT NEEDED A PICC REPLACEMENT DUE TO A BREAK NEAR THE HUB OF THE CATHETER. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7500687 ↗MalfunctionApr 5, 2018Fluid/Blood Leak
IT WAS REPORTED THAT DURING PICC LINE PLACEMENT, BLOOD STARTED COMING OUT OF THE PURPLE LUMEN EVEN THOUGH IT WAS CLAMPED. THEY CLAMPED IT AGAIN AND THEN OBTAINED A NEW CATHETER TO CONTINUE WITH THE PROCEDURE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7401460 ↗MalfunctionMar 23, 2018Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT WHEN THE VAS TEAM ARRIVED TO ASSESS THE PATIENT¿S PICC, THE PATIENT'S DRESSING HAD BEEN REMOVED AND THE LUMENS WERE WRAPPED IN AN ABDOMINAL PAD DUE TO LEAKING. IT WAS NOTED THAT IT WAS FRACTURED ¿WHERE THE TUBING ENTERS INTO THE END PORTION OF THE LUMENS¿ AND THAT THE FRACTURE WAS VISIBLE. IT WAS ALSO NOTED THAT ¿BLOOD WAS LEAKING OUT.¿ THE PICC WAS REMOVED AND A STERILE DRESSING WAS PLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7367107 ↗MalfunctionFeb 7, 2018Break; Defective Device; Device Dislodged or Dislocated
THE FACILITY REPORTED THE NIGHT SHIFT WAS CALLED LAST NIGHT FOR A DEFECTIVE PICC. ON ARRIVAL VAST IT WAS FOUND THAT THE PORT WAS "HANGING BY A THREAD". IT WAS STATED WHEN THE NURSE PULLED IT OUT, THE END BROKE OFF.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7250187 ↗MalfunctionFeb 6, 2018Break
THE FACILITY REPORTED THE PICC WAS REMOVED AND REPLACED DUE TO A HUB/CATHETER BREAK. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7246316 ↗MalfunctionFeb 6, 2018Break; Fluid/Blood Leak
THE FACILITY REPORTED THE CATHETER WAS REPLACED DUE TO BREAK/LEAK AT THE HUB. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7246314 ↗MalfunctionFeb 6, 2018Detachment Of Device Component
THE FACILITY REPORTED THAT THE "CAP CAME OF THE PICC PORT". THE CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7246311 ↗MalfunctionFeb 6, 2018Fluid/Blood Leak
THE FACILITY REPORTED THE CATHETER WAS REPLACED DUE TO BREAK/LEAK AT THE HUB. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7246259 ↗MalfunctionJan 4, 2018Crack; Chemical Spillage
IT WAS REPORTED BY THE SALES REP THAT THE FACILITY REPORTED A CRACK IN THE HUB OF A PROVENA. THE SALES REP STATED THAT CHEMO LEAKED EVERYWHERE INCLUDING "ONTO THE PATIENT". THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7164181 ↗MalfunctionJan 3, 2018Leak/Splash
IT WAS REPORTED BY THE FACILITY THAT THE PICC WAS PLACED ON (B)(6) 2017. ON (B)(6) 2017, IT WAS DISCONNECTED AS A LEAK WAS FOUND BENEATH THE RED HUB AND EXTENSION JUNCTURE. PATIENT WAS ON "VASOSUPRESSORS" AND COULDN'T KEEP HIS BLOOD PRESSURE UP. WHEN TRACING THE IV TUBING BACK TO THE PICC, IT WAS NOTED THAT THE MEDICATION WAS LEAKING ONTO THE BED AND THE PATIENT WASN'T RECEIVING MEDICATION.
Patient problems: Underdose
MAUDE report 7162840 ↗MalfunctionDec 14, 2017Fluid/Blood Leak
IT WAS REPORTED BY THE SALES REP THAT THE FACILITY REPORTED HAVING A PICC LINE THAT WAS LEAKING FROM THE RED HUB THAT CONNECTED TO THE EXTENSION LEG. THE SALES REP STATED THAT THERE WAS NO VISIBLE CRACK. IT WAS REPORTED THAT ONLY ONE LEG WAS USED, AS THEY DID NOT NEED THE DUAL LUMEN AND THE LINE WAS PULLED UPON COMPLETION OF THERAPY. NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118125 ↗MalfunctionDec 14, 2017Fluid/Blood Leak
IT WAS REPORTED BY THE SALES REP THAT THE FACILITY REPORTED HAVING A PICC LINE THAT WAS LEAKING FROM THE PURPLE HUB THAT CONNECTED TO THE EXTENSION LEG. THE SALES REP STATED THAT THERE WAS NO VISIBLE CRACK. IT WAS REPORTED THAT ONLY ONE LEG WAS USED, AS THEY DID NOT NEED THE DUAL LUMEN AND THE LINE WAS PULLED UPON COMPLETION OF THERAPY. NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7118005 ↗MalfunctionNov 1, 2017Hole In Material; Leak/Splash
THE FACILITY REPORTED THAT THE PROVENA CATHETER LEAKED FROM A PINHOLE BETWEEN THE HUB AND TUBING. THE LINE WAS REMOVED. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6993882 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page