PowerPICC Provena S3274108Recalls & reported problems
Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741130168), received between Oct 20, 2017 and Mar 18, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Break1
- Device Or Device Fragments Location Unknown1
- Failure to Infuse1
- Loss of or Failure to Bond1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2026Failure to Infuse
IT WAS REPORTED THAT DURING A DIFFICULT CATHETER INSERTION, THE RN NOTED THAT THE RED LUMEN WAS DIFFICULT TO FLUSH AND THEY COULD NOT OBTAIN A BLOOD RETURN. STYLET THAT IS IN THE RED LUMEN FROM THE MANUFACTURER WAS REMOVED DURING PROCEDURE. PURPLE LUMEN HOUSED THE ADDED ARROW BRAND VASCULAR POSITIONING STYLET AND THIS LUMEN REMAINED PATENT THROUGHOUT THE PROCEDURE, EASY TO FLUSH WITH BLOOD RETURN. SUSPECTING THE PICC MAY BE KINKED OR "KNOTTED", SHE RETRACTED THE PICC A FEW CM AT A TIME TO ASSESS PATENCY. NO AMOUNT OF RETRACTING THE PICC DURING THE PROCEDURE RE-ESTABLISHED PATENCY. THIS PICC WAS REMOVED AND ANOTHER PICC KIT WAS OPENED TO COMPLETE THE PROCEDURE. ONCE THE FAULTY PICC WAS REMOVED FROM THE PT, WRITER ATTEMPTED TO FLUSH THE RED LUMEN, AGITATE WITH FLUID, ETC WITH ONLY DRIPS OF FLUID PASSING THROUGH THE LUMEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24634081 ↗MalfunctionFeb 13, 2026Fluid/Blood Leak
IT WAS REPORTED CATHETER WAS FOUND WITH A HOLE JUST DISTAL TO THE HUB, DISCOVERED DUE TO LEAKAGE UNDER THE DRESSING. IV MEDICATION TREATMENT WAS INTERRUPTED, AND PATIENT HAD TO UNDERGO AN EXCHANGE OF THE LINE. LINE WAS SECURED WITH SECURACATH. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24337395 ↗MalfunctionFeb 13, 2026Fluid/Blood Leak
IT WAS REPORTED CATHETER WAS FOUND WITH A HOLE JUST DISTAL TO THE HUB, DISCOVERED DUE TO LEAKAGE UNDER THE DRESSING. IV MEDICATION TREATMENT WAS INTERRUPTED, AND PATIENT HAD TO UNDERGO AN EXCHANGE OF THE LINE. LINE WAS SECURED WITH A SUTURE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24337392 ↗MalfunctionJul 1, 2021Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THAT ON 3 FR PICC KITS THAT HAVE THE NEW DRESSING DOES NOT ADHERE TO THE PATIENT'S SKIN. IT WAS STATED THIS HAS OCCURRED WITH EVERY PATIENT HOWEVER AN EXACT QUANTITY WAS NOT PROVIDED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12103686 ↗InjuryOct 20, 2017Break; Device Or Device Fragments Location Unknown
IT WAS REPORTED BY THE FACILITY THAT THE DEVICE WAS PLACED IN THE LEFT CEPHALIC VEIN AND THEY WERE UNABLE TO DETERMINE THE EXACT LOCATION OF THE SUSPECTED GUIDEWIRE BREAK. THEY REPORTED THAT THE GUIDEWIRE EMBOLI ZED IN THE DISTAL PULMONARY ARTERY BRANCH AND THAT THE GUIDEWIRE HAS NOT BEEN REMOVED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6965378 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page