PowerPICC Provena S3294108DRecalls & reported problems
Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741130410), received between Aug 8, 2023 and Aug 8, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE STIFFENING WIRE IN THE BARD PICC BENDS AND CANNOT BE ADVANCED ONCE THE STIFFENING WIRE IS ALTERED. HAD TO USE 3 PICC LINE KITS ON THIS PATIENT. STIFFENING WIRE INSIDE PICC LINE BENT AT A 90 DEGREE APPROXIMATELY 15 CM FROM HUB. UNABLE TO FULLY THREAD PICC LINE AFTER WIRE BENT. HAD TO TRIM THE PICC LINE TO WHERE THE STIFFENING WIRE BENT TO USE PICC LINE. IT WAS REPORTED THAT THREE DIFFERENT PICC KITS WITH THREE DIFFERENT MATERIAL NUMBERS WERE USED. THIS REPORT ADDRESSES THE FIRST KIT AND MATERIAL NUMBER USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17480448 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page