PowerPICC CK000417BRecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC (118)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741134043), received between Jul 17, 2019 and Jul 17, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Failure to Advance1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2019Break; Retraction Problem; Failure to Advance
IT WAS REPORTED THAT "PT WAS ASKED TO HAVE PICC PLACED IN ICU FOR DIFFICULT ACCESS. PT IS A CHRONIC CHILD BUT HAS NEVER HAD A PICC BEFORE. PICC TEAM ATTEMPTED TO PLACE PICC TO RIGHT BASILIC AND WAS ABLE TO ACCESS VEIN AND THREAD WIRE EASILY AS WELL AS DILATATE THE VEIN. HOWEVER; WHEN ATTEMPTING TO PLACE THE 3FR TO RIGHT BASILIC PICC LINE WOULD NOT THREAD PAST 10CM IN RIGHT ARM. PICC RN ATTEMPTED TO RETRACT PICC LINE AND PICC LINE WOULD NOT RETRACT. AFTER TWISTING PICC LINE IN VESSEL , PICC LINE WAS ABLE TO BE RETRACTED. UPON EXAMINING THE PICC BEFORE ATTEMPTING TO THREAD AGAIN PICC RN NOTICED WIRE PROTRUDING FROM THE TIP OF CATHETER. ATTEMPTED AT THAT TIME TO RETRACT WIRE AT THE LUMEN OF THE PICC AND MAGNET ON TIP OF PICC WOULD NOT RETRACT. PICC RN THEN ATTEMPT TO ADVANCE WIRE THROUGH TIP OF PICC AND WHEN SHE DID A PIECE OF THE WIRE WAS NOTED TO BE BROKEN AND FELL TO THE STERILE DRAPE. ATTEMPTED TO ADVANCE WIRE AGAIN AND ANOTHER PIECE ALSO FELL ONTO THE STERILE DRAPE. THAT PICC KIT WAS THEN DISCARDED AND ANOTHER 3 FR KIT WAS THEN USED." IT WAS STATED THAT THERE WAS NO HARM TO THE PATIENT AS "PICC CATHETER REMAINED INTACT AND NONE OF THE BROKEN WIRE PIECES WERE RETAINED IN THE PATIE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8800161 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page