Skip to content

PowerPICC Provena S1274108PDRecalls & reported problems

Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant

All models of PowerPICC Provena (116)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard PowerPICC catheter is used for long-term intravascular therapy. The 40 reports mostly describe the introducer deforming, leaking, or being hard to position, sometimes delaying the procedure; nine mention injuries and no deaths. There are no recalls naming this device identifier.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00801741134166), received between Jun 15, 2017 and Dec 17, 2025 — classified as 9 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation10
  • Adverse Event Without Identified Device or Use Problem8
  • Leak/Splash7
  • Defective Component2
  • Device Operates Differently Than Expected2
  • Difficult To Position2
  • Fluid/Blood Leak2
  • Hole In Material2

Patient problems

  • No Known Impact Or Consequence To Patient12
  • No Clinical Signs, Symptoms or Conditions8
  • Thrombus8
  • No Consequences Or Impact To Patient7
  • Insufficient Information2
  • Blood Loss1
  • Foreign Body In Patient1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED, THE DARK GRAY PIECE BUNCHED UP ON THE INTRODUCER AND WOULD NOT GO INTO THE VEIN. IT WAS A REPLACEMENT KIT. ADDITIONAL INFORMATION: NO ADVERSE EFFECTS ON THE PATIENT OR HEALTH CARE PROVIDERS. I HAD TO DROP A NEW INTRODUCER SET ON THE STERILE FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23830947 ↗
  • MalfunctionDec 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED, THE DARK GRAY PIECE BUNCHED UP ON THE INTRODUCER AND WOULD NOT GO INTO THE VEIN. IT WAS A REPLACEMENT KIT. ADDITIONAL INFORMATION: NO ADVERSE EFFECTS ON THE PATIENT OR HEALTH CARE PROVIDERS. I HAD TO DROP A NEW INTRODUCER SET ON THE STERILE FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23830133 ↗
  • MalfunctionNov 19, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE NURSE WAS INSERTING A PICC AND THE INTRODUCER WENT THROUGH THE SKIN ON A PEDIATRIC PATIENT BUT GOT HUNG UP ON THE 1 CM INTO THE TISSUE. THE NURSE HAD TO GET A MICROEZ INSERTION KIT AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23596003 ↗
  • MalfunctionOct 28, 2025Material Deformation; Material Deformation

    IT WAS REPORTED NURSE WAS INSERTING THE DILATOR INTO THE PEDIATRIC PATIENT FOR PICC PLACEMENT AND COULD NOT GET IT TO GO THROUGH. IT WAS HUNG UP ON THE PATIENT'S SKIN. SHE ENDED UP GETTING A NEW MICRO-INTRODUCER KIT TO HELP HER INSERT. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23409354 ↗
  • MalfunctionJul 31, 2024Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER THE PEEL AWAY INTRODUCER MUSHROOMS OUT, FLARES. NO PATIENT HARM BUT CAUSED A LONGER PROCEDURE. NO OTHER INFORMATION WAS PROVIDED. TWO DEVICES WERE RETURNED FOR INVESTIGATION. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19877486 ↗
  • MalfunctionJul 31, 2024Material Deformation; Material Deformation

    IT WAS REPORTED BY CUSTOMER THAT PEEL AWAY INTRODUCER MUSHROOMS (FLARES) OUT CAUSING A LONGER TRAUMATIZING PROCEDURE. NO OTHER INFORMATION WAS PROVIDED. TWO DEVICES WERE RETURNED FOR INVESTIGATION. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19873782 ↗
  • MalfunctionFeb 17, 2023Defective Component; Defective Component

    IT WAS REPORTED BY THE CUSTOMER, 'WIRE IN THE KIT IS NOT SMOOTH ON INSERTION'. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 16396496 ↗
  • MalfunctionFeb 14, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER, 'PEEL AWAY ROUTINELY BUCKLES'. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 16375782 ↗
  • MalfunctionOct 7, 2021Material Frayed; Material Frayed

    IT WAS REPORTED "3FR. PICC THAT HAD A BROKEN WIRE OVER THE WEEKEND." ADDITIONAL INFO. RECEIVED (B)(6) 2021 "DIFFICULTY ADVANCING PICC, NOTED TO HAVE BROKEN WIRE AFTER MULTIPLE ATTEMPTS TO PASS THROUGH INTRODUCER. WHEN PULLED OUT, WIRE NOTED TO BE BROKEN, OTHER PIECE OF WIRE RETRIEVED FROM PICC."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12591016 ↗
  • MalfunctionFeb 19, 2021Material Deformation; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEVICE RETURNED FOR EVALUATION WAS FOUND KINKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11349160 ↗
  • InjuryNov 18, 2020Break

    IT WAS REPORTED THE WIRE BROKE INSIDE THE PATIENT. THE RN STATED, "I BELIEVE IT WAS THE STYLET, BUT NOT SURE YET. IT IS FROM A PICC PLACED A MONTH AGO, AND THE PIECE IS NOW LODGED IN THE PULMONARY ARTERY." ADD INFO RCVD 10/27/2020: DIFFICULTY ADVANCING PICC PAST SHOULDER WHAT TYPE OF CATHETER SECUREMENT WAS UTILIZED: STATLOCK WAS THERE PATIENT HARM REPORTED?: SO FAR PT IS ASYMPTOMATIC. WAITING TO SEE IF WIRE CAN BE REMOVED IN IR.

    Patient problems: Foreign Body In Patient

    MAUDE report 10864654 ↗
  • MalfunctionOct 29, 2020Fluid/Blood Leak; Material Puncture/Hole; Communication or Transmission Problem

    IT WAS REPORTED "THE CATHETER HAD A HOLE IN THE LINE THAT CAUSED ALL THE SALINE TO LEAK OUT AND NOT ALLOW THE SHERLOCK TO PICK UP THE SIGNAL FROM THE CATHETER."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10759225 ↗
  • InjuryApr 22, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9985436 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT.

    Patient problems: Thrombus

    MAUDE report 9983613 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983440 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983369 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983140 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983139 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983137 ↗
  • InjuryApr 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PICC WAS PLACED ON (B)(6) 2020 AND PATIENT DEVELOPED A DVT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Thrombus

    MAUDE report 9983133 ↗
  • MalfunctionDec 6, 2019Defective Component; Defective Component

    IT WAS REPORTED THAT THE GALT MICROINTRODUCERS IN THE PROVENA KITS ARE PEELING UP, CREATING A LIP WHEN INSERTING. NO PATIENT HARM REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9428470 ↗
  • MalfunctionNov 6, 2019Material Deformation; Material Deformation

    FACILITY REPORTED, "WE TRIED THE WAY WE HAVE BEEN PRACTICING FOR A WHILE NOW WHICH I BELIEVE IS BEST PRACTICE BUT TAKING THE INSIDE OF THE INTRODUCER AND SLIDING IT OVER THE WIRE TO STRETCH IT OUT ENOUGH TO HOPEFULLY NOT HAVE TO MAKE A NICK BUT THE INTRODUCER BUNCHES UP AND STARTS TO SHRED WITH MINIMAL PRESSURE APPLIED. EVEN WITH NICKS WE HAVE BEEN MOST OF THE TIME UNABLE TO EVEN GET THIS INTRODUCER IN AND HAVING TO DROP A BETTER INTRODUCER LIKE THE GALT." IT WAS REPORTED THAT TWO INTRODUCERS HAVE BUNCHED UP AND SHREDDED. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9288053 ↗
  • MalfunctionNov 6, 2019Material Deformation; Material Deformation

    FACILITY REPORTED, "WE TRIED THE WAY WE HAVE BEEN PRACTICING FOR A WHILE NOW WHICH I BELIEVE IS BEST PRACTICE BUT TAKING THE INSIDE OF THE INTRODUCER AND SLIDING IT OVER THE WIRE TO STRETCH IT OUT ENOUGH TO HOPEFULLY NOT HAVE TO MAKE A NICK BUT THE INTRODUCER BUNCHES UP AND STARTS TO SHRED WITH MINIMAL PRESSURE APPLIED. EVEN WITH NICKS WE HAVE BEEN MOST OF THE TIME UNABLE TO EVEN GET THIS INTRODUCER IN AND HAVING TO DROP A BETTER INTRODUCER LIKE THE GALT." IT WAS REPORTED THAT TWO INTRODUCERS HAVE BUNCHED UP AND SHREDDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9288047 ↗
  • MalfunctionJun 8, 2018Hole In Material; Nonstandard Device

    PER MEDWATCH: "FOUND OUT ABOUT THIS ON (B)(6) 2018 FROM THE PICC TEAM. DEFECT OR FAILURE CREATED AN UNSAFE CONDITION." PATIENT WAS BROUGHT BACK TO FACILITY BY PARENTS ON (B)(6) 2017 WITH A HOLE IN THE PICC LINE IN THE LEG EXTENSION ABOVE THE HUB FOR THE NEEDLELESS CONNECTOR. PATIENT HAD TO GO TO SURGERY TO HAVE A BROVIAC PLACED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7585011 ↗
  • MalfunctionApr 26, 2018Leak/Splash; Nonstandard Device

    IT WAS REPORTED BY THE FACILITY THAT THERE WAS A LEAK AT THE HUB NEAR THE INJECTION CAP OF THE DEVICE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7466148 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page