PowerPICC Provena S3274108PDRecalls & reported problems
Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741134326), received between Sep 26, 2017 and Mar 10, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- Device Dislodged or Dislocated2
- Crack1
- Detachment Of Device Component1
- Kinked1
- Leak/Splash1
- Material Deformation1
- Sticking1
Patient problems
- No Known Impact Or Consequence To Patient6
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THE CLINICIAN EXPERIENCED AN ISSUE IN WHICH THE ITEM BROKE WHILE IN USE. THE LINE HAD BEEN SECURED WITH SUTURES. THE LINE WAS PLACED BY AN OUTSIDE FACILITY, AND THE CLINICIAN NOTICED IT WAS ALREADY BROKEN WHEN THE PATIENT ARRIVED. THE PATIENT REQUIRED PLACEMENT OF A NEW LINE, RESULTING IN DELAYS IN MEDICATION DELIVERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24557960 ↗MalfunctionJun 29, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT LEAK NOTED FROM PURPLE PLASTIC HUB BY BEDSIDE NURSE (B)(6) 2023 ADDITIONAL INFORMATION RECEIVED 06/19/2023 PATIENT HAS HISTORY OF AML WHO PRESENTS DUE TO FATIGUE AND HEMOPTYSIS WITH LABS SIGNIFICANT FOR SEVERE THROMBOCYTOPENIA. HE RECEIVED ALL PREVIOUS CARE IN MEXICO BESIDES RECENT ADMISSION AT OSH AND FURTHER RESEARCH WILL NEED TO BE MADE INTO PRIOR TREATMENTS. S/P BRONCHOSCOPY, BIOPSY, LP, INTRATHECAL CYTARABINE. FEVER S/P PROCEDURES ON 6/9, SPIKED FEVER 6/10 AM. CONTINUES CEFEPIME AND LINEZOLID. PLAN TO REPEAT BLOOD AND URINE CULTURES AFTER 24 HOURS SINCE LAST DRAW IF PATIENT REMAINS FEBRILE. PLAN TO REPLACE PICC TOMORROW DUE TO HOLE IN ONE OF THE LINES, FUNCTIONALLY PATIENT ONLY HAS SINGLE LUMEN PICC. THIS EVENT DIRECTLY INVOLVED A PATIENT. IT DID NOT INVOLVE AN URGENT/LIFE THREATENING MEDICAL SITUATION. PATIENT REQUIRED PLACEMENT OF ANOTHER PICC LINE WHICH ALWAYS RUNS THE RISK OF INTRODUCTION INFECTION. THIS RN IS NOT AWARE OF ANY MEDICATION INTERRUPTION. NEW PICC LINE WAS PLACED IMMEDIATELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17234672 ↗MalfunctionJun 28, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT LEAK NOTED TO PURPLE LUMEN BY BEDSIDE NURSE ON (B)(6) 2023 AT 1857. ADDITIONAL INFORMATION RECEIVED 06/19/2023 : PATIENT IS A PREVIOUSLY HEALTHY MALE WITH PANCYTOPENIA AND PERIPHERAL BLASTS ADMITTED TO THE PICU WITH CONTINUED BLEEDING AND PANCYTOPENIA, WORK UP CONSISTENT WITH ACUTE PROMYELOCYTIC LEUKEMIA. INITIATED ON CHEMOTHERAPY (B)(6) 2023 PER AAML1331 PROTOCOL. THIS EVENT DIRECTLY INVOLVED A PATIENT. THE EVENT DID NOT INVOLVE AN URGENT/LIFE THREATENING MEDICAL SITUATION. PATIENT REQUIRED NEW PICC LINE PLACEMENT IMMEDIATELY TO AVOID ANY INTERRUPTION IN CHEMOTHERAPY PLAN. PATIENT REQUIRED AN ADDITIONAL SEDATION EVENT FOR ANOTHER PICC LINE INSERTION. THIS RN DOES IS NOT AWARE OF ANY MEDICATION INTERRUPTION. NEW PICC LINE PLACED AS SOON AS POSSIBLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17221277 ↗MalfunctionApr 17, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE RN REPORTED THAT THEY'VE "HAD A COUPLE 4 FR PICCS THAT HAVE BEEN PLACED AND THEN MALPOSITION (WAS ABLE TO REPOSITION THEM)." THE EXACT QUANTITY OF AFFECTED PICCS WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9971796 ↗MalfunctionApr 17, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THE RN HAD ISSUE WITH ATTEMPT TO REPOSITION LINE BY FLUSHING. RN STATED, "LINE WAS CROSSING MIDLINE AND TERMINATING IN LEFT BRACHIOCEPHALIC. WHEN A TURBULENT FLUSH WAS DONE, THE LINE LOOPED IN THE RIGHT SUBCLAVIAN/AXILLARY VEIN AND COULD NOT GET LINE TO FLUSH AT ALL. I EXCHANGED THE LINE (TO THE 5FR) TO SEND PATIENT HOME WITH LONG TERM ANTIBIOTICS."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9971791 ↗MalfunctionApr 1, 2020Material Deformation; Material Deformation
IT WAS REPORTED THE RN INSERTED THE DILATOR AS NORMAL BUT IT WOULD NOT GO IN ALL THE WAY, THERE WAS RESISTANCE. IT WAS STATED WHEN IT WAS REMOVED THE EDGE WAS FRAYED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9913814 ↗MalfunctionJun 14, 2018Detachment Of Device Component
IT WAS REPORTED THAT THE PICC CATHETER WAS PLACED IN THE (B)(6) FEMORAL VEIN. THE PURPLE END CAME APART FROM THE CATHETER ABOUT A WEEK LATER. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7602611 ↗MalfunctionFeb 2, 2018Kinked; Sticking
IT WAS REPORTED BY THE NURSE THAT DURING CATHETER PLACEMENT SHE PULLED THE GUIDEWIRE TO TRIM THE CATHETER, BUT THE GUIDEWIRE WAS STUCK AND APPEARED TO BE KINKED. SHE WAS ABLE TO PLACE THE CATHETER. WHEN SHE REMOVED THE GUIDEWIRE, SHE STATED IT LOOKED KINKED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7239280 ↗MalfunctionNov 28, 2017Leak/Splash
THE FACILITY REPORTED TO SALES REP THAT THE PICC BEGAN TO LEAK AT THE RED PORT AND EXTENSION LEG CONNECTION. THE PICC WAS REMOVED. NO OTHER INFORMATION IS AVAILABLE AT THIS TIME. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7065447 ↗MalfunctionNov 2, 2017Crack; Fluid/Blood Leak
THE FACILITY REPORTED TO THE SALES REP THAT THE CATHETER CRACKED AND LEAKED. NO OTHER INFORMATION WAS PROVIDED BUT HAS BEEN REQUESTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6998018 ↗MalfunctionSep 26, 2017Fluid/Blood Leak
IT WAS REPORTED BY FACILITY THAT WHEN THE PICC WAS FLUSHED, IT LEAKED AND SPRAYS AT THE JUNCTION OF THE PURPLE LUMEN AND CATHETER LINE (PURPLE LUMEN BELOW THE HUB). THE CLEANING SOLUTION USED ON THE CATHETER WAS CHLORHEXIDINE 2 OR 4 %. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6895077 ↗MalfunctionSep 26, 2017Fluid/Blood Leak
IT WAS REPORTED BY THE FACILITY THAT THE PROVENA CATHETER LUMEN LEAKED. THERE WAS NO REPORT TO WHICH LUMEN LEAKED OR ANY OTHER DETAILS OF THE EVENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6895019 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page