Skip to content

Peripheral Catheter Insertion Kit F122088PTRecalls & reported problems

Bard Access Systems, Inc. · PowerGlide 22Ga, 8cm · Class 2

All models of Peripheral Catheter Insertion Kit (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741140594), received between May 26, 2022 and Mar 15, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Defective Component1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Foreign Body In Patient3
  • Device Embedded In Tissue or Plaque1
  • Needle Stick/Puncture1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2024Defective Component; Defective Component

    IT WAS REPORTED BY CUSTOMER BROKEN ITEM. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18914890 ↗
  • InjuryMar 7, 2024Break; Break

    IT WAS REPORTED BY CUSTOMER ¿ON 2/13/24, AN EXPERIENCE POWER GLIDE MIDLINE NURSE WAS PLACING A 22 GAUGE 8 CM AST MIDLINE INTO A PATIENT'S RIGHT BASILIC VEIN. HE HAD THE NEEDLE INSIDE THE VEIN THE OF THE ANGLE OF THE DEVICE WAS LOW. THE NURSE WAS ABLE TO PUSH THE GUIDEWIRE IN WITH ONE SMOOTH MOTION. HE SLIDE THE CATHETER DOWN INTO THE PATIENT'S ARM WITH EASE. UPON PULLING THE NEEDLE/GUIDEWIRE OUT, THE CATHETER SHEARED APART ABOUT 5MMS BELOW THE HUB OF THE MIDLINE. THE REST OF THE CATHETER WAS BALLED UP ON THE GUIDEWIRE UNDER THE PATIENT'S SKIN. WE HAD TO DO A SMALL INCISION IN THE ARM SO WE COULD GRAB THE CATHETER WITH THE GUIDEWIRE TO ENSURE THAT IT WOULDN'T GET LOST IN THE PATIENT. THIS WAS WITNESSED BY A DOCTOR, BOTH OF US ARE TRAINED IN MIDLINES.¿ NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18857882 ↗
  • InjuryFeb 27, 2024

    DESCRIPTION OF COMPLAINT: IT WAS REPORTED BY CUSTOMER THAT ON (B)(6) 2024. AN EXPERIENCE POWER GLIDE MIDLINE NURSE WAS PLACING A 22 GAUGE 8 CM (LOT NUMBER REJN2430) AST MIDLINE INTO A PATIENT'S RIGHT BASILIC VEIN. HE HAD THE NEEDLE INSIDE THE VEIN THE OF THE ANGLE OF THE DEVICE WAS LOW. THE NURSE WAS ABLE TO PUSH THE GUIDEWIRE IN WITH ONE SMOOTH MOTION. HE SLIDE THE CATHETER DOWN INTO THE PATIENT'S ARM WITH EASE. UPON PULLING THE NEEDLE/GUIDEWIRE OUT, THE CATHETER SHEARED APART ABOUT 5MMS BELOW THE HUB OF THE MIDLINE. THE REST OF THE CATHETER WAS BALLED UP ON THE GUIDEWIRE UNDER THE PATIENT'S SKIN. WE HAD TO DO A SMALL INCISION IN THE ARM SO WE COULD GRAB THE CATHETER WITH THE GUIDEWIRE. TO ENSURE THAT IT WOULDN'T GET LOST IN THE PATIENT. THIS WAS WITNESSED BY TWO HEALTHCARE EMPLOYEES TRAINED IN MIDLINES. FOR LOT NUMBER. ITEM#F122088PT  LOT#REJN2430. NO FOREIGN OBJECT OF CATHETER FRAGMENT NOTED WITH XRAY RESULT. HAVE PRODUCT AND PICTURES BUT THIS PROGRAM WILL NOT ACCEPT IN PDF FORMAT.

    Patient problems: Device Embedded In Tissue or Plaque; Needle Stick/Puncture

    MAUDE report 18809421 ↗
  • InjuryJun 23, 2022Break; Break

    IT WAS REPORTED VIA MEDWATCH "DURING MIDLINE INSERTION, 8CM OF MIDLINE TIP BROKE OFF IN THE RIGHT FOREARM WHEN GUIDEWIRE BECAME STUCK. PT REQUIRED SURGICAL INTERVENTION ON TO REMOVE. IT APPEARS WIRE WAS REMOVED SUCCESSFULLY WITHOUT LONG TERM HARM. IT IS UNCLEAR WHETHER THIS COULD BE RELATED TO TECHNIQUE BY NEW USER OR IS RELATED TO THE DEVICE DIRECTLY."

    Patient problems: Foreign Body In Patient

    MAUDE report 14793586 ↗
  • InjuryMay 26, 2022Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method

    IT WAS REPORTED "PERFECTLY IN THE CENTER OF THE VESSEL. ADVANCED WIRE WITHOUT RESISTANCE AND ADVANCED CATHETER THREADED WITHOUT RESISTANCE AT THE VERY END SLIGHT RESISTANCE. WENT TO SAFETY THE NEEDLE AND RESISTANCE. COULDN¿T PULL BACK. HUB OF CATHETER BROKE FROM CATHETER. HAD TO USE FORCEPS TO GET NEEDLE AND GUIDE WIRE OUT. CATHETER STUCK IN PATIENT FOREARM."

    Patient problems: Foreign Body In Patient

    MAUDE report 14518128 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →