PowerPICC CK000758Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC (118)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741141140), received between Apr 16, 2021 and Apr 16, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Frayed” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2021Material Frayed; Material Frayed
IT WAS REPORTED "ON PICC LINE PLACEMENT, VAS SPECIALIST WAS UNABLE TO GET THE YELLOW WAVEFORM INTRATHORACIC VISABLE ON THE PICC MONITOR. USUAL SUSPECTS WERE CHECKED. SPECIALIST TOOK THE STYLET OUT OF THE PICC AND RECALIBRATED AND THEN REINSERTED THE STYLET INTO THE PICC LINE. AT THIS TIME DURING THE PROCEDURE, THE VAS SPECIALIST NOTICED APPROXIMATELY 4 CM OF THE TIP OF THE STYLET WAS HANGING LOOSE AND WHEN THIS AREA WAS TOUCHED IT BROKE COMPLETELY OFF OF THE END OF THE STYLET. " ADDITIONAL INFORMATION RECEIVED 04/02/2021: PER SPECIALIST: STARTED OUT STANDARD PROCEDURE BUT NOTICED HE DID NOT GET A NORMAL WAIVE FORM. CHECKED CONNECTIONS, FLUSHED CATHETER AGAIN AND WITH NO LUCK HE THEN PULLED THE WIRE BACK COMPLETELY BACK TO RECALIBRATE. AT THAT POINT HE COULD SEE THE TIP OF THE WIRE, NOTICED SOMETHING DID NOT LOOK RIGHT SO WHEN HE TOUCHED THE WIRE TO LOOK CLOSER, HE NOTICED APPROX. 4 CM¿S HANGING OFF OF STYLET. WIRE WAS BROKE. WAS THERE PATIENT HARM REPORTED?: NO BECAUSE WIRE BROKE OUTSIDE OF PT. THEY DID X-RAY JUST TO MAKE CERTAIN AND SHOWED NO SIGNS OF THE WIRE IN THE PT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11680374 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page