PowerPICC CK000821Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC (118)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741152580), received between Aug 5, 2020 and Oct 10, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING CT IV CONTRAST BOLUS EXTRAVASATED INTO LEFT ARM/AXILLA. ON REMOVAL HOLE NOTED AT 10 CM MARK. 4 FR CATHETER WOULD NOT ADVANCE SO WAS EXCHANGED FOR 3 FR CATHETER. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20423458 ↗
  • MalfunctionJun 27, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED PATIENT ARRIVED TO EMERGENCY DEPARTMENT COMPLAINING OF PICC LINE LEAKING; DRESSING TAKEN DOWN & LINE FLUSHED. LEAKING NOTED AT THE JUNCTION OF THE EXTENSION TUBING AND LUMEN. INDENTATION FROM CLAMP NOTED. POSSIBLE OVERUSE OF CLAMP AT LOCATION DISRUPTED TUBING INTEGRITY. NOTHING WAS INFUSING AT TIME OF LEAK DISCOVERY. INSERTED AC, SECURED WITH STATLOCK. PICC D/C'D. PIV ESTABLISHED TO BRIDGE, PICC SCHEDULED TO BE REPLACED 5/30. NO PATIENT HARM. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630638 ↗
  • MalfunctionJul 12, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED VAT WAS CALLED URGENTLY TO ASSESS LEAKING PICC. NOTED HOLE IN THE "PIGTAIL" PORTION OF THE PICC, CLOSE TO THE HARD PLASTIC PIECE THAT CONNECTS TO THE NEEDLELESS CONNECTOR. UNDERWENT AT LEAST 10 ADDITIONAL STICKS UNTIL NEW PICC WAS PLACED BETWEEN IV ACCESS AND LAB DRAWS. DEVICE WAS PLACED 3CM BELOW THE KNEE, AND NEONATAL GRIPLOCK WAS USED TO SECURE. PATIENT WAS BEING TREATED FOR S/P ENTEROPLASTY AND REDO ILEOCOLIC ANASTOMOSIS, AND HAS A HISTORY OF OMPHALOCEALE WITH JEJUNAL ATRESIA. INFUSING: D5 77 NS 77 ACETATE AT 5 CC HR, HEP FLUSH 5/28 18:57. (LAST 7 DAYS REVIEWED) CEFEPIME 40MG Q 8 LAST DOSE 5/26 21:58, VANCOMYCIN 160 MG Q 6 LAST DOSE 5/26 18:19, TPA 5/26 23:36 BY VAT. LINE FLUSHED BUT DID NOT HAVE BLOOD RETURN ONLY ONE DOSE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17312983 ↗
  • MalfunctionApr 21, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT A 3 FR CATHETER PLACED IN FEBRUARY SPRUNG A LEAK AT THE CONNECTION BETWEEN THE HUB AND THE EXTERNAL LUMEN. THE LINE HAD TO BE REPLACED. ADDITIONAL INFORMATION RECEIVED ON 4/12/2023: IT WAS REPORTED THAT ON (B)(6) AT 2030 FLUID WAS NOTED BY BEDSIDE NURSE AND CONFIRMED BY VAT. CATHETER WAS REMOVED. NEW PICC PLACED (B)(6) 2023. PATIENT UNABLE TO RECEIVE CPN FOR 2 DAYS UNTIL NEW LINE PLACED. CATHETER WAS SECURED WITH SECUREACATH. NO PATIENT HARM OR SYMPTOMS RELATED TO THE FLUID LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16792383 ↗
  • MalfunctionFeb 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT, 3FR SV REPORTED TO HAVE HOLE IN IT. PICC FOUND LEAKING WHERE THE PURPLE HUB ATTACHES TO THE EXTERNAL LUMEN. SECURACATH WAS USED AS THE CATHETER SECUREMENT. PATIENT NEEDED ANOTHER PICC LINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16439749 ↗
  • MalfunctionAug 5, 2020Material Frayed; Material Frayed

    IT WAS REPORTED YESTERDAY THE TEAM WAS INSERTING A 3 FR SINGLE LUMEN PICC FOR A 5 YEAR OLD, AT THE END OF THE PROCEDURE WHEN THE GUIDE WIRE WAS BEING REMOVED FROM THE CATHETER, IT WAS NOTED THAT THE GUIDEWIRE WAS FRACTURED APPROXIMATELY 5 CM FROM THE TIP AND WAS DANGLING. THE WIRE HAD BEEN WITHDRAWN AND THE CATHETER WAS TRIMMED PRIOR TO THE PROCEDURE. THE INSERTER IS 100% CERTAIN THAT THE GUIDEWIRE WAS WITHDRAWN WELL PAST THE TRIM SITE. THE INSERTION WAS UNEVENTFUL, HOWEVER THE 3CG WAS NOT READING CORRECTLY AND A CHEST X-RAY WAS OBTAINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10370935 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →