PowerPICC Provena SOLO S1395108D1Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena SOLO (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741154812), received between Sep 13, 2019 and May 20, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Material Deformation1
- Material Frayed1
- Material Twisted/Bent1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2022Material Deformation; Material Deformation
IT WAS REPORTED "THEIR PICC NURSE WAS HAVING ISSUES WITH THE GUIDEWIRE AND IT NOT THREADING INTO THE VESSEL WITH THEIR 5FR TRIPLE LUMEN PICC KIT. SHE WAS ABLE TO GET BLOOD RETURN OUT OF THE NEEDLE BUT WHEN SHE WENT TO ADVANCE THE GUIDEWIRE IT JUST KEPT COILING AROUND IN THE NEEDLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14461994 ↗InjuryJul 9, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED VIA MEDWATCH "PATIENT'S BRACHIAL VESSEL WAS ACCESSED USING THE SAFETY NEEDLE SUPPLIED IN THE BARD PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) KIT. BLOOD RETURN FROM THE NEEDLE WAS POSITIVE AND THE GUIDEWIRE WAS INTRODUCED THROUGH THE NEEDLE. THE NEEDLE WAS THEN REMOVED FROM THE PATIENT'S VESSEL, LEAVING THE GUIDEWIRE IN PLACE. I THEN MADE A TINY INCISION RIGHT ABOVE THE GUIDEWIRE TO ALLOW THE DILATOR/INTRODUCER TO BE PLACED. I THREADED THE GUIDEWIRE INTO THE NEEDLE INTRODUCER AND BEGAN TO APPLY PRESSURE TO ADVANCE THE INTRODUCER INTO THE VESSEL. AS I CONTINUED TO APPLY PRESSURE TO THE INTRODUCER, THE GUIDEWIRE ADVANCED INTO THE VESSEL AS WELL. I REMOVED THE ENTIRE INTRODUCER SINCE I DIDN'T SEE THE WIRE, AND WAS UNABLE TO SEE THE GUIDEWIRE. I APPLIED PRESSURE TO STOP THE BLEEDING AND THEN ULTRASOUND THE PATIENT'S AREA TO CONFIRM THE WIRE WAS IN FACT IN THE PATIENT'S VASCULATURE."
Patient problems: Foreign Body In Patient
MAUDE report 12144192 ↗MalfunctionFeb 9, 2021Material Frayed; Material Twisted/Bent
IT WAS REPORTED PICC PLACED ON RIGHT UPPER EXTREMITY DIFFICULTY GETTING LINE TO DROP PAST SHOULDER, APPEARING TO COIL AT TIP OF CATHETER, SO LINE WAS REMOVED AND INSPECTED. 3CG WIRE WAS NEVER TAKEN OUT AND RE-ADVANCED BACK THROUGH PICC CATHETER DURING PLACEMENT ATTEMPT. 3CG WIRE NOTED TO BE BENT, ALMOST FRACTURED OFF 1CM FROM DISTAL TIP. ADDITIONAL INFORMATION RECEIVED 01/21/2021: WHAT TYPE OF CATHETER SECUREMENT WAS UTILIZED: CATHETER WAS REMOVED AFTER PLACEMENT ATTEMPT AND BENT WIRE NOTED. NO SECUREMENT DEVICE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11305143 ↗MalfunctionSep 13, 2019Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE VALVED HUB BROKE OFF OF THE EXTENSION LEG ON A PICC. 9/05/2019- ADDITIONAL INFORMATION FROM FACILITY STATED, "PATIENT WAS BEING MOVED IN BED VERY LITTLE TENSION ON IV TUBING AND THE CAP DISCONNECTED."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9035890 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page