PowerPICC Provena S1385108D2Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741154867), received between Jan 26, 2022 and Jun 23, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Material Deformation1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hematoma1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THE GREY PART OF THE MICROINTRODUCER WAS BENT AND NURSE COULD NOT FULLY GET INTO PATIENT. ADDITIONAL INFORMATION RECEIVED 03 JUNE 2025: IT WAS REPORTED "DURING A ROUTINE PICC LINE PLACEMENT IN THE UPPER LEFT ARM, THE PROCEDURE PROCEEDED NORMALLY UNTIL THE INTRODUCTION OF THE PEEL-AWAY INTRODUCER. AFTER MAKING A STANDARD SMALL SKIN INCISION AT THE GUIDEWIRE ENTRY SITE, I ATTEMPTED TO ADVANCE THE INTRODUCER OVER THE GUIDEWIRE AND THROUGH THE SUPERFICIAL TISSUE LAYERS. UNEXPECTED RESISTANCE WAS ENCOUNTERED. UPON WITHDRAWAL OF THE INTRODUCER, I OBSERVED THAT THE GRAY OUTER COMPONENT OF THE DEVICE WAS VISIBLY FOLDED AND TORN, COMPROMISING ITS STRUCTURAL INTEGRITY. THIS OCCURRED DESPITE THE USE OF STANDARD TECHNIQUE AND APPROPRIATE INSERTION FORCE. FOLLOWING REMOVAL, A SPHERICAL HEMATOMA APPROXIMATELY THE SIZE OF A SILVER DOLLAR DEVELOPED UNDER THE SKIN, ALONG WITH ACTIVE BLEEDING AT THE ENTRY SITE. A SECOND TRIPLE LUMEN PICC KIT WAS OPENED, AND THE INTRODUCER FROM THAT KIT ADVANCED WITHOUT COMPLICATION THROUGH THE SKIN AND INTO THE SELECTED VESSEL. DESPITE SUCCESSFUL ADVANCEMENT OF THE SECOND INTRODUCER, THE OVERALL PICC ATTEMPT WAS ULTIMATELY DISCONTINUE
Patient problems: Hematoma
MAUDE report 22309640 ↗MalfunctionFeb 8, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT TRIPLE LUMEN CATHETER WAS CRACKED AT THE BASE OF THE LUMEN. WHILE CONDUCTING THE FLUSH ON THE PORT NOTICED THAT IT WAS LEAKING OUT OF THE SIDE PORT. CHANGED THE WIRE OUT. DISPOSED OF IT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18669058 ↗MalfunctionDec 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER "STYLET STOPPER SEEMED TO BE BROKEN AND LEAK WHEN FLUSHING AFTER PLACEMENT" NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15950188 ↗MalfunctionJan 26, 2022Material Frayed; Material Frayed
IT WAS REPORTED "AFTER REMOVING THE WIRE INSIDE THE PICC DURING INSERTION IT WAS RECOGNIZED THAT THE TIP OF THE WIRE WAS BENT AND AFTER CLOSER OBSERVATION THE TIP WAS NOTED TO BE BROKEN OFF."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13359543 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page