Skip to content

PowerPICC Provena S1385108D2Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC Provena (116)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741154867), received between Jan 26, 2022 and Jun 23, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Material Deformation1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Hematoma1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER THE GREY PART OF THE MICROINTRODUCER WAS BENT AND NURSE COULD NOT FULLY GET INTO PATIENT. ADDITIONAL INFORMATION RECEIVED 03 JUNE 2025: IT WAS REPORTED "DURING A ROUTINE PICC LINE PLACEMENT IN THE UPPER LEFT ARM, THE PROCEDURE PROCEEDED NORMALLY UNTIL THE INTRODUCTION OF THE PEEL-AWAY INTRODUCER. AFTER MAKING A STANDARD SMALL SKIN INCISION AT THE GUIDEWIRE ENTRY SITE, I ATTEMPTED TO ADVANCE THE INTRODUCER OVER THE GUIDEWIRE AND THROUGH THE SUPERFICIAL TISSUE LAYERS. UNEXPECTED RESISTANCE WAS ENCOUNTERED. UPON WITHDRAWAL OF THE INTRODUCER, I OBSERVED THAT THE GRAY OUTER COMPONENT OF THE DEVICE WAS VISIBLY FOLDED AND TORN, COMPROMISING ITS STRUCTURAL INTEGRITY. THIS OCCURRED DESPITE THE USE OF STANDARD TECHNIQUE AND APPROPRIATE INSERTION FORCE. FOLLOWING REMOVAL, A SPHERICAL HEMATOMA APPROXIMATELY THE SIZE OF A SILVER DOLLAR DEVELOPED UNDER THE SKIN, ALONG WITH ACTIVE BLEEDING AT THE ENTRY SITE. A SECOND TRIPLE LUMEN PICC KIT WAS OPENED, AND THE INTRODUCER FROM THAT KIT ADVANCED WITHOUT COMPLICATION THROUGH THE SKIN AND INTO THE SELECTED VESSEL. DESPITE SUCCESSFUL ADVANCEMENT OF THE SECOND INTRODUCER, THE OVERALL PICC ATTEMPT WAS ULTIMATELY DISCONTINUE

    Patient problems: Hematoma

    MAUDE report 22309640 ↗
  • MalfunctionFeb 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TRIPLE LUMEN CATHETER WAS CRACKED AT THE BASE OF THE LUMEN. WHILE CONDUCTING THE FLUSH ON THE PORT NOTICED THAT IT WAS LEAKING OUT OF THE SIDE PORT. CHANGED THE WIRE OUT. DISPOSED OF IT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669058 ↗
  • MalfunctionDec 9, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "STYLET STOPPER SEEMED TO BE BROKEN AND LEAK WHEN FLUSHING AFTER PLACEMENT" NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15950188 ↗
  • MalfunctionJan 26, 2022Material Frayed; Material Frayed

    IT WAS REPORTED "AFTER REMOVING THE WIRE INSIDE THE PICC DURING INSERTION IT WAS RECOGNIZED THAT THE TIP OF THE WIRE WAS BENT AND AFTER CLOSER OBSERVATION THE TIP WAS NOTED TO BE BROKEN OFF."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13359543 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →