PowerPICC CK000550ARecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC (118)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741155758), received between Apr 1, 2021 and Aug 20, 2021 — classified as 4 injuries.
The most frequently reported specific device problem is “Material Frayed” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Frayed4
Patient problems
- Foreign Body In Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2021Material Frayed; Material Frayed
IT WAS REPORTED "PATIENT HAD A LOT OF SCAR TISSUE FROM PICCS PLACED IN THE PAST. PATIENT WAS A DIFFICULT ACCESS PATIENT. WHEN ADVANCING PICC THE STYLET GOT STUCK IN THE SCAR TISSUE AND WOULD NOT RETRACT OUT. THE TIP BROKE OFF AND IR HAD TO REMOVE THE STYLET THAT BROKE OFF IN PATIENT." ADDITIONAL INFORMATION RECEIVED 08/03/2021: WAS THERE ANY PATIENT HARM REPORTED? PT HAD TO GO TO THE OR TO HAVE PIECE REMOVED. WHAT THEY'RE BEING TREATED FOR: PICC WAS NEEDED FOR IV ANTIBIOTICS.
Patient problems: Foreign Body In Patient
MAUDE report 12351443 ↗InjuryApr 26, 2021Material Frayed; Material Frayed
IT WAS REPORTED VIA MEDWATCH "FLUOROSCOPIC IMAGING OF RIGHT UPPER ARM SHOWED FOREIGN OBJECT - OBJECT WAS RETRIEVED SUCCESSFULLY BY IR MD - OBJECT APPEARED TO BE PORTION OF GUIDEWIRE (PICC WAS ATTEMPTED IN RIGHT ARM ON (B)(6) 2021)."
Patient problems: Foreign Body In Patient
MAUDE report 11728465 ↗InjuryApr 9, 2021Material Frayed; Material Frayed
IT WAS REPORTED PATIENT HAD TOUGH PICC STICK THAT ENDED UP BEING CUT TO MIDLINES. AFTER PLACEMENT, TIP OF GUIDEWIRE WAS NEEDING TO BE EXTRACTED FROM PATIENT BY RADIOLOGY.
Patient problems: Foreign Body In Patient
MAUDE report 11645325 ↗InjuryApr 1, 2021Material Frayed; Material Frayed
IT WAS REPORTED PATIENT HAD TOUGH PICC STICK THAT ENDED UP BEING CUT TO MIDLINES. AFTER PLACEMENT, TIP OF GUIDEWIRE WERE NEEDING TO BE EXTRACTED FROM PATIENT BY RADIOLOGY. ADDITIONAL INFORMATION RECEIVED 03/19/2021: IT WAS REPORTED THE PATIENT IS BEING TREATED FOR TPN. THE PICC WOULD NOT THREAD TO SVC. MET RESISTANCE. PT REPOSITIONED, ETC TO ATTEMPT TO THREAD TO CENTRAL LOCATION. PICC CUT AT 10 CMS AND LEFT IS PLACE AS A MIDLINE. FLUSHES AND HAS A POSITIVE BLOOD RETURN AT THIS SITE, RN NOTIFIED. LONG TERM TNA USE AND PICC WAS BEING REPLACED. DOES THE PATIENT HAVE AN INFECTIOUS DISEASE?: YES, BEING EVALUATED FOR INTRA-ABDOMINAL ABSCESS.
Patient problems: Foreign Body In Patient
MAUDE report 11602806 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page