PowerPICC Provena CK000291CRecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741161162), received between Aug 8, 2019 and Jan 5, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Fracture2
- Contamination /Decontamination Problem1
- Leak/Splash1
- Material Frayed1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2026Fluid/Blood Leak
IT WAS REPORTED, VERTICAL CRACK IN CATHETER NEAR THE 7CM MARK. ADDITIONAL INFORMATION RECEIVED 12/19/2025: THE PATIENT HAD TO RECEIVE A NEW PICC LINE DUE TO THE MALFUNCTION WITH THE LINE THAT WAS IN PLACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23958395 ↗MalfunctionMay 1, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED HAIR WAS FOUND IN PICC KIT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21942095 ↗MalfunctionAug 18, 2021Material Frayed; Material Frayed
IT WAS REPORTED THE WIRE TIP BROKE OFF AFTER UNSUCCESSFUL PICC PLACEMENT. MET RESISTANCE AT THE 15CM MARK AND PULLED THE WIRE TO CHECK ON INTEGRITY AND IT BROKE OFF ONCE OUTSIDE OF THE PATIENT AND AFTER INTEGRITY CHECK. ADD INFO RCVD 08/11/2021: PLACEMENT INFO: INCIDENT OCCURRED 7/28/2021. INSERTER ATTEMPTED DOUBLE LUMEN POWERPICC PROVENA INSERTION FOR PATIENT NEEDING CHEMOTHERAPY TREATMENTS. ENCOUNTERED RESISTANCE, UNABLE TO PASS INTO CHEST. WHEN PICC REMOVED AND GUIDEWIRE INSPECTED, COPPER TIP OF WIRE FELL OUT OF END OF PICC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12340421 ↗MalfunctionJan 20, 2021Material Twisted/Bent
IT WAS REPORTED UNABLE TO GET WIRE TO CROSS THE SHOULDER DURING INSERTION. THE WIRE WAS PULLED OUT OF THE PATIENT AND A BEND IN THE WIRE WAS NOTICED. ON (B)(6) 2020: DURING DEVICE EVALUATION THE WIRE WAS FOUND TO BE KINKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11205379 ↗MalfunctionOct 13, 2020Fracture
IT WAS REPORTED THE ECG WIRE FRACTURED ON THE 5FR. TRIPLE LUMEN PICC. IT WAS STATED IT WAS AN UNSUCCESSFUL PICC AND DID NOT REACH THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10674146 ↗InjuryOct 13, 2020Fracture
IT WAS REPORTED THE ECG WIRE FRACTURED ON THE 5FR. TRIPLE LUMEN PICC. IT WAS STATED IT WAS A SUCCESSFUL PICC PLACEMENT AND DID REACH THE PATIENT BUT WAS NOT RETRIEVED. DETERMINATION OF SOURCE OF THE WIRE WAS CONFIRMED DUE TO PRE AND POST PLACEMENT CXR.
Patient problems: Foreign Body In Patient
MAUDE report 10672927 ↗MalfunctionJan 24, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT PICC RUPTURED NEAR THE EXTERNAL GREY LUMEN NEAR THE HUB DURING FLUSHING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9627039 ↗MalfunctionAug 8, 2019Leak/Splash
IT WAS REPORTED THAT THE PICC WAS LEAKING. THE NURSE STATED IT WAS NOT LEAKING FROM THE INSERTION SITE. WHEN FACILITY CONTACT ASSESSED THE PICC, SHE FOUND IT WAS LEAKING FROM NEAR WHERE THE GRAY PORT ATTACHES TO THE SUTURE WING. IT WAS STATED THE PICC DID NOT LEAK ALL THE TIME, JUST WHEN THE LINE WAS MOVED IN A CERTAIN POSITION. THE PICC WAS ORIGINALLY PLACED JUST YESTERDAY. RN HAD TO OVERLINE EXCHANGE THE LINE CAUSING THE PT AN EXTRA PROCEDURE AND INFECTION RISK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8875586 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page