PowerPICC CK000875Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC (118)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741178702), received between Jan 4, 2024 and Jan 4, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
Patient problems
- Extravasation1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 4, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED A FRACTURED PICC. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED: "I CAN CONFIRM THIS WAS A PARTIAL FRACTURE OF THE CATHETER."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18447336 ↗InjuryJan 4, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED A FRACTURED PICC. NO OTHER INFORMATION WAS PROVIDED. CUSTOMER RESPONSE RECEIVED 12/18/2023: PATIENT HAD AN EXTRAVASATION INJURY FROM THE PICC FRACTURE. CUSTOMER RESPONSE RECEIVED 12/27/2023: I CAN CONFIRM THIS WAS A PARTIAL FRACTURE OF THE CATHETER.
Patient problems: Extravasation
MAUDE report 18446578 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page