Skip to content

SureStep Intermittent Catheter Tray INTS16Recalls & reported problems

C. R. Bard, Inc. · SureStep Intermittent Catheter Tray · Class 2

All models of SureStep Intermittent Catheter Tray (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741183225), received between May 20, 2022 and Jun 19, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Too Rigid or Stiff” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Class IIOpen, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3117-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Too Rigid or Stiff2
  • Disconnection1
  • Inadequacy of Device Shape and/or Size1
  • Leak/Splash1
  • Restricted Flow rate1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Exposure to Body Fluids1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THERE ARE CONCERNS THAT DUE TO SIZE OF INTERMITTENT CATHETERS IN THE TRAY, THAT MULTIPLE INSERTIONS ON A PATIENT CAUSE TRAUMA. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19566414 ↗
  • MalfunctionAug 3, 2023Restricted Flow rate; Restricted Flow rate

    IT WAS REPORTED THAT THE BAG COULD NOT DROP DOWN FAR ENOUGH FOR THE URINE TO DRAIN. ALSO STATED THAT IT STARTED, THEN STOPPED AND DID NOT DRAIN FURTHER. SO IN THE MEANTIME, THE NURSES ON THAT FLOOR WERE OPENING A BRAND NEW FULL FOLEY KIT AND TAKING THE CATHETER OUT, THEN DISCARDING THE REST. (PCN INTS14), (PCN INTS16). PER FOLLOW UP VIA EMAIL ON 21JUL2023, IT WAS STATED THAT THEY HAD WORKED WITH THE REPRESENTATIVE ON EDUCATION FOR THE UNIT. CURRENTLY THAT HAD A POSITIVE EFFECT ON STAFF¿S USAGE OF THE PRODUCT. PER FOLLOW UP VIA EMAIL ON 21JUL2023, STATED THAT THE CUSTOMER MADE IT CLEAR THAT THE ISSUE HAS BEEN RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17452470 ↗
  • MalfunctionMar 6, 2023Disconnection; Leak/Splash

    AFTER REMOVING STRAIGHT CATHETER FROM PATIENT, THE STAFF BROUGHT CATHETER TO THE TRASH AND CATHETER SPONTANEOUSLY DISCONNECTED FROM TEAL HUB ON CATHETER BAG, EJECTING URINE ON STAFF'S CHEST AND NECK AREA.

    Patient problems: Exposure to Body Fluids

    MAUDE report 16484891 ↗
  • InjuryMay 20, 2022Material Too Rigid or Stiff; Material Too Rigid or Stiff

    IT WAS REPORTED THAT THE INTERMITTENT CATHETER TRAY WAS STIFFER THAN THE OLDER ONES AND HAD RESULTED IN FOUR PATIENT INJURIES THAT REQUIRED IRRIGATION DUE TO BLEEDING. NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 14454865 ↗
  • MalfunctionMay 20, 2022Material Too Rigid or Stiff; Material Too Rigid or Stiff

    IT WAS REPORTED THAT THE CUSTOMER RECENTLY CONVERTED FROM URETERAL CATHETER TRAY TO INTERMITTENT CATHETER TRAY AND SAID THAT THE INTERMITTENT CATHETER TRAY WAS STIFFER AND HAD RESULTED IN PATIENT INJURY BY PENETRATING THE URETHRA. NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 14454620 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page