SureStep Intermittent Catheter Tray recallZ-3117-2026 · C.R. Bard Inc · initiated Jun 25, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
FDA root cause category: Under Investigation by firm
What the company is doing
On 06/25/2026, an "URGENT Medical Device Recall" Letter was via regional courier services or other electronic means to inform customers that based on customer feedback, BD has identified that the CliniClean Pericare Wipes provided within Urinary Catheter Trays may be potentially contaminated with an unknown non-biological substance. Customer are instructed to: Do not use CliniClean Pericare Wipes contained within the affected products listed in Attachment 2. The CliniClean Pericare Wipes are the only impacted component within the tray and all remaining components can continue to be safely used. Users should source alternative, cleansing wipes for the impacted product lots listed in Attachment 2 and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. If affected CliniClean Pericare Wipes have been used without issue, no additional treatment or clinical follow up is required. 1. For affected lots, remove and destroy wipes within the product package at the point of use and do not use. Source alternative cleansing wipes for all impacted urinary catheter trays listed in Attachment 2. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 3. Post this notice in all locations where inventory of these trays are stored. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification. For Questions - contact: BD Medical Information Services: Available 24/7 Select #2 when prompted 1-800-555-7422 or Email: [email protected] BD Technical Support 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: [email protected]
Recall details
- Recall number
- Z-3117-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 25, 2026
- Posted by the FDA
- Sep 4, 2026
- Recalling firm
- C.R. Bard Inc
- Quantity
- 391720 units/kits
- Product code
- KOD
- FDA submissions
- K000723, K802665
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.
Product description
Products that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF: INTP16 INTS14 INTS14C INTS16 INTS16C PEDINTS06 PEDINTS08 PEDINTS10 UDI code: 00801741183300 10801741183307 00801741183218 10801741183215 00801741183232 10801741183239 00801741183225 10801741183222 00801741183249 10801741183246 00801741183263 10801741183260 00801741183270 10801741183277 00801741183287 10801741183284 Preamendment REF: INTL14 INTL15 INTL16 INTP14 UDI code: 00801741183164 10801741183161 00801741183171 10801741183178 00801741183188 10801741183185 00801741183294 10801741183291 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
SureStep Intermittent Catheter Tray INTP16 / 00801741183300¿ 10801741183307 Lot Numbers: NGKT3672 NGKU1467 NGKU3559 INTS14 / 00801741183218 10801741183215 Lot Numbers: NGKT2630 NGKT2687 NGKT3636 NGKT3673 NGKU0119 NGKU0134 NGKU1412 NGKU1439 NGKU1494 NGKU2451 NGKU2487 NGKU2510 INTS14C / 00801741183232 10801741183239 Lot Numbers: NGKT2660 NGKU0149 NGKU1468 INTS16 / 00801741183225 10801741183222 Lot Numbers: NGKT2688 NGKT3587 NGKU0135 NGKU2423 INTS16C / 00801741183249 10801741183246 Lot Numbers: NGKT2631 NGKU0120 NGKU1413 PEDINTS06 / 00801741183263¿ 10801741183260 Lot Number: NGKU2452 PEDINTS08 / 00801741183270 10801741183277 Lot Numbers: NGKT3637 NGKU0166 NGKU1495 PEDINTS10 / 00801741183287¿ 10801741183284 Lot Number: NGKT3674 Pre-amendment INTL14 / 00801741183164 10801741183161 Lot Numbers: NGKT3670 NGKU0132 NGKU1491 INTL15 / 00801741183171 10801741183178 Lot Numbers: NGKT2658 NGKT3584 NGKT3609 NGKU0116 NGKU0118 NGKU0133 NGKU0147 NGKU0163 NGKU1411 NGKU1436 NGKU1437 NGKU1465 NGKU1466 NGKU1492 NGKU2421 NGKU2449 NGKU2450 NGKU2485 NGKU2486 NGKU2508 NGKU2509 NGKU4598 NGKU4601 INTL16 / 00801741183188 10801741183185 Lot Number: NGKU3558 INTP14 / 00801741183294 10801741183291 Lot Numbers: NGKT2659 NGKT3585 NGKT3610 NGKT3635 NGKU0148 NGKU1438 NGKU3567
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- SureStep Intermittent Catheter Tray INTS14CC. R. Bard, Inc. · CATHETER, UROLOGICAL2 recalls4 reports
- SureStep Intermittent Catheter Tray INTS16CC. R. Bard, Inc. · CATHETER, UROLOGICAL2 recalls2 reports
- SureStep Intermittent Catheter Tray INTP14C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall44 reports
- SureStep Intermittent Catheter Tray INTS14C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall31 reports
- SureStep Intermittent Catheter Tray INTL15C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall9 reports
- SureStep Intermittent Catheter Tray INTS16C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall5 reports
- SureStep Intermittent Catheter Tray INTP16C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall5 reports
- SureStep Intermittent Catheter Tray INTL14C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall2 reports
- SureStep Intermittent Catheter Tray PEDINTS06C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall1 report
- SureStep Intermittent Catheter Tray INTL16C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall0 reports
- SureStep Intermittent Catheter Tray PEDINTS08C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall0 reports
- SureStep Intermittent Catheter Tray PEDINTS10C. R. Bard, Inc. · CATHETER, UROLOGICAL1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98840).
- Z-3114-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801 10801741024808 00801741024818 10801741024815 008…
- Z-3115-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: Bard Foley Tray With Lubri-Sil All-Silicone Foley Catheter REF: 947314 947316 947318 UDI codes: 00801741024924 10801741024921 00801741024…
- Z-3116-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: *Bardex I.C. Complete Care Foley Catheter Tray REF: 300316A UDI codes: 00801741024344 10801741024341 *Bardex I.C. Complete Care Foley Ca…
- Z-3118-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067 10801741121064 SureStep" Foley Tray System Comple…
- Z-3119-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: (Canada Only) SureStep" Foley Tray System Bardex IC Complete Care REF: A300314AC UDI-DI code: 00801741143243 SureStep" Foley Tray System …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page