Skip to content

SureStep Intermittent Catheter Tray INTS14CRecalls & reported problems

C. R. Bard, Inc. · SureStep Intermittent Catheter Tray · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741183232), received between Jul 7, 2021 and Jul 24, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Open, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3117-2026 · Root cause: Under Investigation by firm · FDA record ↗

  2. Jan 27, 2022Terminated

    There is a potential to leak from the catheter where it connects to the drain bag

    Z-0666-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 24, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A STRAIGHT CATHETER KIT WAS IDENTIFIED WITH AN UNKNOWN ITEM/MATERIAL IN IT. CUSTOMER HAVE NOT HEARD OF THIS OCCURRENCE WITH ANY OTHER STRAIGHT CATHETER KITS. NO MEDICAL INTERVENTION AND NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22588907 ↗
  • MalfunctionSep 23, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT SURESTEP INTERMITTENT CATHETER KITS WERE LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519372 ↗
  • MalfunctionJul 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT STRAIGHT CATHETER KITS WERE LEAKING AT THE CONNECTION SITE TO THE DRAIN BAG. PER FOLLOW-UP ON 17JUN21, IT WAS REPORTED THAT TWO STAFF HAD BROUGHT THIS FORWARD. IT WAS STATED THAT ONE STAFF INSERTED THE STRAIGHT CATHETER IN TWO PATIENTS, AND IT LEAKED BOTH TIMES. PER FOLLOW-UP ON 18JUN21, IT WAS REPORTED THAT THE CATHETER THAT LEAKED BELONGED TO LOT# NGEWX143 AND LOT#NGFQ0368. IT WAS ALSO REPORTED THAT IT LOOKED LIKE MORE THAN ONE LOT# WAS LEAKING. IT WAS UNKNOWN WHERE THE LEAK OCCURRED. PER FOLLOW-UP ON 21JUN21, IT WAS STATED THAT AT LEAST 6 LEAKING EVENTS HAVE TAKEN PLACE AS OF LAST WEEK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12164267 ↗
  • MalfunctionJul 7, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE STRAIGHT CATHETER KITS WERE LEAKING AT THE CONNECTION SITE TO THE DRAIN BAG. PER FOLLOW UP ON (B)(6) 2021 IT WAS STATED THAT TWO STAFF MEMBERS HAD BROUGHT THIS FORWARD. IT WAS STATED THAT ONE STAFF INSERTED THE STRAIGHT CATHETER IN TWO PATIENTS AND IT LEAKED BOTH THE TIMES. PER FOLLOW UP ON (B)(6) 2021 IT WAS STATED THAT THE CATHETER THAT LEAKED BELONGED TO LOT NGEWX143 AND LOT#NGFQ0368. IT WAS ALSO REPORTED THAT IT LOOKS LIKE MORE THAN ONE LOT WAS LEAKING. IT WAS UNKNOWN WHERE THE LEAKAGE WAS OCCURRING. PER FOLLOW UP ON (B)(2021 IT WAS STATED THAT AT LEAST 6 LEAKING EVENTS HAVE TAKEN PLACE AS OF LAST WEEK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12127721 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.