Skip to content

SureStep Intermittent Catheter Tray INTS16CRecalls & reported problems

C. R. Bard, Inc. · SureStep Intermittent Catheter Tray · Class 2

All models of SureStep Intermittent Catheter Tray (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741183249), received between Jan 30, 2026 and Aug 4, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Open, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3117-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jan 27, 2022Terminated

    There is a potential to leak from the catheter where it connects to the drain bag

    Z-0667-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Urinary Retention1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2026Disconnection

    It was reported that during a straight in and out catheterization procedure, the coude catheter tubing became disconnected from the collection bag. The device was reported to be intact prior to insertion. Urine initially drained into the collection bag, however, while the user was holding the coude tubing, the tubing separated from the bag and urine drainage stopped. Approximately 300 mL of urine had collected in the bag at the time of disconnection. A subsequent bladder scan indicated greater than 600 mL of urine remained in the patient's bladder. Photographs of the damaged tubing were reportedly taken. No patient injury was reported.

    Patient problems: Urinary Retention

    MAUDE report 50131780 ↗
  • MalfunctionJan 30, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE SURESTEP INTERMITTENT CATHETER HAD KINKED AT THE END AND THE SURESTEP INTERMITTENT CATHETER HAD NO OPEN HOLE AT THE END OF THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24228792 ↗