Skip to content

SureStep Intermittent Catheter Tray INTAAIRecalls & reported problems

C. R. Bard, Inc. · SureStep Intermittent Catheter Tray · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741183317), received between Nov 22, 2023 and Nov 22, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Inaccurate Flow Rate” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Open, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3118-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Inaccurate Flow Rate1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 22, 2023Inaccurate Flow Rate; Inaccurate Flow Rate

    IT WAS REPORTED THAT URINE STAYED IN THE INTERMITTENT CATHETER TUBE DURING STRAIGHT CATHETERIZATION THEREFORE CUSTOMER COULD NOT TELL WHEN THEIR BLADDER WAS DONE DRAINING AS URINE REMAINED IN THE TUBE. BLADDER SCANNED PATIENT FOR 900CC AND GOT OUT 600CC OF URINE. WAITED FOR PROLONGED AMOUNT OF TIME TO SEE IF ANYMORE URINE WOULD DRAIN BUT COULD NOT TELL WHEN BLADDER WAS FINISHED DRAINING SINCE URINE SAT STAGNANT IN THE FOLEY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18192804 ↗

Other Catheter care tray devices

All Catheter care tray devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.