Skip to content

Medline DYND10407Recalls & reported problems

Medline Industries, Inc. · TRAY,URETHRAL,PRE-CONN,RED RUBBER,15FR · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10080196990740), received between Sep 28, 2026 and Sep 28, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 4, 2026Class IIOpen, Classified

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

    Z-2481-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Tear, Rip or Hole in Device Packaging1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916002 ↗

FDA clearance and classification

The FDA regulates product code OHR (catheter care tray) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Medline DYND10407

Has the Medline DYND10407 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2481-2026 (Class II), began on May 4, 2026: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Medline DYND10407?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 28, 2026. That report lists device problems “Tear, Rip or Hole in Device Packaging” (1) and “Unsealed Device Packaging” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Medline DYND10407?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Medline DYND10407, and is it still sold?

The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline, model DYND10407). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter care tray devices

All Catheter care tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page