GlidePath 5303100Recalls & reported problems
Bard Peripheral Vascular, Inc. · GlidePath 10F Long-Term Dialysis Catheter, 10 cm · Class 2 · Implant
All models of GlidePath (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741187285), received between Nov 20, 2024 and Nov 14, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Deformation Due to Compressive Stress2
- Restricted Flow rate2
- Air/Gas in Device1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 14, 2025Fracture; Loose or Intermittent Connection; Air/Gas in Device; Fracture; Loose or Intermittent Connection; Air/Gas in Device
¿ON (B)(6), A PATIENT UNDERWENT A DIALYSIS CATHETER PLACEMENT PROCEDURE USING A GLIDEPATH HEMODIALYSIS CATHETER. ON (B)(6) 2025, FOLLOWING INITIATION OF DIALYSIS TREATMENT, THE PRESENCE OF AIR BUBBLES WAS OBSERVED IN THE HEMODIALYSIS CIRCUIT, SPECIFICALLY IN THE RED LINE CONNECTED TO THE BLUE LUMEN OF THE GLIDEPATH CATHETER. NO VISIBLE FISSURES WERE INITIALLY IDENTIFIED, AS THE SUSPECTED BREAK WAS LOCATED BENEATH THE PATIENT'S TEGADERM DRESSING. THE CATHETER WAS LATER FOUND TO BE RUPTURED AT THE JUNCTION BETWEEN THE OUTER CATHETER TUBING AND THE PLASTIC HUB. THE PATIENT WAS TRANSFERRED TO INTERVENTIONAL RADIOLOGY FOR AN EMERGENCY EVALUATION OF THE DIALYSIS LINE. AFTER THE DIALYSIS LINE WAS CHANGED AND THE DEFECTIVE CATHETER WAS RETRIEVED, THE CATHETER BREAK WAS CONFIRMED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 23551408 ↗MalfunctionDec 15, 2024Restricted Flow rate; Fracture; Deformation Due to Compressive Stress; Restricted Flow rate; Fracture; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THREE MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE OUTER PART OF THE LINES ALLEGEDLY TENDS TO REMAIN CRUSHED AT THE CLAMPS, AFFECTING THE LUMEN OF THE CATHETER. IT WAS FURTHER REPORTED THAT IT DID NOT BREAK COMPLETELY, HOWEVER IT LOOKS LIKE IT CAN'T GET BACK TO ITS ORIGINAL ROUND SHAPE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20935848 ↗MalfunctionNov 20, 2024Restricted Flow rate; Fracture; Deformation Due to Compressive Stress; Restricted Flow rate; Fracture; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THREE MONTHS AND NINETEEN DAYS POST DIALYSIS CATHETER PLACEMENT PROCEDURE, THE OUTER PART OF THE LINES ALLEGEDLY TENDS TO REMAIN CRUSHED AT THE CLAMPS, AFFECTING THE LUMEN OF THE CATHETER. IT WAS FURTHER REPORTED THAT IT DID NOT BREAK COMPLETELY, HOWEVER IT LOOKS LIKE IT CANNOT GET BACK TO ITS ORIGINAL ROUND SHAPE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20732757 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page