Skip to content

GlidePath 5303190Recalls & reported problems

Bard Peripheral Vascular, Inc. · GlidePath 10F Long-Term Dialysis Catheter, 19 cm · Class 2 · Implant

All models of GlidePath (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741187315), received between Jul 25, 2024 and Nov 22, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Loss of or Failure to Bond2
  • Material Protrusion/Extrusion2
  • Stretched2
  • Deformation Due to Compressive Stress1
  • Fluid/Blood Leak1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETERS WERE DISLODGING. IT WAS FURTHER REPORTED THAT THE CATHETER PLACED WAS TOO CLOSE TO THE EXIT SITE AND NOT ALLOWED FOR ADEQUATE TISSUE ATTACHMENT. REPORTEDLY, IT HAD AN ISSUE WITH PUTTING TOO MUCH GLUE ON THE CUFFS SO THAT THEY WERE NOT FLUFFY ENOUGH. IT AFFECTS THE TISSUE INGROWTH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20755552 ↗
  • MalfunctionNov 22, 2024Fracture; Stretched; Material Protrusion/Extrusion; Fracture; Stretched; Material Protrusion/Extrusion

    IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER ALLEGEDLY CRACKED AND ALLEGEDLY BALLOONED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20755408 ↗
  • MalfunctionNov 21, 2024Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETERS WERE DISLODGING. IT WAS FURTHER REPORTED THAT THE CATHETER PLACED WAS TOO CLOSE TO THE EXIT SITE AND NOT ALLOWED FOR ADEQUATE TISSUE ATTACHMENT. REPORTEDLY, IT HAD AN ISSUE WITH PUTTING TOO MUCH GLUE ON THE CUFFS SO THAT THEY WERE NOT FLUFFY ENOUGH. IT AFFECTS THE TISSUE INGROWTH. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20744285 ↗
  • MalfunctionNov 21, 2024Fracture; Stretched; Material Protrusion/Extrusion; Fracture; Stretched; Material Protrusion/Extrusion

    IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER ALLEGEDLY CRACKED AND ALLEGEDLY BALLOONED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20744088 ↗
  • MalfunctionJul 25, 2024Fluid/Blood Leak; Fracture; Material Separation; Deformation Due to Compressive Stress; Fluid/Blood Leak; Fracture; Material Separation; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT EIGHT MONTHS AND THIRTY DAYS POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE PHYSICIAN FOUND THAT THERE WAS ALLEGEDLY A LEAK AT THE CONNECTION BETWEEN THE HUB AND THE CATHETER. FURTHERMORE, KINK WAS ALSO FOUND BETWEEN THE HUB AND CATHETER. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19834322 ↗

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page