GlidePath 5343190Recalls & reported problems

Bard Peripheral Vascular, Inc. · GlidePath 13F Long-Term Hemodialysis Catheter Standard Kit, 19 cm · Class 2 · Implant

All models of GlidePath (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741187391), received between Jul 31, 2024 and Jan 14, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Fracture2
  • Loss of or Failure to Bond2
  • Connection Problem1
  • Contamination /Decontamination Problem1
  • Detachment of Device or Device Component1
  • Expulsion1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 14, 2026Suction Problem; Malposition of Device

    ON (B)(6) 2025, A PATIENT UNDERWENT FOR A DIALYSIS CATHETER PLACEMENT PROCEDURE USING GLIDEPATH. DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY POSITIONED SHALLOWER THAN INTENDED, RESULTING IN POOR BLOOD WITHDRAWAL. DUE TO INSUFFICIENT DEPTH AND INADEQUATE BLOOD RETURN, THE CATHETER WAS REMOVED AND REPLACED WITH A LONGER SIZE CATHETER TO ACHIEVE PROPER PLACEMENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064967 ↗
  • InjuryNov 25, 2025Loss of or Failure to Bond; Material Separation; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Material Separation; Improper or Incorrect Procedure or Method

    ON NOV 04, 2025, IT WAS REPORTED THAT AFTER REMOVING A 13F X 19CM GLIDEPATH CATHETER FROM A PATIENT, THE CUFF WAS FOUND TO BE DETACHED FROM THE CATHETER. THE CUFF WAS SUBSEQUENTLY RETRIEVED FROM THE SOFT TISSUE, AND IT WAS NOTED THAT BOTH THE CUFF AND A SLEEVE PORTION OF THE CATHETER HAD SEPARATED FROM THE ORIGINAL DEVICE. FURTHER PROBING AND DISSECTION WITH INSTRUMENTS AND ULTRASOUND GUIDANCE USED TO VISUALLY SEE AND REMOVE THE CUFF PORTION. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23641577 ↗
  • MalfunctionAug 25, 2025Loss of or Failure to Bond; Expulsion; Loss of or Failure to Bond; Expulsion

    ON (B)(6) 2025 A PATIENT UNDERWENT A DIALYSIS CATHETER PLACEMENT BY USING GLIDEPATH. DURING THE PROCEDURE, THE CUFF ALLEGEDLY SLIPPED OFF WHILE MANIPULATING THE POSITION. IT WAS FURTHER REPORTED THAT IT WAS REPOSITIONED WITH FLUOROSCOPY AND NOTICED THAT THE CATHETER WAS MOVING BUT THE CUFF REMAINED IN PLACE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882331 ↗
  • MalfunctionApr 1, 2025Contamination /Decontamination Problem; Connection Problem; Contamination /Decontamination Problem; Connection Problem

    IT WAS REPORTED THAT PRIOR TO A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE TUNNELER AND THE CATHETER CONNECTION WAS ALLEGEDLY NOT MADE WELL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21737950 ↗
  • MalfunctionMar 5, 2025Fracture; Material Separation; Detachment of Device or Device Component; Fracture; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE CONNECTION OF THE CUFF ALLEGEDLY CAME OFF WHEN THE PROTECTIVE CAP OF THE TUNNELER WAS REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21527159 ↗
  • MalfunctionJul 31, 2024Fracture; Material Separation; Fracture; Material Separation

    IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE PIECE OF PLASTIC ALLEGEDLY BROKE OFF OF CATHETER UPON REMOVAL. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED PAIN. FURTHERMORE, THERE IS NO INFORMATION REGARDING MEDICAL INTERVENTION NOR MEDICATIONS PRESCRIBED TO TREAT PAIN. REPORTEDLY, THE CATHETER WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19875207 ↗

Other Catheter, hemodialysis, implanted devices