Crosser iQ XCTO146Recalls & reported problems
Bard Peripheral Vascular, Inc. · Crosser iQ Ultrasonic CTO Device · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Crosser iQ catheter is an ultrasonic device used to open fully blocked arteries. Its 154 reports mostly describe parts separating, such as tips or marker bands detaching, and core wires stretching. The reports include no deaths and 9 injuries, and 145 are malfunctions. No recalls name this device. Reports rose from 50 in the year before to 81 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 144
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 144 adverse event reports that list this exact device (DI 00801741201332), received between Feb 25, 2022 and Sep 29, 2026 — classified as 9 injuries, 135 malfunctions.
The most frequently reported specific device problem is “Material Separation” (80 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation80
- Detachment of Device or Device Component47
- Device-Device Incompatibility28
- Material Deformation25
- Peeled/Delaminated14
- Failure to Advance13
- Material Split, Cut or Torn13
- Retraction Problem12
Patient problems
- No Clinical Signs, Symptoms or Conditions135
- Foreign Body In Patient7
- Perforation of Vessels3
Recent adverse event reports
The 25 most recent of 144 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Failure to Advance; Detachment of Device or Device Component
On (b)(6) 2026, a patient underwent a recanalization procedure using the Crosser iQ catheter for treatment of a calcified lesion. During the procedure, approximately four minutes into use, it became difficult to cross the lesion. Upon pulling the device back, the tip of the device detached. As the patient is not fit for open surgery at that stage, the tip remained embedded in the calcified vessel. The procedure was not completed. The surgeon is following up with the patient for another surgery. The patient was later discharged.
Patient problems: Foreign Body In Patient
MAUDE report 50928868 ↗MalfunctionSep 15, 2026Material Separation
On (b)(6) 2026, a patient underwent a recanalization procedure via contralateral approach using the catheter. During the procedure, material separation was observed at the catheter tip approximately 5 minutes of use but was still attached to the corewire. The procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50770848 ↗MalfunctionAug 13, 2026Material Separation; Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50233643 ↗MalfunctionJun 23, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, THE TIP ALLEGEDLY DETACHED AFTER APPROXIMATELY THREE MINUTES, BUT IT WAS STILL ATTACHED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25589726 ↗MalfunctionJun 20, 2026Material Separation; Failure to Advance
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, ON FIRST ATTEMPT THE CROSSER WAS UNABLE TO CROSS THE LESION AND ON SECOND ATTEMPT THE TIP SEPARATED FROM THE COREWIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25569097 ↗MalfunctionJun 20, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING A CROSSER IQ CATHETER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CATHETER TIP ALLEGEDLY BROKE FROM THE SHAFT AND WAS OBSERVED TO BE DETACHED BUT NOT COMPLETELY SEPARATED. THE TIP REMAINED ATTACHED TO THE CORE WIRE, AND NO FRAGMENT WAS LEFT BEHIND IN THE PATIENT'S BODY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25568958 ↗MalfunctionJun 9, 2026Material Separation; Improper or Incorrect Procedure or Method
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25441809 ↗MalfunctionJun 8, 2026Break; Material Separation; Improper or Incorrect Procedure or Method; Device-Device Incompatibility; Material Deformation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PROCEDURE, AFTER PASSING THROUGH THE WIRE VASSALLO SUPPORT IN A RIGHT CFA CALCIFIED LESION VIA A LEFT CONTRALATERAL APPROACH, THE PATIENT ATTEMPTED TO INSERT THE WIRE INTO THE CROSSER IQ, BUT THE GUIDEWIRE WAS RELUCTANT TO GO IN AT ALL, AND THERE WAS NO WAY TO GET THE WIRE CAUGHT IN THE GUIDEWIRE PORT AND NO WAY TO GET THE WIRE OUT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25423276 ↗MalfunctionJun 1, 2026Material Separation; Device-Device Incompatibility; Material Deformation
ON (B)(6) 2026, THE PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. PRIOR TO THE PROCEDURE, THE GUIDEWIRE ALLEGEDLY FAILED TO ADVANCE OVER THE GUIDE WIRE LUMEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25345885 ↗MalfunctionMay 28, 2026Break; Device-Device Incompatibility; Material Deformation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE GUIDEWIRE ALLEGEDLY DID NOT PASS THROUGH THE EXIT PORT OF THE CATHETER AND STUCK. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25312345 ↗MalfunctionMay 26, 2026Material Separation
A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER WAS DISCONTINUED DUE TO POSSIBLE DAMAGE TO THE TIP AND THE END POINT. IT WAS LATER CONFIRMED THAT THE TIP, END POINT AND SHAFT WERE DETACHED FROM THE ADHESIVE, AND NO COMPLETE SEPARATION WAS OBSERVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273714 ↗MalfunctionMay 22, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED BUT IT WAS STILL CONNECTED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25251439 ↗MalfunctionMay 20, 2026Material Split, Cut or Torn
ON (B)(6) 2026, THE PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER SHAFT WAS ALLEGEDLY SPLINTERED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25226390 ↗MalfunctionMay 18, 2026Fracture; Failure to Advance
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE GUIDEWIRE THE TIP OF THE GUIDEWIRE SNAPPED AS THE DEVICE UNABLE TO PASS THROUGH THE LESION. THE GUIDEWIRE WAS RETRIEVED USING A SNARE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25197369 ↗MalfunctionMay 13, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE TIP ALLEGEDLY DETACHED AND STILL ATTACHED TO THE COREWIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25157300 ↗MalfunctionMay 8, 2026Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25123098 ↗MalfunctionMay 8, 2026Break
ON (B)(6) 2026, A PATIENT UNDERWENT RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. PRIOR TO THE PROCEDURE, THE AIR PREVENTION VALVE WAS ALLEGEDLY DAMAGED. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25121242 ↗MalfunctionMay 8, 2026Material Separation; Failure to Advance; Detachment of Device or Device Component
ON ()(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE VIA IPSILATERAL ANTEGRADE APPROACH USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE DEVICE WAS INACCESSIBLE TO THE ATA. FURTHER IT WAS REPORTED THAT THE TIP DETACHED AFTER FIFTY-FIVE SECONDS OF USE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25120687 ↗MalfunctionMay 6, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE BY IPSILATERAL APPROACH USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED, BUT IT WAS STILL ATTACHED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25090309 ↗MalfunctionMay 6, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER TO TREAT AN IPSILATERAL CIRCUMFERENTIAL PTA CTO LESION. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED AND STILL REMAINS CONNECTED TO CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25090307 ↗MalfunctionMay 6, 2026Material Separation
ON (b)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE IN THE SFA USING IPSILATERAL APPROACH BY USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED BUT STILL ATTACHED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25090103 ↗MalfunctionMay 4, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED BUT STILL ATTACHED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25075893 ↗MalfunctionMay 4, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE TIP OF THE CATHETER ALLEGEDLY DETACHED BUT STILL REMAIN ATTACHED TO THE COREWIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25066008 ↗MalfunctionApr 30, 2026Material Separation; Material Deformation
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, IN A POP CASE, THE CATHETER TIP WAS ALLEGEDLY DETACHED AFTER FIFTEEN SECONDS PASSING THROUGH THE GUIDEWIRE BUT STILL ATTACHED TO THE CORE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25039182 ↗MalfunctionApr 29, 2026Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER IQ CATHETER. DURING THE PROCEDURE, THE CATHETER TIP WAS ALLEGEDLY DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25030549 ↗
FDA clearance and classification
The FDA regulates product code PDU (catheter for crossing total occlusions) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Crosser iQ XCTO146
Has the Crosser iQ XCTO146 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741201332), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Crosser iQ XCTO146?
The FDA's MAUDE database lists 144 adverse event reports for this device, received between Feb 25, 2022 and Sep 29, 2026. In those reports, the most common specific device problems are “Material Separation” (80), “Detachment of Device or Device Component” (47) and “Device-Device Incompatibility” (28); the most common patient problems are “Foreign Body In Patient” (7) and “Perforation of Vessels” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Crosser iQ XCTO146?
None of the reports are classified as deaths, 9 as injuries and 135 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Crosser iQ XCTO146, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand Crosser iQ, model XCTO146). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter for crossing total occlusions devices
- Approach G50791Cook Incorporated · Catheter for crossing total occlusions2 recalls15 reports
- Approach G50792Cook Incorporated · Catheter for crossing total occlusions2 recalls7 reports
- Approach G50787Cook Incorporated · Catheter for crossing total occlusions2 recalls2 reports
- Approach G50786Cook Incorporated · Catheter for crossing total occlusions2 recalls1 report
- Approach G50788Cook Incorporated · Catheter for crossing total occlusions2 recalls1 report
- Approach G50789Cook Incorporated · Catheter for crossing total occlusions2 recalls0 reports
- OUTBACK OTB59120ACordis US Corp. · Catheter for crossing total occlusions82 reports
- Approach G50793Cook Incorporated · Catheter for crossing total occlusions1 recall17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.