Approach G50791Recalls & reported problems
Cook Incorporated · Approach, CTO-12 Micro Wire Guide · Class 2
All models of Approach (11)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00827002507913), received between Mar 7, 2016 and Sep 14, 2021 — classified as 2 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Material Separation” (6 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 5, 2026Class IIOpen, Classified
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Z-1776-2026 · Root cause: Incorrect or no expiration date · Recall details · FDA record ↗
- Oct 8, 2024Class IIOpen, Classified
Affected device lots have labels that state the incorrect expiration dates.
Z-0346-2025 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation6
- Sticking2
- Stretched2
- Uncoiled2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
- Unraveled Material1
Patient problems
- No Known Impact Or Consequence To Patient10
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Sepsis1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2021Unraveled Material; Unraveled Material
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12469400 ↗InjuryMar 19, 2021Material Separation; Material Separation
NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11529266 ↗InjuryMay 10, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Sepsis
MAUDE report 8598105 ↗MalfunctionMar 28, 2019Stretched; Stretched
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8459758 ↗MalfunctionMar 25, 2019Stretched; Stretched
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8448404 ↗MalfunctionAug 11, 2017Sticking; Sticking
IT WAS REPORTED THAT DURING A LOWER LEG ANGIOGRAM, AN APPROACH CTO MICROWIRE WIRE GUIDE BECAME STUCK IN THE CATHETER WHEN REMOVING. THE SAME PROBLEM WAS REPORTED WITH TWO OTHER WIRE GUIDES. EACH DEVICE IS BEING REPORTED AND DOCUMENTED IN THE FOLLOWING REPORT NUMBERS: 1820334-2017-02072, 1820334-2017-02073, AND 1820334-2017-02114. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6789153 ↗MalfunctionJul 24, 2017Sticking; Sticking
IT WAS REPORTED THAT DURING A LOWER LEG ANGIOGRAM, AN APPROACH CTO MICROWIRE WIRE GUIDE BECAME STUCK IN THE CATHETER WHEN REMOVING. THE SAME PROBLEM WAS REPORTED WITH TWO OTHER WIRE GUIDES. EACH DEVICE IS BEING REPORTED AND DOCUMENTED IN THE FOLLOWING REPORT NUMBERS: 1820334-2017-02073, 1820334-2017-02114, AND 1820334-2017-02115. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6737702 ↗MalfunctionJun 13, 2017Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING A PTA PROCEDURE ON A (B)(6) YEAR OLD MALE PATIENT, THEY WERE UNABLE TO REMOVE THE APPROACH CTO MICROWIRE WIRE GUIDE FROM ANOTHER MANUFACTURER'S CATHETER. THE PATIENT'S ANATOMY WAS VERY CALCIFIED AND THE APPROACH CTO MICROWIRE WIRE GUIDE WAS INSERTED WITH SOME FRICTION INITIALLY. WHEN THEY ATTEMPTED TO REMOVE THE WIRE, IT WAS FIXED IN ANOTHER MANUFACTURER'S CATHETER. AS A RESULT, THEY REMOVED THE SUPPORT CATHETER WITH THE WIRE AND COMPLETED THE CASE. AFTER THE PROCEDURE, THEY ATTEMPTED TO PUSH THE WIRE OUT OF THE OTHER MANUFACTURER'S CATHETER BACKWARDS BUT THEY WERE UNSUCCESSFUL. THEY WERE ONLY ABLE TO PUSH THE WIRE FORWARD. DUE TO THE AMOUNT OF CALCIFICATION, THE PATIENT WAS TAKEN TO SURGERY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6638609 ↗MalfunctionMay 11, 2017Uncoiled; Uncoiled
DURING A LOWER LEG PERIPHERAL ANGIOGRAM, THE TIP OF THE APPROACH CTO MICROWIRE WIRE GUIDE UNCOILED (1820334-2017-00946). THE PHYSICIAN OPENED A SECOND APPROACH CTO MICROWIRE WIRE GUIDE AND IT PROCEEDED TO UNCOIL AS WELL (1820334-2017-00947). THE PHYSICIAN USED A COMPETITOR'S WIRE TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6561699 ↗MalfunctionMay 11, 2017Uncoiled; Uncoiled
DURING A LOWER LEG PERIPHERAL ANGIOGRAM, THE TIP OF THE APPROACH CTO MICROWIRE WIRE GUIDE UNCOILED (1820334-2017-00946). THE PHYSICIAN OPENED A SECOND APPROACH CTO MICROWIRE WIRE GUIDE AND IT PROCEEDED TO UNCOIL AS WELL (1820334-2017-00947). THE PHYSICIAN USED A COMPETITOR'S WIRE TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6561553 ↗MalfunctionApr 6, 2017Material Separation; Material Separation
IT WAS REPORTED THAT DURING A LEFT LOWER EXTREMITY ANGIOGRAM, USING AN APPROACH CTO MICROWAVE WIRE GUIDE, WITHIN TWO MINUTES OF ATTEMPTING TO PASS THE CTO, THE TIP OF TWO APPROACH CTO MICROWIRE GUIDES SEPARATED. THE TIP OF THE WIRES WERE STILL ATTACHED AND ABLE TO BE REMOVED. THE PROCEDURE WAS THEN COMPLETED USING A COMPETITOR'S WIRE. THERE WAS NOTHING LEFT BEHIND AND NO HARM TO THE PATIENT. THIS MEDWATCH IS FOR DEVICE LOT #7611876, PLEASE SEE MEDWATCH WITH G9# 1820334-2017-00553 FOR DEVICE LOT #7638094.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6463361 ↗MalfunctionApr 5, 2017Material Separation; Material Separation
IT WAS REPORTED THAT DURING A LEFT LOWER EXTREMITY ANGIOGRAM, USING AN APPROACH CTO MICROWAVE WIRE GUIDE, WITHIN TWO MINUTES OF ATTEMPTING TO PASS THE CTO, THE TIP OF TWO APPROACH CTO MICROWIRE GUIDES SEPARATED. THE TIP OF THE WIRES WERE STILL ATTACHED AND ABLE TO BE REMOVED. THE PROCEDURE WAS THEN COMPLETED USING A COMPETITOR'S WIRE. THERE WAS NOTHING LEFT BEHIND AND NO HARM TO THE PATIENT. THIS MEDWATCH IS FOR DEVICE LOT #7638094, PLEASE SEE MEDWATCH WITH MFR REPORT# 1820334-2017-00556 FOR DEVICE LOT #7611876.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6461749 ↗MalfunctionMar 10, 2017Material Separation; Material Separation
IT WAS REPORTED, DURING LEFT LEG LOWER EXTREMITY ANGIOGRAM, THE COILING OF THE TIP SEPARATED FROM THE APPROACH CTO MICROWIRE WIRE GUIDE. IT WAS STILL CONNECTED TO WIRE. PER ADDITIONAL INFORMATION, THERE WAS RESISTANCE ENCOUNTERED DURING THE REMOVAL. AS THE WIRE BEING REMOVED THROUGH A CXI CATHETER. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6396590 ↗MalfunctionJan 13, 2017Material Separation; Material Separation
IT WAS REPORTED THE TIP OF THE DEVICE SEPARATED. A NEW DEVICE WAS OPENED AND USED TO COMPLETE THE PROCEDURE. THE DEVICE SEPARATION OCCURRED OUTSIDE OF THE PATIENTS BODY THEREFORE A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT PROVIDED AT THE TIME OF THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6249219 ↗MalfunctionMar 7, 2016Material Separation; Material Separation
THE PHYSICIAN WAS USING THE WIRE WITH A ANOTHER MANUFACTURER'S DEVICE AND CXI SUPPORT CATHETER. UPON ATTEMPTING TO CROSS CTO, THE PHYSICIAN NOTED THE WIRE HAD BROKEN OFF. UPON REMOVING THE DEVICES USED, ALL DEVICES AND DEVICE PORTIONS CAME OUT OF THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT铠BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES NOR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 5483026 ↗
FDA clearance and classification
The FDA regulates product code PDU (catheter for crossing total occlusions) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Approach G50791
Has the Approach G50791 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1776-2026 (Class II), began on Mar 5, 2026: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Approach G50791?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Mar 7, 2016 and Sep 14, 2021. In those reports, the most common specific device problems are “Material Separation” (6), “Sticking” (2) and “Stretched” (2); the most common patient problem is “Sepsis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Approach G50791?
None of the reports are classified as deaths, 2 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Approach G50791, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Approach, model G50791). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter for crossing total occlusions devices
- Crosser iQ XCTO146Bard Peripheral Vascular, Inc. · Catheter for crossing total occlusions144 reports
- Approach G50792Cook Incorporated · Catheter for crossing total occlusions2 recalls7 reports
- Approach G50787Cook Incorporated · Catheter for crossing total occlusions2 recalls2 reports
- Approach G50786Cook Incorporated · Catheter for crossing total occlusions2 recalls1 report
- Approach G50788Cook Incorporated · Catheter for crossing total occlusions2 recalls1 report
- Approach G50789Cook Incorporated · Catheter for crossing total occlusions2 recalls0 reports
- OUTBACK OTB59120ACordis US Corp. · Catheter for crossing total occlusions82 reports
- Approach G50793Cook Incorporated · Catheter for crossing total occlusions1 recall17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page