PowerPICC Provena SOLO CK000409ARecalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena SOLO (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741215070), received between Feb 19, 2021 and Jul 22, 2021 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Deformation4
- Material Frayed2
- Material Twisted/Bent2
- Difficult to Flush1
- No Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2021Material Deformation; Material Deformation
IT WAS REPORTED TWO 4 FRENCH DUAL LUMEN PROVENA PICC¿S, KINKED IN THE AXILLARY REGION OVER THE LAST TWO DAYS. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12211271 ↗MalfunctionJul 22, 2021Material Deformation; Material Deformation
IT WAS REPORTED TWO 4 FRENCH DUAL LUMEN PROVENA PICC¿S, KINKED IN THE AXILLARY REGION OVER THE LAST TWO DAYS. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12211260 ↗MalfunctionMay 13, 2021Material Frayed; Material Frayed
IT WAS REPORTED "ATTEMPTED PICC PLACEMENT - ABLE TO ACCESS VEIN AND THREAD WIRE WITHOUT DIFFICULTY BUT UNABLE TO PASS PICC PAST ~20 CM MARKING. UNABLE TO ADVANCE OR RETRACT PICC OR INNER WIRE. ATTENDING PHYSICIAN AND VASCULAR SURGERY PHYSICIAN NOTIFIED. VASCULAR PHYSICIAN TO BEDSIDE AND ABLE TO REMOVE PICC. PICC CATHETER AND GUIDEWIRE INTACT BUT GUIDEWIRE WAS KINKED AND FRAYED IN SEVERAL AREAS." ADD INFO RCVD D 05/03/2021: PLACEMENT INFO: RIGHT BRACHIAL VEIN. WAS THERE PATIENT HARM REPORTED?: NO. MEDWATCH RECEIVED STATED: PICC CATHETER INTACT, UNABLE TO ADVANCE, CATHETER REMOVED, INNER WIRE WAS KINKED AND FRAYED IN SEVERAL AREAS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821321 ↗MalfunctionApr 15, 2021Material Frayed; Material Frayed
IT WAS REPORTED "THIS EVENING THAT A 3CG WIRE BROKE OFF INSIDE OF A PATIENT DURING A PICC INSERTION. NURSE STATED HE HAD NO RESISTANCE INSERTING BUT DID HAVE TO MAKE TWO PASSES TO GET THE PICC TO DROP. HE ALSO HAD NO RESISTANCE WHEN REMOVING THE WIRE. WHEN HE PULLED THE WIRE, HE NOTICED SOMETHING ¿HANGING¿ OUT OF THE SOLO VALVE WHICH ENDED UP BEING 6 3CG MAGNETS THAT WERE ALL STUCK TOGETHER." "THE PICC WAS REMOVED AND A CT OF THE HEAD, CHEST, ABDOMEN AND PELVIS SHOWED NO EVIDENCE OF FOREIGN BODIES. THE PATIENT HAS HAD NO SYMPTOMS, AND A NEW PICC WAS PLACED IN IR." "UPON INSPECTION OF THE WIRE IT HAD BROKEN, SOME DETAILS FROM FACILITY PRODUCT INSPECTION:" 3 MAGNETS IN TACT IN THE WIRE STILL. 6 MAGNETS PULLED OUT OF THE VALVE AND ACCOUNTED FOR. KINK THAT CAN BE SEEN IN THE WIRE HAPPENED DURING TRANSPORT, THEY SHOVED IT IN A BAG, BUT IT WAS STRAIGHT WHEN IT WAS REMOVED FROM THE PATIENT. ABOUT 1¿ OF THE CASING HAS NO MAGNETS IN IT. ABOUT .8¿ OF THE CASING IS MISSING. IF THERE ARE 21 MAGNETS IN EACH WIRE, 12 ARE NOT ACCOUNTED FOR. PATIENT INFO: COMORBIDITIES: TRIPLE A AORTIC MITRALTIS, AORTIC ROOT DISSECTION, ASTHMA, PNEUMONIA, GERD, MELANOMA, AORTIC ROOT AND HEMIAORTIC ARCH REPLACEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11670751 ↗MalfunctionMar 5, 2021Material Deformation; Material Deformation
IT WAS REPORTED "CK000409A, POWERPICC SOLO 4 FRENCH PROVENA DUAL, KINKING ISSUE. PLACE ON 2/1, PATIENT OBESE. THEY RETHREADED WITH A 5 FRENCH DUAL IN IR."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11421866 ↗MalfunctionFeb 25, 2021Material Twisted/Bent; No Flow
IT WAS REPORTED "PLACED ON 1/26 CK000409A 4 FRENCH DUAL LUMEN POWERPICC PROVENA SOLO, THE NEXT DAY NO BLOOD RETURN, XRAY TAKEN OF ARM AND KINK FOUND. "
Patient problems: Insufficient Information
MAUDE report 11379883 ↗MalfunctionFeb 19, 2021Material Deformation; Difficult to Flush; Material Deformation; Material Twisted/Bent
IT WAS REPORTED "PICC PLACED ON 12/20 EXTERNAL 2 CM. RIGHT BRACHIAL PLACEMENT. 12/27 ISSUES WITH BLOOD RETURN/FLUSHING BUT COULD BE FIXED POSITIONALLY. 12/28 PICC WAS KINKING FOUND WITH XRAY, WAS RETHREADED BY IR, THEY DID NOT RE-ACCESS THROUGH A NEW HOLE. 1/1 TPA USED AND PULLED BACK, NO BLOOD RETURN STILL. 1/7 IT WAS REMOVED BECAUSE THEY WERE STILL HAVING ISSUES."
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11348767 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page