PowerPICC Provena CK000988Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741216022), received between Oct 26, 2023 and Sep 5, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Adverse Event Without Identified Device or Use Problem1
- Device Markings/Labelling Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hemorrhage/Blood Loss/Bleeding1
- Skin Tears1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT PICC WAS LEAKING AND BLEEDING. INSERTED PICC IN PATIENT ON (B)(6) 2025. AFTER HAVING A PROCEDURE, PACU NOTIFIED THE MANAGER THAT THE PATIENT'S PICC WAS LEAKING AND BLEEDING. WAS ABLE TO SEE THAT THE PICC WAS FRACTURED 2CM PAST THE INSERTION SITE WHERE IT HAD KINKED. PATIENT WAS RECEIVING ANTIBIOTIC THERAPY. PATIENT HAD TO BE STUCK AGAIN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22979987 ↗MalfunctionFeb 19, 2025Fluid/Blood Leak; Device Markings/Labelling Problem; Fluid/Blood Leak; Device Markings/Labelling Problem
IT WAS REPORTED BY THE CUSTOMER, NURSE WAS CUTTING THE PICC TO INSERT WHEN SHE NOTICED A POTENTIAL CUT OR MARK ON THE PICC 3.5CM UP FROM THE CUT. SHE DID NOT INSERT THE PICC AND GOT A NEW ONE TO INSERT. ADDITIONAL INFORMATION RECEIVED 01/29/2025: IT WAS REPORTED BY THE CUSTOMER, "NOT SURE IF IT WAS A CUT OR A MARK BUT THE PICC IS IN ONE PIECE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21411485 ↗MalfunctionJan 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER, KINKING OF THE CATHETER NOTED ON CXR. WHEN THE CATHETER WAS REMOVED FROM THE PATIENT THERE WERE 2 FRACTURES IN THE PICC LINE WHERE THE CATHETER WAS KINKED. THIS WAS APPROXIMATELY MIDWAY OF THE LENGTH OF THE CUT CATHETER (35 CM). PATIENT HAD TO BE STUCK AGAIN. THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21118698 ↗MalfunctionApr 18, 2024Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT TRANSITION FROM THE PEEL AWAY SHEATH WILL ROLL UP ON TISSUE. IT SPLITS IT AND HAVE TO GET A NEW ONE. IT IS SOMETHING WITH THE TRANSITION THAT CATCHES. HAVE TO GO BACK IN AND KNICK A SECOND TIME TO TRY TO ADVANCE THE INTRODUCER. THERE WAS NO ADVERSE OUTCOME OR SERIOUS INJURY TO THE PATIENT OR HEALTHCARE PROVIDER. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19135065 ↗InjuryOct 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY CUSTOMER THAT REMOVING THE STATLOCK DURING A DRESSING CHANGE, THE PATIENT'S SKIN CAME OFF IN A PERFECT OUTLINE OF WHERE THE PICC STAT-LOCK WAS LOCATED. NURSE DID NOT RE-APPLY A STATLOCK WHEN THEY SECURED THE PICC AGAIN. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Skin Tears
MAUDE report 18014072 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page