Skip to content

ARROW IPN035912Recalls & reported problems

Teleflex Incorporated · Radial Artery Catheterization Set · Class 2

Quick summary

This Teleflex Arrow radial artery catheterization set is a kit used to place a catheter in the radial artery. Its 92 reports, with 14 injuries and no deaths, often describe guidewires that kink, bend, or break, and catheters that split or get stuck. In a few cases, pieces broke off inside the artery. No recalls name this device. Reports for the last 12 months (7) are fewer than in the previous 12 months (19).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
92
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 92 adverse event reports that list this exact device (DI 00801902002730), received between May 25, 2016 and May 5, 2026 — classified as 14 injuries, 78 malfunctions.

The most frequently reported specific device problem is “Material Separation” (19 reports).

32 other recalls cover devices cleared under the same FDA submission (K810675); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 5 reportsMar 2024: 13 reportsApr 2024: 0 reportsMay 2024: 7 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 3 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation19
  • Material Twisted/Bent14
  • Obstruction of Flow12
  • Material Split, Cut or Torn11
  • Physical Resistance/Sticking8
  • Deformation Due to Compressive Stress7
  • Difficult to Remove6
  • Failure to Advance6

Patient problems

  • No Clinical Signs, Symptoms or Conditions49
  • Insufficient Information22
  • Foreign Body In Patient8
  • Needle Stick/Puncture3
  • Device Embedded In Tissue or Plaque2
  • Hematoma2
  • No Known Impact Or Consequence To Patient2
  • Hemodynamic instability1

Recent adverse event reports

The 25 most recent of 92 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Material Twisted/Bent

    IT WAS REPORTED THAT, "DURING PREPARATION PRIOR TO THE ARTERIAL LINE PROCEDURE, THE GUIDEWIRE KINKED AND COULD NOT BE ADVANCED." THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25081827 ↗
  • MalfunctionApr 17, 2026Break

    IT WAS REPORTED THAT "WE HAD A RADIAL ARTERY CATHETERIZATION SET RA-04020 WHERE THE WIRE BROKE IN HALF WHEN INSERTING THE CATHETER INTO THE RADIAL ARTERY". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24910612 ↗
  • InjuryMar 23, 2026Break

    IT WAS REPORTED "DURING THE PROCEDURE, A PORTION OF THE CATHETER BROKE OFF WITHIN THE PATIENT'S RADIAL ARTERY. THE OPERATING SURGEON WAS IMMEDIATELY NOTIFIED, AND A VASCULAR SURGERY CONSULT WAS OBTAINED. ULTRASOUND IMAGING WAS PERFORMED, AND THE RETAINED CATHETER FRAGMENT WAS VISUALIZED WITHIN THE PATIENT'S RADIAL ARTERY". THERE WAS NO PATIENT HARM OR INJURY. ADDITIONAL INFORMATION RECEIVED STATES "THE CATHETER FRAGMENT REMAINS IN SITU, AND THE VASCULAR SURGERY TEAM IS ACTIVELY FOLLOWING AND MANAGING THE PATIENT".

    Patient problems: Foreign Body In Patient

    MAUDE report 24662726 ↗
  • InjuryFeb 4, 2026Material Separation

    IT WAS REPORTED THAT " WHILE PLACING THE A-LINE A PIECE OF THE WIRE FROM THE RADIAL ARTERY CATHETERIZATION SET SEVERED OFF AND WAS RETAINED IN THE PATIENT. PIECE OF WIRE THAT WAS RETAINED IN THE PATIENT'S RIGHT WRIST WAS CONFIRMED VIA FLUORO AND ULTRASOUND. SURGICAL CONSULT IS BEING PLACED TO REMOVE THE FOREIGN BODY". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Foreign Body In Patient

    MAUDE report 24257538 ↗
  • MalfunctionDec 16, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT "NOT GETTING BLOOD FLASHBACK IN CHAMBER". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23817203 ↗
  • MalfunctionDec 3, 2025Material Puncture/Hole; Material Twisted/Bent; Material Puncture/Hole; Material Twisted/Bent

    IT WAS REPORTED "X2 ARTERIAL CATHETERIZATION SETS WERE NOTED TO BE FAULTY. IT SEEMS THE CONNECTION POINT BETWEEN THE CANNULA AND THE NEEDLE DISLODGED INAPPROPRIATELY CAUSING THE NEEDLE TO BEND AND PUT A HOLE THROUGH THE CANNULA." A THIRD CANNULA WAS INSERTED AND USED SUCCESSFULLY. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE:9680794-2025-01045 AND 9680794-2025-01033.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705601 ↗
  • MalfunctionDec 3, 2025Material Puncture/Hole; Material Twisted/Bent; Material Puncture/Hole; Material Twisted/Bent

    IT WAS REPORTED "X2 ARTERIAL CATHETERIZATION SETS WERE NOTED TO BE FAULTY. IT SEEMS THE CONNECTION POINT BETWEEN THE CANNULA AND THE NEEDLE DISLODGED INAPPROPRIATELY CAUSING THE NEEDLE TO BEND AND PUT A HOLE THROUGH THE CANNULA." A THIRD CANNULA WAS INSERTED AND USED SUCCESSFULLY. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE:9680794-2025-01045 AND 9680794-2025-01033.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705600 ↗
  • MalfunctionOct 21, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT "DURING ARTERIAL PRESSURE MONITORING IN THE OPERATING ROOM USING AN RA CATHETER, THE NEEDLE TIP WAS NOT VISIBLE." A NEW KIT WAS USED. THERE WAS NO PATIENT HARM OR INJURY. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS UNKNOWN AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23347381 ↗
  • MalfunctionSep 10, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED "THE SET BROKE DOWN WHEN ATTEMPTING TO ADVANCE THE ACTUAL CATHETER OVER THE INCLUDED WIRE. THE WHOLE SET LOCKED UP AND HAD TO BE TAKEN OUT EN BLOC. AFTER INSPECTION, IT APPEARS AS IF THE CATHETER ITSELF WAS SPLIT FROM THE NORMAL DISTAL HOLE AND UP THE SIDE OF THE CATHETER FOR ABOUT 5-6 MM." ANOTHER CATHETER WAS USED. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: Insufficient Information

    MAUDE report 23016672 ↗
  • MalfunctionAug 18, 2025Material Separation; Material Separation

    IT WAS REPORTED " 07/17: FELLOW ATTEMPTED TO PLACE RIGHT RADIAL A-LINE. THE FIRST SET THAT HE ATTEMPTED WITH WAS THE DEFECTIVE SET. THE NEEDLE SLID OUT AFTER FELLOW HAD PUNCTURED PATIENT'S SKIN. HE WAS ABLE TO REMOVE IT AND THE NEEDLE SLID COMPLETELY OUT OF THE PLASTIC SHEATH. THE RESIDENT, WENT TO RETRIEVE A SECOND INSERTION SET." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE AND UNCHANGED".

    Patient problems: Needle Stick/Puncture

    MAUDE report 22828003 ↗
  • MalfunctionAug 18, 2025Break; Difficult to Remove; Failure to Advance; Break; Difficult to Remove; Failure to Advance

    IT WAS REPORTED "THE LEFT RADIAL ARTERY WAS IDENTIFIED WITH PALPATION AND ULTRASOUND AND A-LINE NEEDLE WAS ADVANCED UNTIL FLASH OF PULSATILE BRIGHT RED BLOOD WAS OBSERVED UP TO MARKER LINE. GUIDEWIRE WAS SMOOTHLY ADVANCED OVER THE CATHETER BUT THE CATHETER WOULD NOT ADVANCE OVER THE WIRE. THE WIRE WOULD NOT REMOVE AND IT WAS FOUND TO BE BROKEN. THE APPARATUS WAS REMOVED AND PRESSURE HELD. ON SCANNING A SMALL HEMATOMA WAS VISIBLE BUT NOT RETAINED WIRE." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Hematoma

    MAUDE report 22826680 ↗
  • MalfunctionAug 11, 2025Material Split, Cut or Torn; Physical Resistance/Sticking; Material Split, Cut or Torn; Physical Resistance/Sticking

    IT WAS REPORTED "RADIAL ARTERIAL ART LINE INSERTION. CATHETER IN ARTERY AS PER ULTRASOUND GUIDE. GUIDE WIRE THREADED THROUGH CATHETER EASILY. UPON CATHETER INSERTION OVER GUIDE WIRE, RESISTANCE FELT. CATHETER PULLED BACK AND GUIDE WIRE PULLED BACK. GUIDE WIRE FLOATED INTO VESSEL ON ITS OWN. AT THIS POINT, BOTH CATHETER AND GUIDE WIRE REMOVED FROM PATIENT. GUIDE WIRE "COATING" WAS SEVERED. X RAY DONE AND VASCULAR SERVICE CONTACTED FOR FURTHER MANAGEMENT. XRAY SHOWED NO GUIDE WIRE IN PATIENT". SURGERY WAS DELAYED FOR 45 MINUTES. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: Insufficient Information

    MAUDE report 22771449 ↗
  • MalfunctionAug 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED "RA INSERTED AND THEN REMOVED AS THE STEEL CANNULA PART OF THE ART LINE HAS DETACHED. IT HAD FALLEN OFF WHILE STILL PARTIALLY WITHIN THE PLASTIC INTRA- ARTERIAL CANNULA. LUCKILY ENOUGH WAS LEFT THAT THEY WERE ABLE TO GRAB AND REMOVE." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22727612 ↗
  • MalfunctionJul 8, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT "DURING A PROCEDURE USING THE RA-04020 CATHETER IN THE OPERATING ROOM, THERE WAS NO BLOOD RETURN AFTER PUNCTURE. REPLACED WITH THE NEW PRODUCT." A NEW KIT WAS OPENED. THERE WAS NO MEDICAL INTERVENTION REQUIRED. NO ADVERSE EVENT REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22438838 ↗
  • MalfunctionJun 20, 2025Material Separation; Material Separation

    IT WAS REPORTED "AFTER SUCCESSFUL CANNULATION OF RADIAL ARTERY CATHETER WAS ADVANCED. BACK PORTION OF NEEDLE WAS REMOVED BUT THIS ONLY REMOVED THE WIRE LEAVING THE NEEDLE IN PATIENT. THIS WAS REMOVED AND CATHETER HAD TO BE REPLACED." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22300607 ↗
  • MalfunctionJun 13, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED "DURING A PROCEDURE USING THE RA-04020 CATHETER IN THE OPERATING ROOM, THERE WAS NO BLOOD RETURN AFTER PUNCTURE." A NEW KIT WAS USED TO RESOLVE THE ISSUE. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: Insufficient Information

    MAUDE report 22201884 ↗
  • InjuryFeb 21, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT THE INSERTION WAS SUCCESSFUL WITH FLASHBACK. THE GUIDEWIRE ADVANCED WITHOUT ISSUE. HOWEVER, WHEN ATTEMPTING TO REMOVE THE DEVICE, SIGNIFICANT RESISTANCE WAS ENCOUNTERED. THE GUIDEWIRE COULD NOT BE RETRACTED, AND THE CATHETER NEEDLE COULD NOT BE REMOVED. THE MDA CAME TO REMOVE THE ENTIRE DEVICE AND DISCOVERED THAT THE CATHETER TIP HAD BEEN SHEARED OFF. A FOREIGN BODY WAS VISIBLE ON ULTRASOUND. THE CATHETER WAS REMOVED VIA SURGICAL CUTDOWN BY VASCULAR SURGERY. THE PATIENT IS FINE AND HAD NO HARM OR INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 21432641 ↗
  • MalfunctionFeb 10, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED "ANAETHETIST INSERTING ARTERIAL CATHETER HAD 2 FAILED ATTEMPTS WITH NO EVIDENCE OF FLASH BACK SEEN. ON REMOVAL STRONG BRIGHT BLOOD FLOW INDICATING WAS IN THE ARTERY ON BOTH OCCASIONS". A NEW KIT WAS USED TO RESOLVE THE ISSUE. THE PATIENT IS FINE AND HAD NO HARM OR INJURY. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00143 AND 9680794-2025-00142.

    Patient problems: Needle Stick/Puncture; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344543 ↗
  • MalfunctionFeb 10, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED, "ANAETHETIST INSERTING ARTERIAL CATHETER HAD 2 FAILED ATTEMPTS WITH NO EVIDENCE OF FLASH BACK SEEN. ON REMOVAL STRONG BRIGHT BLOOD FLOW INDICATING WAS IN THE ARTERY ON BOTH OCCASIONS". A NEW KIT WAS USED TO RESOLVE THE ISSUE. THE PATIENT IS FINE AND HAD NO HARM OR INJURY. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00143 AND 9680794-2025-00142.

    Patient problems: Needle Stick/Puncture; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344523 ↗
  • MalfunctionJan 9, 2025Difficult to Remove; Failure to Advance; No Flow; Difficult to Remove; Failure to Advance; No Flow

    IT WAS REPORTED "BLOOD WAS NOT ABLE TO FLASH BACK. ALSO, WOULD NOT ALLOW THE WIRE TO ADVANCE. AND THEN HAD TROUBLE GETTING THE NEEDLE AND WIRE TO COME OUT OF THE PATIENT'S ARM." THE CUSTOMER REPORTED THIS EVENT HAS OCCURED SEVERAL TIMES. ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER; HOWEVER, FURTHER DETAILS HAVE NOT BEEN PROVIDED AT THIS TIME.THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2024-01234, 9680794-2024-01254 AND 9680794-2025-00043.

    Patient problems: Insufficient Information

    MAUDE report 21112717 ↗
  • MalfunctionJan 7, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT TWO A-LINE ARROWS BENT UPON INSERTION INTO THE PATIENT DURING THE PROCEDURE. THE PATIENT CONDITION WAS REPORTED AS "UNKNOWN". ASSOCIATED MDR#: 9680794-2025-00023.

    Patient problems: Insufficient Information

    MAUDE report 21090849 ↗
  • MalfunctionJan 7, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT TWO A-LINE ARROWS BENT UPON INSERTION INTO THE PATIENT DURING THE PROCEDURE. THE PATIENT CONDITION WAS REPORTED AS "UNKNOWN". ASSOCIATED MDR#: 9680794-2025-00024.

    Patient problems: Insufficient Information

    MAUDE report 21090731 ↗
  • MalfunctionDec 26, 2024Difficult to Remove; Obstruction of Flow; Failure to Advance; Material Twisted/Bent; Difficult to Remove; Obstruction of Flow; Failure to Advance; Material Twisted/Bent

    IT WAS REPORTED THAT 'THE DEVICE WAS NOT ALLOWING BLOOD TO FLASH BACK. ALSO, THE WIRE COULD NOT BE ADVANCED, AND THEN HAD TROUBLE GETTING THE NEEDLE AND WIRE TO COME OUT OF THE PATIENT'S ARM'. THE ISSUE HAD OCCURRED SEVERAL TIMES AND NO PATIENT INJURIES WERE NOTED. THE CATHETERS WERE REPLACED TO RESOLVE THE ISSUE. NO PATIENT INJURY OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21019626 ↗
  • MalfunctionDec 17, 2024Difficult to Remove; Obstruction of Flow; Failure to Advance; Difficult to Remove; Obstruction of Flow; Failure to Advance

    IT WAS REPORTED, "REFERENCE NUMBER (B)(4) IS NOT ALLOWING A BLOOD TO FLASH BACK. ALSO, WOULD NOT ALLOW THE WIRE TO ADVANCE. AND THEN HAD TROUBLE GETTING THE NEEDLE AND WIRE TO COME OUT OF THE PATIENT'S ARM." THE CUSTOMER REPORTED THIS EVENT HAS OCCURED SEVERAL TIMES. ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER; HOWEVER, FURTHER DETAILS HAVE NOT BEEN PROVIDED AT THIS TIME. ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2024-01234 AND 9680794-2024-01254.

    Patient problems: Insufficient Information

    MAUDE report 20951584 ↗
  • MalfunctionDec 12, 2024Difficult to Remove; Obstruction of Flow; Failure to Advance; Material Twisted/Bent; Difficult to Remove; Obstruction of Flow; Failure to Advance; Material Twisted/Bent

    IT WAS REPORTED "IS NOT ALLOWING A BLOOD TO FLASH BACK. ALSO, WOULD NOT ALLOW THE WIRE TO ADVANCE. AND THEN HAD TROUBLE GETTING THE NEEDLE AND WIRE TO COME OUT OF THE PATIENT'S ARM." ADDITIONAL INFORMATION RECEIVED REPORTS "THE WIRE WHEN PULLED OUT HAD A Z IN IT ONE TIME". THE ARTERIAL CATHETER WAS REMOVED AND ANOTHER LINE WAS PLACED. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20912823 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K810675. One submission often covers many models, so check the product and lot details in each record.

  • Z-2397-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18
  • Z-3299-2024 · Jul 26, 2024 · Open, Classified
    Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray
  • Z-0844-2020 · Dec 12, 2019 · Open, Classified
    Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04
  • Z-2019-2018 · Apr 11, 2018 · Terminated
    Arterial Catheterization Kit; Product Code:NA-04550-1A
  • Z-1968-2018 · Apr 11, 2018 · Terminated
    Radial Arterial Access Kit; Product Code: ASK-04001-BID2
  • Z-1970-2018 · Apr 11, 2018 · Terminated
    Radial Artery Catheterization Kit; Product Codes: ASK-04020-AH, ASK-04020-MIHS, ASK-04120-HF1
  • Z-1972-2018 · Apr 11, 2018 · Terminated
    Radial Artery Catheterization Kit with Sharps Safety Features; Product Codes: ASK-04020-RIHM
  • Z-1969-2018 · Apr 11, 2018 · Terminated
    Arterial Line Kit with Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-04018-OH1
  • Z-1971-2018 · Apr 11, 2018 · Terminated
    Arterial Catheterization Kit; Product Codes: ASK-04020-DH
  • Z-1330-2018 · Dec 22, 2017 · Terminated
    (1) Arterial Catheterization Kit, Part Number (PN): ASK-04018-CC; (2) Arterial Line Kit with Sharps Safety Features, PN: ASK-04510-HUM; (3) Arterial Line Kit with Sharps Safety Fea
  • Z-1286-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04220
  • Z-1281-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04018
  • Z-1282-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04020
  • Z-1284-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04120
  • Z-1283-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04020-C

…and 17 more.

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.