ARROW IPN046311 (REF JC-05400-B)Recalls & reported problems
Teleflex Incorporated · Epidural Catheterization Set · Class 2
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801902029676), received between Mar 26, 2022 and Jan 26, 2023 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
4 other recalls cover devices cleared under the same FDA submission (K103658); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak3
- Obstruction of Flow2
- Difficult to Advance1
- Material Separation1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information2
- Device Embedded In Tissue or Plaque1
- Spinal Column Injury1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2023Material Twisted/Bent; Material Twisted/Bent
REPORTED EVENT: THE CATHETER KINKED INSIDE THE SNAPLOCK WHICH CAUSED AN OCCLUSION; TREATMENT FLOW STOPPED. THE SNAPLOCK AND THE CATHETER WERE REPLACED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16246255 ↗MalfunctionJan 24, 2023Fluid/Blood Leak; Fluid/Blood Leak
REPORTED EVENT: INCIDENT OCCURED ON (B)(6) 2022. THE PLUNGER OF THE LOR SYRINGE LEAKED.
Patient problems: Insufficient Information
MAUDE report 16233870 ↗MalfunctionJan 10, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT DURING THE USE OF AN OBSTETRICAL EPIDURAL IT WAS OBSERVED THAT THE CATHETER WAS NOT PERFORATED. SO THE INJECTION WAS IMPOSSIBLE. IT WAS NECESSARY TO USE ANOTHER DEVICE AND THUS THE PATIENT HAD 2 CATHETERS INSERTED IN A ROW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16131655 ↗MalfunctionDec 12, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT ON 30 OCTOBER 2022 IN THE MATERNITY DEPARTMENT, THE CATHETER HAD A HOLE ABOUT 5 CM FROM THE END ON THE PUNCTURE POINT SIDE AND LEAKED. THIS ISSUE WAS OBSERVED AFTER THE INSERTION OF THE DEVICE WHEN CHECKING THAT IT WAS WORKING PROPERLY. THE DEVICE WAS REMOVED. INCIDENT WITH NO CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15959778 ↗MalfunctionNov 14, 2022Fluid/Blood Leak; Fluid/Blood Leak
REPORTED ISSUE: INCIDENT OCCURRED ON 16 OCTOBER 2022; THE PATIENT IS FEMALE 25-YEAR-OLD IN THE MATERNITY DEPARTMENT. THE EPIDURAL CATHETER WAS INSERTED AND DURING THE INFUSION OF THE TEST DOSE, WE OBSERVED A LEAK COMING FROM A HOLE IN THE BODY OF THE CATHETER. THE HOLE IS LOCATED 20-30 CM FROM THE END OF THE TUBING; THE CATHETER WAS REMOVED AND REPLACED. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15789250 ↗MalfunctionJun 27, 2022Material Separation; Material Separation
DURING THE REMOVAL OF THE EPIDURAL CATHETER 3 HOURS BEFORE GIVING BIRTH, THE CATHETER SEPARATED DURING THE REMOVAL. IT LOOKS LIKE 7 CM REMAINED INSIDE THE FEMALE PATIENT. THERE WAS NO CONSEQUENCE FOR THE PATIENT. A SCAN HAS BEEN REQUESTED TO CHECK THE POSITION OF THE PART INSIDE THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions
MAUDE report 14829626 ↗MalfunctionJun 6, 2022Difficult to Advance; Difficult to Advance
SEVERAL KITS FACED AN ISSUE. ANESTHETISTS HAVE GREAT DIFFICULTY ADVANCING THE CATHETER ONCE THE NEEDLE IS IN PLACE. SEVERAL CATHETERS COULD NOT BE INSERTED AND OTHERS NEEDED ADDITIONAL FORCE DURING INSERTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14606494 ↗MalfunctionMay 20, 2022Obstruction of Flow; Obstruction of Flow
INCIDENT HAPPENED IN THE SURGERY. IT WAS IMPOSSIBLE TO INJECT INTO THE CATHETER (BLOCKAGE). WE HAD TO REMOVE THE CATHETER AND INSERT A NEW DEVICE IN THE FEMALE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14455237 ↗InjuryMar 26, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
INVOLVED A FEMALE PATIENT WHO WAS IN MATERNITY ROOM. THE LOR SYRINGE DOES NOT IDENTIFY THE LOSS OF RESISTANCE DURING THE PROCEDURE OF INSERTION OF THE EPIDURAL CATHETER. WITH PHYSIOLOGICAL SALINE, THE SYRINGE STIFFENED AND RESISTANCE OCCURED LEADING TO A DURAL BREACH. ADDITIONAL INFORMATION: PATIENT WENT HOME. A BLOOD PATCH WAS APPLIED AND ANALGESICS PARACETAMOL PLUS IBUPROFENE. THE DEVICE IS NOT AVAILABLE FOR INVESTIGATION.
Patient problems: Spinal Column Injury
MAUDE report 13915357 ↗
FDA clearance and classification
Cleared through 510(k) K103658 on May 16, 2012 (decision: Substantially Equivalent). Submitted by Teleflex, Inc. as “Flextip Plus Closed Tip Epidural Catherter”. The FDA review took 518 days. FDA record ↗
The FDA regulates product code CAZ (anesthesia conduction kit) under 21 CFR 868.5140 as a Class II device. 21 CFR 868.5140 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K103658. One submission often covers many models, so check the product and lot details in each record.
- Z-2364-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;
- Z-1179-2020 · Jan 15, 2020 · TerminatedFlexTip Plus Epidural Catheterization Kit, Product Code ASK-05400-CA1
- Z-1962-2018 · Apr 11, 2018 · TerminatedSnapLock" Catheter/Syringe Adapter; Product Code: AS-05500
- Z-2462-2015 · Jul 30, 2015 · TerminatedPICC Kits with Peel-Away Sheath Component; catheter, intravascular, therapeutic, ultrasound Product Usage: catheter, intravascular, therapeutic, ultrasound
Common questions about the ARROW IPN046311
Has the ARROW IPN046311 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801902029676), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ARROW IPN046311?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Mar 26, 2022 and Jan 26, 2023. In those reports, the most common specific device problems are “Fluid/Blood Leak” (3), “Obstruction of Flow” (2) and “Difficult to Advance” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (1) and “Spinal Column Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN046311?
None of the reports are classified as deaths, 1 as an injury and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN046311 FDA approved?
It was cleared through 510(k) K103658 on May 16, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the ARROW IPN046311, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN046311). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Anesthesia conduction kit devices
- BD Luer-Lok, BD Glaspak 405671Becton, Dickinson And Company · Anesthesia conduction kit1 recall68 reports
- ARROW IPN922678Teleflex Incorporated · Anesthesia conduction kit2 recalls0 reports
- BD Luer-Lok, BD Glaspak 405672Becton, Dickinson And Company · Anesthesia conduction kit1 recall26 reports
- BD Luer-Lok, BD Glaspak 405673Becton, Dickinson And Company · Anesthesia conduction kit1 recall10 reports
- BD Luer-Lok, BD Glaspak 405637Becton, Dickinson And Company · Anesthesia conduction kit1 recall7 reports
- BD Luer-Lok, BD Glaspak 405707Becton, Dickinson And Company · Anesthesia conduction kit1 recall4 reports
- BD Luer-Lok, BD Glaspak 405609Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
- BD Luer-Lok, BD Glaspak 405735Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.