ARROW IPN032961Recalls & reported problems
Teleflex Incorporated · Arterial Catheterization Kit · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801902067050), received between Feb 15, 2023 and Feb 15, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
32 other recalls cover devices cleared under the same FDA submission (K810675); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- Loss of consciousness1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 15, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
CUSTOMER REPORTED PATIENT LOST CONSCIOUSNESS DUE TO "A CLOT IN THE CATHETER OR AIR". UNKNOWN IF THE ISSUE WAS WITH THE CATHETER OR EDWARDS TUBING. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Loss of consciousness
MAUDE report 16382103 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K810675. One submission often covers many models, so check the product and lot details in each record.
- Z-2397-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18
- Z-3299-2024 · Jul 26, 2024 · Open, ClassifiedArrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray
- Z-0844-2020 · Dec 12, 2019 · Open, ClassifiedArrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04
- Z-1969-2018 · Apr 11, 2018 · TerminatedArterial Line Kit with Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-04018-OH1
- Z-1970-2018 · Apr 11, 2018 · TerminatedRadial Artery Catheterization Kit; Product Codes: ASK-04020-AH, ASK-04020-MIHS, ASK-04120-HF1
- Z-1968-2018 · Apr 11, 2018 · TerminatedRadial Arterial Access Kit; Product Code: ASK-04001-BID2
- Z-1971-2018 · Apr 11, 2018 · TerminatedArterial Catheterization Kit; Product Codes: ASK-04020-DH
- Z-2019-2018 · Apr 11, 2018 · TerminatedArterial Catheterization Kit; Product Code:NA-04550-1A
- Z-1972-2018 · Apr 11, 2018 · TerminatedRadial Artery Catheterization Kit with Sharps Safety Features; Product Codes: ASK-04020-RIHM
- Z-1330-2018 · Dec 22, 2017 · Terminated(1) Arterial Catheterization Kit, Part Number (PN): ASK-04018-CC; (2) Arterial Line Kit with Sharps Safety Features, PN: ASK-04510-HUM; (3) Arterial Line Kit with Sharps Safety Fea
- Z-1281-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04018
- Z-1282-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04020
- Z-1286-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04220
- Z-1284-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04120
- Z-1283-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04020-C
…and 17 more.
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page