Skip to content

ARROW IPN033472Recalls & reported problems

Teleflex Incorporated · ARROW(R) PICC Set · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801902095657), received between Aug 30, 2024 and Jul 31, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Separation” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K080604); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation4
  • Material Split, Cut or Torn2
  • Crack1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2025Material Separation; Material Separation

    IT WAS REPORTED "PINK LUMEN OF PICC LINE SNAPED FROM PATIENT FORCE. THE PINK LUMEN HAD NO CLAP WHEN IT SNAPPED, SHE TOOK ONE OF THE CLAMPS FROM ANOTHER LUMEN TO CLAMP IT, END OF LUMEN ALSO COVERED WITH TEGADERM. 7F EDUCATOR DISCUSSED WITH VA NURSE PRACTITIONER KATIE, REQUESTED PICC TO BE REMOVED. PICC LINE IS REMOVED UNDER ASEPTIC TECHNIQUE AND PROTOCOL, PATIENT VITAL SIGNS ARE BETWEEN THE FLAGS, NO SIGNS OF INFECTION AT THE INSERTION SITE." A PIV WAS INSERTED TO CONTINUE THE TREATMENTS. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22675579 ↗
  • MalfunctionJan 17, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE PICC PEEL-AWAY SHEATH RIPPED DURING INSERTION OF DILATOR. THE DEVICE WAS REMOVED IN ITS ENTIRETY AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21179360 ↗
  • MalfunctionOct 16, 2024Material Separation; Material Split, Cut or Torn; Material Separation; Material Split, Cut or Torn

    IT WAS REPORTED "THE INTRODUCER SHEATH IS MEANT TO SPLIT CLEANLY AS PART OF THE REMOVAL PROCESS AFTER THE LINE IS INSERTED. INSTEAD THE SHEATH TORE PARTIALLY AND THEN ONE OF THE HANDLES SNAPPED OFF." THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN". ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER, HOWEVER FURTHER DETAILS HAVE NOT BEEN PROVIDED AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 20460370 ↗
  • MalfunctionSep 5, 2024Crack; Material Rupture; Crack; Material Rupture

    IT WAS REPORTED THAT: "EXTRAVASATION NOTED TO PT R) ARM WITH ++ SWELLING TO ARM AND R) HAND. TPN FLUID NOTED UNDER DRESSING. ALL INFUSIONS (HEPARIN + TPN) STOPPED AND PICC REMOVED. ON REMOVAL OF PICC LINE, PICC NOTED TO HAVE SMALL HAIRLINE FRACTURE DISTALLY ON PICC. VISIBLE SPLIT TO TUBING JUST PROXIMAL TO SKIN LEVEL. TPN AND HEPARIN INFUSIONS WERE STOPPED UNTIL NEW PICC LINE INSERTED. THE DEVICE WAS REMOVED IN ITS ENTIRETY AND PICC WAS REPLACED." THE RETURNED DEVICE PROVED TO BE RUPTURED. THERAPY WAS DELAYED BUT THE PATIENT WAS REPORTED AS FINE POST THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20143737 ↗
  • MalfunctionAug 30, 2024Material Separation; Material Separation

    IT WAS REPORTED: "PICC INSERTED THROUGH SHEATH AND WHEN THEY CRACKED PEEL AWAY SHEATH TO PULL IT APART, THE WHITE PLASTIC PART AT THE END BROKE OFF. SHEATH WAS ABLE TO BE REMOVED AND PICC REMAINED INSITU. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20105258 ↗
  • MalfunctionAug 30, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: "PICC INSERTED THROUGH SHEATH AND WHEN THEY CRACKED PEEL AWAY SHEATH TO PULL IT APART, THE WHITE PLASTIC PART AT THE END BROKE OFF. SHEATH WAS ABLE TO BE REMOVED AND PICC REMAINED INSITU. THERE WAS NO PATIENT HARM OR CONSEQUENCE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20105225 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080604. One submission often covers many models, so check the product and lot details in each record.

  • Z-2024-2018 · Apr 11, 2018 · Terminated
    ARROW PICC Set; Product codes: PR-35563-HPHNL and PR-35563-HPHNM

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates