Skip to content

ARROW IPN032602Recalls & reported problems

Teleflex Incorporated · Percutaneous Cavity Drainage Catheterization Set · Class 1

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801902097903), received between Mar 4, 2024 and Dec 26, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent4
  • Crack1
  • Device Damaged Prior to Use1
  • Device Dislodged or Dislocated1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "DOCTOR ATTEMPTED TO INSERT THE CAVITY DRAIN, EVERY TIME HE TRIED TO THREAD THE GUIDE WIRE IT KINKED. HE THEN OPENED A SECOND SET AND THERE WAS THE SAME OUTCOME. HE EVENTUALLY OPENED THE CVC SET JUST TO TAKE OUT THE GUIDE WIRE AND HE USED THAT AND IT WAS SUCCESSFUL". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23896308 ↗
  • MalfunctionDec 26, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "DOCTOR ATTEMPTED TO INSERT THE CAVITY DRAIN, EVERY TIME HE TRIED TO THREAD THE GUIDE WIRE IT KINKED. HE THEN OPENED A SECOND SET AND THERE WAS THE SAME OUTCOME. HE EVENTUALLY OPENED THE CVC SET JUST TO TAKE OUT THE GUIDE WIRE AND HE USED THAT AND IT WAS SUCCESSFUL". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23896302 ↗
  • MalfunctionSep 5, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED "THE DOCTOR INSERTED NEEDLE AND SYRINGE AS SOON AS HE STARTED THREADING THE GUIDEWIRE THE GUIDEWIRE KINKED. THE DEVICE REMOVED ENTIRELY, MADE TWO ATTEMPTS, THIRD ATTEMPT WAS SUCCESSFUL BUT THERAPY GOT DELAYED." IT WAS REPORTED THERE WAS NO HARM TO THE PATIENT DUE TO THE DEVICE. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2024-00857.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20146054 ↗
  • MalfunctionSep 3, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED "THEY OPENED THE ITEM, BUT THE INTRODUCER GUIDEWIRE WAS FAULTY, AND THEY COULD NOT USE THE SPECIFIC WIRE , BECAUSE THEY COULDN'T GAIN ACCESS INTO CAVITY." ADDITIONAL INFORMATION RECEIVED REPORTS THE GUIDEWIRE WAS KINKED. THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20127444 ↗
  • MalfunctionSep 2, 2024Crack; Device Damaged Prior to Use; Crack; Device Damaged Prior to Use

    IT WAS REPORTED THAT: "ON 21/5/2024 DR (PULMONOGIST) OPENED PACK TO USE ON PATIENT AND NOTED NEEDLE BROKEN IN SET. PACKAGING WAS NOT DAMAGED." THE RETURNED DEVICE SHOWED A CRACK IN THE NEEDLE HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20117116 ↗
  • MalfunctionMar 4, 2024Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration

    IT WAS REPORTED THAT: PATIENT HAD DRAIN FOR PLEURAL EFFUSION AND THE CATHETER SLIPPED OUT THROUGH THE TIED DOWN WINGS. CATHETER WAS FOUND IN BED AND WINGS STILL SUTURED TO SKIN. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. THE PATIENT WAS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18824752 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page