ARROW IPN033168Recalls & reported problems
Teleflex Incorporated · Arterial Catheterization Set · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801902100597), received between Aug 11, 2023 and Jul 7, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
32 other recalls cover devices cleared under the same FDA submission (K810675); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Deformation Due to Compressive Stress1
- Leak/Splash1
- Material Separation1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED "COPIOUS OOZING NOTED FROM A-LINE SITE DURING DRESSING CHANGE. ASSESSMENT REVEALED THE A-LINE CATHETER WAS DAMAGED AND NEEDED TO BE REPLACED. A SMALL PIECE OF THE CATHETER WAS BROKEN." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22413931 ↗InjuryMar 4, 2024Material Separation; Physical Resistance/Sticking; Material Separation; Physical Resistance/Sticking
IT WAS REPORTED THAT: CLINICIAN WAS ATTEMPTING TO INSERT ARTERIAL CATHETER HOWEVER ENCOUNTERED TROUBLE WITH CHALLENGING ACCESS. WHEN DEVICE WAS INSPECTED, IT WAS OBSERVED THAT A FRAGMENT OF THE CATHETER REMAINED IN THE PATIENT. ADDITIONAL INFORMATION: RESISTANCE WAS MET WHEN ATTEMPTING TO ADVANCE THE CATHETER. THE FRAGMENT OF THE SWG WAS FULLY REMOVED WITH SURGICAL INTERVENTION. THE PATIENT WAS CODING AT THE TIME OF THE INSERTION ATTEMPT. THEY ARE IN ICU AT PRESENT BECAUSE OF THEIR UNDERLYING CONDITION. THERE WAS NO OTHER REPORTED PATIENT HARM OR CONSEQUENCE FROM THE SWG SEPARATION.
Patient problems: Foreign Body In Patient
MAUDE report 18824692 ↗MalfunctionFeb 2, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT "SWG KINKING." THE ISSUE WAS IDENTIFIED PRIOR TO USE ON A PATIENT DURING INSPECTION/FUNCTIONAL TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18635022 ↗MalfunctionAug 11, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED "UNPACK THE PACKAGE, THE GUIDE WIRE WAS KINKING." NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17511101 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K810675. One submission often covers many models, so check the product and lot details in each record.
- Z-2397-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18
- Z-3299-2024 · Jul 26, 2024 · Open, ClassifiedArrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray
- Z-0844-2020 · Dec 12, 2019 · Open, ClassifiedArrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04
- Z-2019-2018 · Apr 11, 2018 · TerminatedArterial Catheterization Kit; Product Code:NA-04550-1A
- Z-1970-2018 · Apr 11, 2018 · TerminatedRadial Artery Catheterization Kit; Product Codes: ASK-04020-AH, ASK-04020-MIHS, ASK-04120-HF1
- Z-1971-2018 · Apr 11, 2018 · TerminatedArterial Catheterization Kit; Product Codes: ASK-04020-DH
- Z-1972-2018 · Apr 11, 2018 · TerminatedRadial Artery Catheterization Kit with Sharps Safety Features; Product Codes: ASK-04020-RIHM
- Z-1969-2018 · Apr 11, 2018 · TerminatedArterial Line Kit with Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-04018-OH1
- Z-1968-2018 · Apr 11, 2018 · TerminatedRadial Arterial Access Kit; Product Code: ASK-04001-BID2
- Z-1330-2018 · Dec 22, 2017 · Terminated(1) Arterial Catheterization Kit, Part Number (PN): ASK-04018-CC; (2) Arterial Line Kit with Sharps Safety Features, PN: ASK-04510-HUM; (3) Arterial Line Kit with Sharps Safety Fea
- Z-1282-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04020
- Z-1286-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04220
- Z-1285-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04122
- Z-1284-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04120
- Z-1281-2018 · Oct 26, 2017 · TerminatedRadial Artery Catheterization Kit; Model: AK-04018
…and 17 more.
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page